Continuous flow system for lead determination by faas in spirituous beverages with solid phase extraction and on-line copper removal

被引:30
作者
de Peña, YP
Paredes, B
Rondón, W
Burguera, M
Burguera, JL
Rondón, C
Carrero, P
Capote, T
机构
[1] Univ Los Andes, Fac Ciencias, Dept Quim, Lab Espectroscopia Mol, Merida 5101 A, Venezuela
[2] Univ Ctr Occidental Lisandro Alvarado, Dept Quim & Suelos, Unidad Anal Instrumental, Decanato Agr, Barquisimeto, Venezuela
关键词
flame atomic absorption spectrometry; continuous flow system; synthetic zeolite; lead; alcoholic drinks; copper removal; spirituous beverages;
D O I
10.1016/j.talanta.2004.05.053
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A continuous flow system for the determination of lead in home made spirituous beverages was developed. The determination was based on the formation of a neutral chelate of the element with ammonium pyrrolidine dithiocarbamate, its adsorption onto a minicolumn packed with sodium faujasite type Y synthetic zeolite, followed by elution with methyl isobutyl ketone and determination by flame atomic absorption spectrometry. Ethanol and copper interfere strongly in the determination and therefore, must be separated prior to the analysis. Copper is removed by precipitation with rubeanic acid, while ethanol is eliminated by rotaevaporation. Sample solutions containing Pb2+ in the concentration range from 5 to 120 mug = l(-1) at pH 2.5 could be analyzed, by using a preconcentration time of 3 min. Preconcentration factors from 80 to 140 were achieved for a sample volume of 6 ml and the detection limit varied from 1.4 to 3.5 mug l(-1), depending on the matrix composition. The relative standard deviations for 60 mug l(-1) Pb was 3.2% (n = 10) and the recovery of spikes (20, 40, 60 and 80 mug l(-1)) added to the samples was estimated within 92-105% range, suggesting that lead can be quantitatively determined in such samples. Determining lead in several samples by an alternative technique further checked the accuracy. Finally, the concentrations of Pb2+ determined in 28 samples of Venezuelan spirituous beverages were in 12.6-370.0 mug l(-1) range, depending on the fermenting material based on different mixtures of agave, raw sugar cane and white sugar. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1351 / 1358
页数:8
相关论文
共 34 条
[1]   DETERMINATION OF LEAD AND ALUMINUM IN PORT WINE BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
ALMEIDA, AA ;
BASTOS, ML ;
CARDOSO, MI ;
FERREIRA, MA ;
LIMA, JLFC ;
SOARES, ME .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (08) :1281-1285
[2]  
[Anonymous], 1990, REC METH INT AN VINS
[3]  
Baldo MA, 1997, ANALYST, V122, P1, DOI 10.1039/a605370j
[4]   DETERMINATION OF LEAD AND ARSENIC IN WINES BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BRUNO, SNF ;
CAMPOS, RC ;
CURTIUS, AJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :341-344
[5]   DETERMINATION OF LEAD IN WINE, OTHER BEVERAGES AND FRUIT SLURRIES BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH ONLINE MICROWAVE DIGESTION [J].
CABRERA, C ;
MADRID, Y ;
CAMARA, C .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1423-1426
[6]   DETERMINATION OF LEAD IN WINES BY HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
CACHO, J ;
FERREIRA, V ;
NERIN, C .
ANALYST, 1992, 117 (01) :31-33
[7]  
CONTRERAS R, 1999, THESIS U LOS ANDES
[8]  
De Peña YP, 2001, ANAL CHIM ACTA, V438, P259
[9]   Synthetic zeolites as sorbent material for on-line preconcentration of copper traces and its determination using flame atomic absorption spectrometry [J].
de Peña, YP ;
López, W ;
Burguera, JL ;
Burguera, M ;
Gallignani, M ;
Brunetto, R ;
Carrero, P ;
Rondon, C ;
Imbert, F .
ANALYTICA CHIMICA ACTA, 2000, 403 (1-2) :249-258
[10]   PRECONCENTRATION OF COPPER TRACES ON C-60-C-70 FULLERENES BY FORMATION OF ION-PAIRS AND CHELATES [J].
DEPENA, YP ;
GALLEGO, M ;
VALCARCEL, M .
ANALYTICAL CHEMISTRY, 1995, 67 (15) :2524-2529