Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers

被引:20
作者
Cho, Jooyoung [1 ,2 ]
Oh, Joowon [1 ,3 ]
Lee, Sang-Guk [1 ]
Lee, You-Hee [1 ]
Song, Jaewoo [1 ]
Kim, Jeong-Ho [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Dept Lab Med, Coll Med, 50-1 Yonsei Ro, Seoul 03722, South Korea
[2] Yonsei Univ, Wonju Severance Christian Hosp, Dept Lab Med, Wonju Coll Med, Wonju, South Korea
[3] Yonsei Univ, Gangnam Severance Hosp, Dept Lab Med, Coll Med, Seoul, South Korea
关键词
Body fluid; Automated cellular analyzer; Differential cell counting; UniCel DxH 800; XN-350; UF-5000; Analytical performance; REFLEX TESTING RULES; COUNTING BLOOD-CELLS; CEREBROSPINAL-FLUID; HEMATOLOGY ANALYZERS; MODE; VALIDATION; UF-1000I; XN-1000;
D O I
10.3343/alm.2020.40.2.122
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Automated cellular analyzers are expected to improve the analytical performance in body fluid (BF) analysis. We evaluated the analytical performance of three automated cellular analyzers and established optimum reflex analysis guidelines. Methods: A total of 542 BF samples (88 cerebrospinal fluid [CSF] samples and 454 non-CSF samples) were examined using manual counting and three automated cellular analyzers: UniCel DxH 800 (Beckman Coulter), XN-350 (Sysmex), and UF-5000 (Sysmex). Additionally, 2,779 BF analysis results were retrospectively reviewed. For malignant cell analysis, the receiver operating characteristic (ROC) curve was used, and the detection of high fluorescence-BF cells (HF-BFs) using the XN-350 analyzer was compared with cytology results. Results: All three analyzers showed good agreement for total nucleated cell (TNC) and red blood cell (RBC) counts, except for the RBC count in CSF samples using the UniCel DxH 800. However, variable degrees of differences were observed during differential cell counting. For malignant cell analysis, the area under the curve was 0.63 for the XN-350 analyzer and 0.76 for manual counting. We established our own reflex analysis guidelines as follows: HF-BFs <0.7/100 white blood cells (WBCs) is the criterion for quick scans with 100x magnification microscopic examination as a rule-out cut-off, while HF-BFs >83.4/100 WBCs or eosinophils >3.8% are the criteria for mandatory double check confirmation with 1,000x magnification examination. Conclusions: The three automated analyzers showed good analytical performances. Application of reflex analysis guidelines is recommended for eosinophils and HF-BFs, and manual confirmation is warranted.
引用
收藏
页码:122 / 131
页数:10
相关论文
共 25 条
[1]  
[Anonymous], DESIRABLE BIOL VARIA
[2]  
Beckman Coulter Inc. UniCel DxH 800, 2017, B26647AE PN BECKM CO, P127
[3]   ICSH guidelines for the verification and performance of automated cell counters for body fluids [J].
Bourner, G. ;
De la Salle, B. ;
George, T. ;
Tabe, Y. ;
Baum, H. ;
Culp, N. ;
Keng, T. B. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (06) :598-612
[4]   Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i [J].
Buoro, Sabrina ;
Esposito, Sara Apassiti ;
Alessio, MariaGrazia ;
Crippa, Alberto ;
Ottomano, Cosimo ;
Lippi, Giuseppe .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2016, 30 (05) :381-391
[5]   Reflex Testing Rules for Cell Count and Differentiation of Nucleated Elements in Pleural and Ascitic Fluids on Sysmex XE-5000 [J].
Buoro, Sabrina ;
Esposito, Sara Appassiti ;
Vavassori, Mauro ;
Mecca, Tommaso ;
Ottomano, Cosimo ;
Dominoni, Paola ;
Seghezzi, Michela ;
Candiago, Elisabetta ;
Farina, Claudio ;
Gianatti, Andrea ;
Crippa, Alberto ;
Lippi, Giuseppe .
JALA, 2016, 21 (02) :297-304
[6]   Body fluid cellular analysis using the Sysmex XN-2000 automatic hematology analyzer: focusing on malignant samples [J].
Cho, Y. -U. ;
Chi, H. -S. ;
Park, S. H. ;
Jang, S. ;
Kim, Y. -J. ;
Park, C. -J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (03) :346-356
[7]   Validation of reflex testing rules and establishment of a new workflow for body fluid cell analysis using a Sysmex XN-550 automatic hematology analyzer [J].
Cho, Y-U ;
You, E. ;
Jang, S. ;
Park, C-J .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (03) :258-267
[8]  
Clinical and Laboratory Standards Institution (CLSI), 2006, Body fluid analysis for cellular composition
[9]  
CLSI, 2006, 15A2 CLSI
[10]  
CLSI, 2003, EP064 CLSI