A Monte Carlo model for organ dose reconstruction of patients in pencil beam scanning (PBS) proton therapy for epidemiologic studies of late effects

被引:12
|
作者
Yeom, Yeon Soo [1 ]
Kuzmin, Gleb [2 ]
Griffin, Keith [1 ]
Mille, Matthew [1 ]
Polf, Jerimy [3 ]
Langner, Ulrich [4 ]
Jung, Jae Won [5 ]
Lee, Choonik [6 ]
Ellis, Dillon [5 ]
Shin, Jungwook [7 ]
Lee, Choonsik [1 ]
机构
[1] NCI, Div Canc Epidemiol & Genet, NIH, Rockville, MD 20850 USA
[2] Cleveland Clin, Dept Radiat Oncol, Cleveland, OH 44195 USA
[3] Univ Maryland, Sch Med, Dept Radiat Oncol, Baltimore, MD 21201 USA
[4] Boston Univ, Dept Radiat Oncol, Boston, MA 02215 USA
[5] East Carolina Univ, Dept Phys, Greenville, NC 27858 USA
[6] Univ Michigan, Div Radiat Oncol, Ann Arbor, MI 48109 USA
[7] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
基金
美国国家卫生研究院;
关键词
proton; pencil beam scanning; Monte Carlo; dose reconstruction; RADIOTHERAPY; SIMULATION; EXPOSURES; DOSIMETRY; CANCER; RISK;
D O I
10.1088/1361-6498/ab437d
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Significant efforts such as the Pediatric Proton/Photon Consortium Registry (PPCR) involving multiple proton therapy centers have been made to conduct collaborative studies evaluating outcomes following proton therapy. As a groundwork dosimetry effort for the late effect investigation, we developed a Monte Carlo (MC) model of proton pencil beam scanning (PBS) to estimate organ/tissue doses of pediatric patients at the Maryland Proton Treatment Center (MPTC), one of the proton centers involved in the PPCR. The MC beam modeling was performed using the TOPAS (TOol for PArticle Simulation) MC code and commissioned to match measurement data within 1% for range, and 0.3 mm for spot sizes. The established MC model was then tested by calculating organ/tissue doses for sample intracranial and craniospinal irradiations on whole-body pediatric computational human phantoms. The simulated dose distributions were compared with the treatment planning system dose distributions, showing the 3 mm/3% gamma index passing rates of 94%-99%, validating our simulations with the MC model. The calculated organ/tissue doses per prescribed doses for the craniospinal irradiations (1 mGy Gy(-1) to 1 Gy Gy(-1)) were generally much higher than those for the intracranial irradiations (2.1 mu Gy Gy(-1) to 0.1 Gy Gy(-1)), which is due to the larger field coverage of the craniospinal irradiations. The largest difference was observed at the adrenal dose, i.e. similar to 3000 times. In addition, the calculated organ/tissue doses were compared with those calculated with a simplified MC model, showing that the beam properties (i.e. spot size, spot divergence, mean energy, and energy spread) do not significantly influence dose calculations despite the limited irradiation cases. This implies that the use of the MC model commissioned to the MPTC measurement data might be dosimetrically acceptable for patient dose reconstructions at other proton centers particularly when their measurement data are unavailable. The developed MC model will be used to reconstruct organ/tissue doses for MPTC pediatric patients collected in the PPCR.
引用
收藏
页码:225 / 242
页数:18
相关论文
共 50 条
  • [21] Evaluation of monte carlo to support commissioning of the treatment planning system of new pencil beam scanning proton therapy facilities
    Botnariuc, D.
    Court, S.
    Lourenco, A.
    Gosling, A.
    Royle, G.
    Hussein, M.
    Rompokos, V
    Veiga, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (04)
  • [22] Determination of surface dose in pencil beam scanning proton therapy
    Kern, A.
    Baeumer, C.
    Kroeninger, K.
    Mertens, L.
    Timmermann, B.
    Walbersloh, J.
    Wulff, J.
    MEDICAL PHYSICS, 2020, 47 (05) : 2277 - 2288
  • [23] Secondary Neutron Dose From a Dynamic Collimation System During Intracranial Pencil Beam Scanning Proton Therapy: A Monte Carlo Investigation
    Smith, Blake R.
    Hyer, Daniel E.
    Hill, Patrick M.
    Culberson, Wesley S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 103 (01): : 241 - 250
  • [24] Commissioning dose computation model for proton source in pencil beam scanning therapy by convolution neural networks
    Liu, Yaoying
    Shang, Xuying
    Zhao, Wei
    Li, Nan
    Qu, Baolin
    Zou, Yue
    Le, Xiaoyun
    Zhang, Gaolong
    Xu, Shouping
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (15)
  • [25] Commissioning of pencil beam and Monte Carlo dose engines for non-isocentric treatments in scanned proton beam therapy
    Carlino, A.
    Boehlen, T.
    Vatnitsky, S.
    Grevillot, L.
    Osorio, J.
    Dreindl, R.
    Palmans, H.
    Stock, M.
    Kragl, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (17)
  • [26] Criteria of spot asymmetry in proton radiotherapy pencil beam scanning - a Monte Carlo study
    Klodowska, M.
    Gajewski, J.
    Skowronska, K.
    Olko, P.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S57 - S58
  • [27] Parametric characterization of penumbra reduction for aperture-collimated pencil beam scanning (PBS) proton therapy
    Maes, Dominic
    Regmi, Rajesh
    Taddei, Phillip
    Bloch, Charles
    Bowen, Steven
    Nevitt, Alexander
    Leuro, Erick
    Wong, Tony
    Rosenfeld, Anatoly
    Saini, Jatinder
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (03):
  • [28] A study of lateral fall-off (penumbra) optimisation for pencil beam scanning (PBS) proton therapy
    Winterhalter, C.
    Lomax, A.
    Oxley, D.
    Weber, D. C.
    Safai, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02)
  • [29] A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system
    Lin, Liyong
    Huang, Sheng
    Kang, Minglei
    Hiltunen, Petri
    Vanderstraeten, Reynald
    Lindberg, Jari
    Siljamaki, Sami
    Wareing, Todd
    Davis, Ian
    Barnett, Allen
    McGhee, John
    Simone, Charles B., II
    Solberg, Timothy D.
    McDonough, James E.
    Ainsley, Christopher
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 44 - 49
  • [30] Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy
    Dowdell, Stephen J.
    Clasie, Benjamin
    Depauw, Nicolas
    Metcalfe, Peter
    Rosenfeld, Anatoly B.
    Kooy, Hanne M.
    Flanz, Jacob B.
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (10) : 2829 - 2842