In vitro collection and posttransfusion engraftment characteristics of MNCs obtained by using a new separator for autologous PBPC transplantation

被引:34
作者
Snyder, EL
Baril, L
Cooper, DL
Min, K
Mechanic, S
Stoddart, L
Burtness, B
Seagraves, P
Debelak, J
Gudino, M
McCullough, J
机构
[1] Yale Univ, Sch Med, Dept Lab Med, New Haven, CT 06510 USA
[2] Yale Univ, Sch Med, Dept Med, New Haven, CT 06510 USA
[3] Univ Minnesota, Dept Pathol, Minneapolis, MN 55455 USA
[4] Baxter Healthcare Corp, Round Lake, IL 60073 USA
关键词
D O I
10.1046/j.1537-2995.2000.40080961.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: A clinical study was performed to evaluate the peripheral blood progenitor cell (PBPC) collection, transfusion, and engraftment characteristics associated with use of a blood cell separator (Amicus, Baxter Healthcare). STUDY DESIGN AND METHODS: Oncology patients (n = 31) scheduled for an autologous PBPC transplant following myeloablative therapy were studied. PBPCs were mobilized by a variety of chemotherapeutic regimens and the use of G-CSF. As no prior studies evaluated whether PBPCs collected on the Amicus separator would be viable after transfusion, to ensure patient safety, PBPCs were first collected on another cell separator (CS-3000 Plus, Baxter) and stored as backup. The day after the CS-3000 Plus collections were completed, PBPC collections intended for transfusion were performed using the Amicus instrument. For each transplant, >2.5 x 10(6) CD34+ PBPCs per kg of body weight were transfused. RESULTS: Clinical data collected on the donors immediately before and after PBPC collection with the Amicus device were comparable to donor data similarly obtained for the CS-3000 Plus collections. While the number of CD34+ cells and the RBC volume in the collected products were equivalent for the two devices, the platelet content of the Amicus collections was significantly lower than that of the CS-3000 Plus collections (4.35 x 10(10) platelets/ bag vs. 6.61 x 10(10) platelets/bag, p<0.05). Collection efficiencies for CD34+ cells were 64 +/- 23 percent for the Amicus device and 43 +/- 14 percent for the CS-3000 Plus device (p<0.05). The mean time to engraftment for cells collected via the Amicus device was 8.7 +/- 0.7 days for >500 PMNs per mu L and 9.7 +/- 1.5 days to attain a platelet count of >20,000 per mu L-equivalent to data in the literature. No CS-3000 Plus backup cells were transfused and no serious adverse events attributable to the Amicus device were encountered. CONCLUSIONS: The mean Amicus CD34+ cell collection efficiency was better (p<0.05) than that of the CS-3000 Plus collection. Short-term engraftment was durable. The PBPCs collected with the Amicus separator are safe and effective for use for autologous transplant patients requiring PBPC rescue from high-dose myeloablative chemotherapy.
引用
收藏
页码:961 / 967
页数:7
相关论文
共 14 条
[1]  
BURGSTALER E, 1996, APHERESIS PRINCIPLES, P96
[2]  
Burgstaler EA, 1997, J CLIN APHERESIS, V12, P55, DOI 10.1002/(SICI)1098-1101(1997)12:2<55::AID-JCA1>3.0.CO
[3]  
2-8
[4]   Factors affecting the efficiency of collection of CD34-positive peripheral blood cells by a blood cell separator [J].
Ford, CD ;
Pace, N ;
Lehman, C .
TRANSFUSION, 1998, 38 (11-12) :1046-1050
[5]  
LANGENMAYER I, 1995, BONE MARROW TRANSPL, V15, P241
[6]  
Leitman SF, 1996, SEMIN HEMATOL, V33, P341
[7]   Collection of peripheral blood progenitor cells with an automated leukapheresis system [J].
Rowley, SD ;
Prather, K ;
Bui, KT ;
Appel, M ;
Felt, T ;
Bensinger, WI .
TRANSFUSION, 1999, 39 (11-12) :1200-1206
[8]   Collection of allogeneic peripheral blood progenitor cells by two protocols on an apheresis system [J].
Schreiner, T ;
Wiesneth, M ;
Krug, E ;
Kalinova, J ;
Büchele, H ;
Maccari, B ;
Erne, E ;
Bischof, C ;
De Reys, S .
TRANSFUSION, 1998, 38 (11-12) :1051-1055
[9]  
Smith Brian R., 1997, P2621
[10]   Kinetics of peripheral blood stem cell harvests during a single apheresis [J].
Smolowicz, AG ;
Villman, K ;
Berlin, G ;
Tidefelt, U .
TRANSFUSION, 1999, 39 (04) :403-409