A comparison of intraocular pressure values obtained using a Goldmann applanation tonometer and a handheld version of applanation resonance tonometer: A preliminary report

被引:1
作者
Mulak, Malgorzata [1 ]
Czak, Wojciech A. [1 ]
Mimier, Malgorzata [1 ]
Kaczmarek, Radoslaw [1 ]
机构
[1] Wroclaw Med Univ, Dept & Clin Ophthalmol, Wroclaw, Poland
来源
ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE | 2018年 / 27卷 / 04期
关键词
comparison; glaucoma; ART; tonometry; GAT; SENSOR;
D O I
10.17219/acem/68559
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background. Despite the development of various methods of intraocular pressure (IOP) measurement, Goldmann applanation tonometry (GAT) is still the most popular. The measurement using GAT depends on the biomechanical properties of the cornea, such as the thickness, the radius of curvature, as well as the amount of the fluorescein used. Objectives. The aim of the study was to compare IOP values measured by GAT with those measured by applanation resonance tonometry (ART). Material and methods. A total of 47 patients (94 eyes), including 28 patients with primary open-angle glaucoma (POAG) and 19 subjects from the control group, were examined at the Glaucoma Outpatient Clinic of the Department and Clinic of Ophthalmology at Wroclaw Medical University (Poland). The measurements of IOP were performed using GAT and a handheld version of ART. Also, the central corneal thickness (CCT) of all patients was measured. Results. The study showed that the IOP values measured by both tonometers were comparable, but ART-acquired values were higher than GAT-obtained values both in the glaucomatous group and in the control group. CCT had little impact on mean IOP difference between GAT- and ART-obtained values. Conclusions. Applanation resonance tonometry is a precise method of IOP measurement and is less affected by biomechanical properties of the cornea than GAT. Our results show that ART is a new, promising, comfortable for both patients and doctors method of IOP measurement, which, in the future, can replace GAT.
引用
收藏
页码:481 / 485
页数:5
相关论文
共 18 条
  • [1] [Anonymous], 2014, TERM GUID GLAUC
  • [2] Beckman RJ, 2014, JAMA OPHTHALMOL, V132, P1426, DOI DOI 10.1001/jamaophthalmol.2014.3029
  • [3] Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry
    Cacho, Isabel
    Sanchez-Naves, Juan
    Batres, Laura
    Pintor, Jesus
    Carracedo, Gonzalo
    [J]. JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [4] Clement Colin I, 2016, Open Ophthalmol J, V10, P35, DOI 10.2174/1874364101610010035
  • [5] An applanation resonator sensor for measuring intraocular pressure using combined continuous force and area measurement
    Eklund, A
    Hallberg, P
    Lindén, C
    Lindahl, OA
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (07) : 3017 - 3024
  • [6] GOLDMANN H, 1957, Ophthalmologica, V134, P221
  • [7] Applanation resonance tonometry for intraocular pressure in humans
    Hallberg, P
    Lindén, C
    Lindahl, OA
    Bäcklund, T
    Eklund, A
    [J]. PHYSIOLOGICAL MEASUREMENT, 2004, 25 (04) : 1053 - 1065
  • [8] Hallberg P., 2006, Journal of Medical Engineering & Technology, V30, P345, DOI 10.1080/03091900512331332528
  • [9] Clinical evaluation of applanation resonance tonometry -: A comparison with Goldmann applanation tonometry
    Hallberg, Per
    Eklund, Anders
    Backlund, Tomas
    Linden, Christina
    [J]. JOURNAL OF GLAUCOMA, 2007, 16 (01) : 88 - 93
  • [10] Symmetric sensor for applanation resonance tomometry of the eye
    Hallberg, Per
    Linden, Christina
    Backlund, Tomas
    Eklund, Anders
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (1-2) : 54 - 60