Effects of medically generated electromagnetic interference from medical devices on cardiac implantable electronic devices: A review

被引:7
作者
Barmore, Walker [1 ,4 ]
Patel, Himax [2 ]
Voong, Cassandra [2 ]
Tarallo, Caroline [3 ]
Calkins, Joe B., Jr. [1 ]
机构
[1] Augusta Univ, Dept Cardiol, Med Ctr, Augusta, GA 30912 USA
[2] Augusta Univ, Dept Internal Med, Med Ctr, Augusta, GA 30912 USA
[3] Med Coll Georgia, Augusta, GA 30912 USA
[4] Augusta Univ, Med Ctr, Dept Cardiol, 1120 15th St, Augusta, GA 30912 USA
关键词
Electromagnetic interference; Pacemaker; Implantable cardioverter defibrillator; Permanent pacemakers; Cardiac implantable electronic devices; Left ventricular assist device; Endoscopy; Bronchoscopy; Electrocautery; Capsule endoscopy; Transcutaneous electrical nerve stimulators unit; Spinal cord stimulator; AMERICAN-HEART-ASSOCIATION; CARDIOVERTER-DEFIBRILLATOR; RHYTHM SOCIETY; MANAGEMENT; STIMULATION; PACEMAKERS; THERAPY; PATIENT;
D O I
10.4330/wjc.v14.i8.446
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
As cardiac implantable electronic devices (CIED) become more prevalent, it is important to acknowledge potential electromagnetic interference (EMI) from other sources, such as internal and external electronic devices and procedures and its effect on these devices. EMI from other sources can potentially inhibit pacing and trigger shocks in permanent pacemakers (PPM) and implantable cardioverter defibrillators (ICD), respectively. This review analyzes potential EMI amongst CIED and left ventricular assist device, deep brain stimulators, spinal cord stimulators, transcutaneous electrical nerve stimulators, and throughout an array of procedures, such as endoscopy, bronchoscopy, and procedures involving electrocautery. Although there is evidence to support EMI from internal and external devices and during procedures, there is a lack of large multicenter studies, and, as a result, current management guidelines are based primarily on expert opinion and anecdotal experience. We aim to provide a general overview of PPM/ICD function, review documented EMI effect on these devices, and acknowledge current management of CIED interference.
引用
收藏
页码:446 / 453
页数:8
相关论文
共 27 条
[1]  
Al-Khatib SM, 2018, CIRCULATION, V138, pE210, DOI [10.1161/CIR.0000000000000548, 10.1161/CIR.0000000000000549]
[2]   Do endoscopy capsules interfere with implantable cardioverter-defibrillators? [J].
Bandorski, D. ;
Irnich, W. ;
Brueck, M. ;
Kramer, W. ;
Jakobs, R. .
ENDOSCOPY, 2009, 41 (05) :457-461
[3]  
Chaiban GM, 2013, PAIN PHYSICIAN, V16, P489
[4]   Effects of surgical and endoscopic electrocautery on modern-day permanent pacemaker and implantable cardioverter-defibrillator systems [J].
Cheng, Alan ;
Nazarian, Saman ;
Spragg, David D. ;
Bilchick, Kenneth ;
Tandri, Harikrishna ;
Mark, Lynette ;
Halperin, Henry ;
Calkins, Hugh ;
Berger, Ronald D. ;
Henrikson, Charles A. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (03) :344-350
[5]   Biventricular Pacing Versus Right Ventricular Pacing in Patients Supported With LVAD [J].
Chung, Ben B. ;
Grinstein, Jonathan S. ;
Imamura, Teruhiko ;
Kruse, Eric ;
Nguyen, Ann B. ;
Narang, Nikhil ;
Holzhauser, Luise H. ;
Burkhoff, Daniel ;
Lang, Roberto M. ;
Sayer, Gabriel T. ;
Uriel, Nir Y. .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (08) :1003-1009
[6]  
Crossley GH, 2011, HEART RHYTHM, V8, pE1, DOI 10.1016/j.hrthm.2011.05.010
[7]  
ecgwaves, COMPONENTS ANDCONSTR
[8]   Pacemaker and Defibrillator Implantation and Programming in Patients with Deep Brain Stimulation [J].
Elliott, Mark ;
Momin, Sheikh ;
Fiddes, Barnaby ;
Farooqi, Fahad ;
Sohaib, S. M. Afzal .
ARRHYTHMIA & ELECTROPHYSIOLOGY REVIEW, 2019, 8 (02) :138-142
[9]   Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device [J].
Erqou, Sebhat ;
Kormos, Robert L. ;
Wang, Norman C. ;
McNamara, Dennis M. ;
Bazaz, Raveen .
CASE REPORTS IN CARDIOLOGY, 2018, 2018
[10]   16-Year Trends in the Infection Burden for Pacemakers and Implantable Cardioverter-Defibrillators in the United States 1993 to 2008 [J].
Greenspon, Arnold J. ;
Patel, Jasmine D. ;
Lau, Edmund ;
Ochoa, Jorge A. ;
Frisch, Daniel R. ;
Ho, Reginald T. ;
Pavri, Behzad B. ;
Kurtz, Steven M. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 58 (10) :1001-1006