Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display

被引:31
作者
Alawa, Karam AlRahman [1 ]
Nolan, Ryan P. [1 ]
Han, Elaine [1 ]
Arboleda, Alejandro [2 ]
Durkee, Heather [2 ]
Sayed, Mohamed S. [1 ]
Aguilar, Mariela C. [2 ]
Lee, Richard K. [1 ]
机构
[1] Univ Miami, Miller Sch Med, Dept Ophthalmol, Bascom Palmer Eye Inst, Miami, FL 33136 USA
[2] Univ Miami, Miller Sch Med, Ophthalm Biophys Ctr, Bascom Palmer Eye Inst, Miami, FL 33136 USA
关键词
Glaucoma; Field of vision; Telemedicine;
D O I
10.1136/bjophthalmol-2019-314031
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Current visual field screening machines are bulky and expensive, limiting their accessibility, affordability and use. We report the design and evaluation of a novel, portable, cost-effective system for glaucoma screening using smartphone-based visual field screening using frequency doubling technology (FDT) and a head-mounted display. Methods Nineteen eyes of 10 subjects with new-onset or chronic primary open angle glaucoma were tested and compared with the Humphrey Zeiss FDT and the newly developed Mobile Virtual Perimetry (MVP) FDT with the C-20 testing pattern. Mann-Whitney, Bland-Altman and linear regression analyses were performed to assess statistical difference, agreement and correlation, respectively, between the two devices. Results The average age of the participants was 58 +/- 15 years. No statistically significant difference was found between the MVP FDT and the Humphrey Zeiss FDT (p>0.05). Bland-Altman and linear regression analyses demonstrated good agreement and correlation between the two devices. Conclusion The MVP FDT is a low-cost, portable visual field screening device that produces comparable results to the Humphrey Zeiss FDT and may be used as an easily accessible screening tool for glaucoma.
引用
收藏
页码:440 / 444
页数:5
相关论文
共 18 条
[1]   Normal aging effects for frequency doubling technology perimetry [J].
Adams, CW ;
Bullimore, MA ;
Wall, M ;
Fingeret, M ;
Johnson, CA .
OPTOMETRY AND VISION SCIENCE, 1999, 76 (08) :582-587
[2]   Frequency doubling technology perimetry for the detection of glaucomatous visual field loss [J].
Alward, WLM .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2000, 129 (03) :376-378
[3]  
Anderson D.R., 1999, Automated static perimetry, V2nd
[4]  
C.Z.M AG, HUMPHREY FDT USER MA
[5]  
Contestabile MT, 2007, J GLAUCOMA, V16, P297
[6]  
Guirguis-Blake J, 2005, AM FAM PHYSICIAN, V72, P1184
[7]  
Hellman JB, 2016, J OCUL DIS THER, V4, P13
[8]  
Johnson CA, 1997, INVEST OPHTH VIS SCI, V38, P413
[9]   Glaucoma Screening in the Haitian Afro-Caribbean Population of South Florida [J].
L. Bokman, Christine ;
Pasquale, Louis R. ;
Parrish, Richard K., II ;
Lee, Richard K. .
PLOS ONE, 2014, 9 (12)