Rigid fixation of the sternum using a new coupled titanium transverse plate fixation system

被引:16
作者
Hallock, Geoffrey G.
Szydlowski, Gary W.
机构
[1] Lehigh Valley Hosp Ctr, Div Plast Surg, Allentown, PA 18102 USA
[2] Lehigh Valley Hosp Ctr, Div Cardiothorac Surg, Allentown, PA 18102 USA
关键词
sternal dehiscence; titanium locked; reconstruction plate; transverse rigid plate fixation; sternum reconstruction;
D O I
10.1097/01.sap.0000248134.37753.1a
中图分类号
R61 [外科手术学];
学科分类号
摘要
Restoration of sternal integrity after median sternotomy for cardiac interventions better ensures optimal postoperative pulmonary function and minimizes overall morbidity. Sternal dehiscence or nonunion mitigates against such a successful outcome. Under such circumstances, if enough viable and uninfected sternum remains, an anatomic reduction should be attempted. Rewiring usually proves unsuccessful, and rigid plate fixation is more rewarding. A new titanium sternal fixation system that permits transverse orientation of plates has been used in 4 patients who had sterile complete or imminent sternal dehiscence to allow eventual sternal union. The specific advantage of this new system is the presence of a releasing pin in the center of coupled plates to allow rapid chest reentry if required without the specific need for cumbersome plate removal. One patient had delayed removal of an infected plate after sternal union was achieved.
引用
收藏
页码:640 / 644
页数:5
相关论文
共 11 条
[1]   Vacuum-assisted closure for sternal wounds: A first-line therapeutic management approach [J].
Agarwal, JP ;
Ogilvie, M ;
Wu, LC ;
Lohman, RF ;
Gottlieb, LJ ;
Franczyk, M ;
Song, DH .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2005, 116 (04) :1035-1040
[2]   Management of sternal wounds with bilateral pectoralis major myocutaneous advancement flaps in 114 consecutively treated patients: Refinements in technique and outcomes analysis [J].
Ascherman, JA ;
Patel, SM ;
Malhotra, SM ;
Smith, CR .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2004, 114 (03) :676-683
[3]   Internal fixation of the sternum in median sternotomy dehiscence [J].
Chase, CW ;
Franklin, JD ;
Guest, DP ;
Barker, DE .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1999, 103 (06) :1667-1673
[4]   Sternal wound reconstruction with transverse plate fixation [J].
Cicilioni, OJ ;
Stieg, FH ;
Papanicolaou, G .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2005, 115 (05) :1297-1303
[5]  
GOTTLIEB LJ, 1994, ARCH SURG-CHICAGO, V129, P489
[6]  
HALLOCK GG, 1995, CAN J PLAST SURG, V3, P127
[7]   Sternal plating for the treatment of sternal nonunion [J].
Hendrickson, SC ;
Koger, KE ;
Morea, CJ ;
Aponte, RL ;
Smith, PK ;
Levin, LS .
ANNALS OF THORACIC SURGERY, 1996, 62 (02) :512-518
[8]   Composite plate and wire fixation for complicated sternal closure [J].
Mitra, A ;
Elahi, MM ;
Tariq, GB ;
Mir, H ;
Powell, R ;
Spears, J .
ANNALS OF PLASTIC SURGERY, 2004, 53 (03) :217-221
[9]   The Waist Coat Flap: a new technique for closure of infected median sternotomy wounds [J].
Morsi, AW ;
Kimble, F ;
Quarmby, C ;
Tucker, A .
BRITISH JOURNAL OF PLASTIC SURGERY, 2004, 57 (08) :728-732
[10]   Paramedian sternal bone plate reinforcement and wiring for difficult sternotomy wounds [J].
Smoot, EC ;
Weiman, D .
ANNALS OF PLASTIC SURGERY, 1998, 41 (05) :464-467