Lab-in-a-syringe using gold nanoparticles for rapid immunosensing of protein biomarkers

被引:47
作者
Nunes Pauli, Gisele Elias [1 ,2 ]
de la Escosura-Muniz, Alfredo [1 ]
Parolo, Claudio [1 ]
Bechtold, Ivan Helmuth [2 ]
Merkoci, Arben [1 ,3 ]
机构
[1] ICN2 Inst Catala Nanociencia & Nanotecnol, Nanobioelect & Biosensors Grp, Barcelona 08193, Spain
[2] Univ Fed Santa Catarina, Dept Fis, Florianopolis, SC, Brazil
[3] ICREA, Barcelona 08010, Spain
关键词
PROSTATE-SPECIFIC ANTIGEN; LOCAL RECURRENCE; 10; NG/ML; PAPER; SENSITIVITY; BIOSENSOR; CANCER; URINE; STRIP; ASSAY;
D O I
10.1039/c4lc01123f
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have developed a paper and gold nanoparticle (AuNP)-based lab-in-a-syringe (LIS) for immunosensing of biomarkers. This simple diagnostic device features simultaneous sampling and vertical-flow operation, which means that unlike typical immunosensors, it does not suffer from any delay between sampling and detection. It can handle large-volume, low-concentration samples for analysis in diverse applications (e.g. biomedical, environmental, food, etc.). Furthermore, its operating range for sample concentration can be tuned by simply changing the volume of the syringed sample, which enables on-demand limits of detection (LOD). The LIS contains two nitrocellulose pads: the conjugate pad (which captures the analyte) and the detection pad (which signals the presence of the captured analyte) both embedded into reusable plastic cartridges. We demonstrated its efficiency in detecting human IgG (HIgG) (LOD: 1.0 ng mL(-1)) and prostate-specific antigen (PSA) (spiked urine samples; LOD: 1.9 ng mL(-1)). In the field, the LIS can be used for complete on-site analysis or to obtain partially analyzed samples (AuNPs with captured analyte) for subsequent detailed testing in specialized laboratories.
引用
收藏
页码:399 / 405
页数:7
相关论文
共 25 条
  • [1] Double-codified gold nanolabels for enhanced immunoanalysis
    Ambrosi, Adriano
    Castaneda, Maria Teresa
    Killard, Anthony J.
    Smyth, Malcolm R.
    Alegret, Salvador
    Merkoci, Arben
    [J]. ANALYTICAL CHEMISTRY, 2007, 79 (14) : 5232 - 5240
  • [2] Bolduc S, 2007, CUAJ-CAN UROL ASSOC, V1, P377
  • [3] Bolduc S, 2009, CUAJ-CAN UROL ASSOC, V3, P213
  • [4] Serum-to-urinary prostate specific antigen ratio: Its impact in distinguishing prostate cancer when serum prostate specific antigen level is 4 to 10 NG/ML.
    Irani, J
    Millet, C
    Levillain, P
    Dore, B
    Begon, F
    Aubert, J
    [J]. JOURNAL OF UROLOGY, 1997, 157 (01) : 185 - 188
  • [5] Biodistribution and toxicity of engineered gold nanoparticles: a review of in vitro and in vivo studies
    Khlebtsov, Nikolai
    Dykman, Lev
    [J]. CHEMICAL SOCIETY REVIEWS, 2011, 40 (03) : 1647 - 1671
  • [6] A SOL PARTICLE AGGLUTINATION ASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    LEUVERING, JHW
    THAL, PJHM
    VANDERWAART, M
    SCHUURS, AHWM
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 1981, 45 (02) : 183 - 194
  • [7] Rapid and Sensitive Detection of Protein Biomarker Using a Portable Fluorescence Biosensor Based on Quantum Dots and a Lateral Flow Test Strip
    Li, Zhaohui
    Wang, Ying
    Wang, Jun
    Tang, Zhiwen
    Pounds, Joel G.
    Lin, Yuehe
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (16) : 7008 - 7014
  • [8] A lateral flow biosensor for detection of nucleic acids with high sensitivity and selectivity
    Lie, Puchang
    Liu, Jie
    Fang, Zhiyuan
    Dun, Boying
    Zeng, Lingwen
    [J]. CHEMICAL COMMUNICATIONS, 2012, 48 (02) : 236 - 238
  • [9] All-Integrated and Highly Sensitive Paper Based Device with Sample Treatment Platform for Cd2+ Immunodetection in Drinking/Tap Waters
    Lopez Marzo, Adaris M.
    Pons, Josefina
    Blake, Diane A.
    Merkoci, Arben
    [J]. ANALYTICAL CHEMISTRY, 2013, 85 (07) : 3532 - 3538
  • [10] High sensitive gold-nanoparticle based lateral flow Immunodevice for Cd2+ detection in drinking waters
    Lopez-Marzo, Adaris M.
    Pons, Josefina
    Blake, Diane A.
    Merkoci, Arben
    [J]. BIOSENSORS & BIOELECTRONICS, 2013, 47 : 190 - 198