Pattern of isthmus conduction recovery using open cooled and solid large-tip catheters for radiofrequency ablation of typical atrial flutter

被引:30
作者
Ventura, R [1 ]
Klemm, H [1 ]
Lutomsky, B [1 ]
Demir, C [1 ]
Rostock, T [1 ]
Weiss, C [1 ]
Meinertz, T [1 ]
Willems, S [1 ]
机构
[1] Univ Hamburg, Hosp Eppendorf, Dept Cardiol, D-20246 Hamburg, Germany
关键词
isthmus recovery atrial flutter ablation;
D O I
10.1046/j.1540-8167.2004.04125.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cooled Versus Large-Tip Catheters for Flutter Ablation. Introduction: Open cooled-tip and solid 8-mm-tip catheters have demonstrated safety and effectiveness for radiofrequency current (RFC) ablation of typical atrial flutter (AFL). However, data from prospective and randomized studies in this setting are lacking. Methods and Results: One hundred thirty consecutive patients (104 men; 61 +/- 11 years) with AFL were randomized to undergo RFC catheter ablation either using a solid 8-mm-tip catheter (group A, 65degreesC, 70 W, 60 s) or an open irrigated-tip catheter (group B, 65degreesC, 50 W, 60 s, 17 mL/min flow). Endpoint was bidirectional conduction isthmus block. In cases of repeated (two times) transient isthmus block, the catheter was changed (crossed over) to the catheter used in the other randomization arm, but patients remained in the original group following intention-to-treat analysis. The selected endpoint could be achieved in all patients after 12 +/- 6 RFC pulses in group A and 10 +/- 7 RFC pulses in group B (P = 0.11). Procedure times were longer (159 +/- 38 min vs 138 +/- 37 min, P = 0.002) and x-ray exposures higher in group A (fluoroscopy time 25 +/- 17 min vs 21 +/- 10 min, P = 0.08; x-ray dosage 3,133 +/- 2,576 cGy . cm(2) vs 2,326 +/- 1,405 cGy . cm, P = 0.03). Transient isthmus block was observed in 23 group A patients and 12 group B patients (P = 0.03). Onset time of transient isthmus block ranged from 0.5 to 27 minutes. Repeated transient isthmus block occurred in 8 of the 23 patients in group A after 19 3 RFC applications. After crossover to the cooled-tip catheter, the endpoint was reached another 5 I RFC pulses. In group B, all patients could be treated without change of ablation catheter. After a follow-up of 14 +/- 2 months, 2 patients (3%) in group A and 1 patient (1.5%) in group B presented with AFL recurrence. Conclusion: Open cooled-tip catheters are more effective than solid large-tip catheters for AFL ablation. The greater effectiveness is evident in cases showing repeated conduction recovery within the cavotricuspid isthmus. Primary use of open irrigated-tip catheters should be considered for AFL ablation.
引用
收藏
页码:1126 / 1130
页数:5
相关论文
共 17 条
[1]   Angiographic anatomy of the inferior right atrial isthmus in patients with and without history of common atrial flutter [J].
Cabrera, JA ;
Sanchez-Quintana, D ;
Ho, SY ;
Medina, A ;
Wanguemert, F ;
Gross, E ;
Grillo, J ;
Hernandez, E ;
Anderson, RH .
CIRCULATION, 1999, 99 (23) :3017-3023
[2]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[3]   Living anatomy of the atrioventricular junctions.: A guide to electrophysiologic mapping -: A consensus statement from the Cardiac Nomenclature Study Group, Working Group of Arrhythmias, European Society of Cardiology, and the Task Force on Cardiac Nomenclature from NASPE [J].
Cosío, FG ;
Anderson, RH ;
Kuck, KH ;
Becker, A ;
Borggrefe, M ;
Campbell, RWF ;
Gaita, F ;
Guiraudon, GM ;
Haïssaguerre, M ;
Rufilanchas, JJ ;
Thiene, G ;
Wellens, HJJ ;
Langberg, J ;
Benditt, DG ;
Bharati, S ;
Klein, G ;
Marchlinski, F ;
Saksena, S .
CIRCULATION, 1999, 100 (05) :E31-E37
[4]   Comparison of irrigated electrode designs for radiofrequency ablation of myocardium [J].
Demazumder, D ;
Mirotznik, MS ;
Schwartzman, D .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (04) :391-400
[5]   Biophysics of radiofrequency ablation using an irrigated electrode [J].
Demazumder, D ;
Mirotznik, MS ;
Schwartzman, D .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (04) :377-389
[6]   RADIOFREQUENCY CATHETER ABLATION FOR THE TREATMENT OF HUMAN TYPE-1 ATRIAL-FLUTTER - IDENTIFICATION OF A CRITICAL ZONE IN THE REENTRANT CIRCUIT BY ENDOCARDIAL MAPPING TECHNIQUES [J].
FELD, GK ;
FLECK, RP ;
CHEN, PS ;
BOYCE, K ;
BAHNSON, TD ;
STEIN, JB ;
CALISI, CM ;
IBARRA, M .
CIRCULATION, 1992, 86 (04) :1233-1240
[7]   Right atrial angiographic evaluation of the posterior isthmus -: Relevance for ablation of typical atrial flutter [J].
Heidbüchel, H ;
Willems, R ;
van Rensburg, H ;
Adams, J ;
Ector, H ;
Van de Werf, F .
CIRCULATION, 2000, 101 (18) :2178-2184
[8]   Prospective randomized comparison of irrigated-tip versus conventional-tip catheters for ablation of common flutter [J].
Jaïs, P ;
Shah, DC ;
Haissaguerre, M ;
Hocini, M ;
Garrigue, S ;
Le Metayer, P ;
Clémenty, J .
CIRCULATION, 2000, 101 (07) :772-776
[9]   Effectiveness of irrigated tip catheter ablation of common atrial flutter [J].
Jaïs, P ;
Hocini, M ;
Gillet, T ;
Shah, DC ;
Haïssaguerre, M ;
Yamane, T ;
Deisenhofer, I ;
Garrigue, S ;
Le Métayer, P ;
Roudaut, R ;
Clémenty, J .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 88 (04) :433-+
[10]   Phased-array intracardiac echocardiography for defining cavotricuspid isthmus anatomy during radiofrequency ablation of typical atrial flutter [J].
Morton, JB ;
Sanders, P ;
Davidson, NC ;
Sparks, PB ;
Vohra, JK ;
Kalman, JM .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (06) :591-597