Pure Cusp Tear of Trifecta Bioprosthesis 2 Years after Aortic Valve Replacement

被引:28
作者
Hamamoto, Masaki [1 ]
Kobayashi, Taira [1 ]
Ozawa, Masamichi [1 ]
Yoshimura, Kosuke [1 ]
机构
[1] JA Hiroshima Gen Hosp, Dept Cardiovasc Surg, 1-3-3 Jigozen, Hatsukaichi, Hiroshima 7388503, Japan
关键词
pericardial bioprosthesis; structural valve deterioration; aortic valve replacement; Trifecta; HEART-VALVES;
D O I
10.5761/atcs.cr.16-00183
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Trifecta is a stented bioprosthetic heart valve with a bovine pericardial sheet externally mounted on a titanium stent. This valve is applied only for aortic valve replacement (AVR), providing excellent hemodynamics and extremely low incidence of structural valve deterioration (SVD). A 76-year-old woman presented with dyspnea on effort 24 months after AVR with a 21-mm Trifecta valve. Echocardiography revealed severe aortic regurgitation with prolapse of a cusp of Trifecta valve, which suggested that she developed acute heart failure due to early SVD. In the operation, Trifecta valve had a cusp tear near the commissure with circumferential fibrous pannus ingrowth only at the inflow side. There was neither calcification nor infection. The Trifecta valve was successfully replaced with a new porcine bioprosthesis.
引用
收藏
页码:157 / 160
页数:4
相关论文
共 10 条
[1]   The St Jude Medical Trifecta aortic pericardial valve: Results from a global, multicenter, prospective clinical study [J].
Bavaria, Joseph E. ;
Desai, Nimesh D. ;
Cheung, Anson ;
Petracek, Michael R. ;
Groh, Mark A. ;
Borger, Michael A. ;
Schaff, Hartzell V. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (02) :590-597
[2]   Early failures of Trifecta aortic bioprosthesis [J].
Campisi, Salvatore ;
Camilleri, Lionel ;
Innorta, Andrea ;
Azarnoush, Kasra .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 148 (01) :E133-E134
[3]   COMPUTER MODELING OF BIOPROSTHETIC HEART-VALVES [J].
CHRISTIE, GW .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 :S95-S101
[4]  
Haziza F, 1996, J HEART VALVE DIS, V5, P35
[5]  
Modi Amit, 2015, Asian Cardiovasc Thorac Ann, V23, P140, DOI 10.1177/0218492314533684
[6]   St. Jude Medical Trifecta™ aortic valve perioperative performance in 200 patients [J].
Permanyer, Eduard ;
Estigarribia, Arnaldo-Javier ;
Ysasi, Alejandro ;
Herrero, Enrique ;
Semper, Omar ;
Llorens, Rafael .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2013, 17 (04) :669-672
[7]   Leaflet tear in a Trifecta aortic bioprosthesis 34 months after implantation [J].
Pinon, Miguel ;
Duran, Dario ;
Pazos, Pablo ;
Pradas, Gonzalo .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2015, 20 (02) :281-282
[8]   Early structural valve deterioration of the Trifecta aortic valve biological prosthesis: A word of caution [J].
Saxena, Pankaj ;
Greason, Kevin L. ;
Schaff, Hartzell V. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (01) :E10-E11
[9]   Bioprosthetic heart valves: modes of failure [J].
Siddiqui, Raheela Fareed ;
Abraham, Johnathan Rajiv ;
Butany, Jagdish .
HISTOPATHOLOGY, 2009, 55 (02) :135-144
[10]   The evolution of bioprosthetic heart valve design and its impact on durability [J].
Vesely, I .
CARDIOVASCULAR PATHOLOGY, 2003, 12 (05) :277-286