A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems

被引:11
|
作者
Yu, Amy S. [1 ]
Fowler, Tyler L. [1 ]
Dubrowski, Piotr [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
关键词
localization positioning system; nonradiographic and radiographic QA; quality assurance; radiotherapy; LINEAR-ACCELERATOR; REAL-TIME; RADIOSURGERY;
D O I
10.1002/mp.12950
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeVarious localization and positioning systems utilizing radiographic or nonradiographic methods have been developed to improve the accuracy of radiation treatment. Each quality assurance (QA) procedure requires its own phantom and is independent from each other, so the deviation between each system is unavailable. The purpose of this work is to develop and evaluate a single-integrated QA phantom for different localization and positioning systems. MethodsThe integrated phantom was designed in three-dimensional (3D) CAD software and 3D printed. The phantom was designed with laser alignment marks, a raised letter S on the anterior surface for optical surface monitoring system registration, a core for radiofrequency (RF) tracking system alignment, eight internal fiducials for image alignment, and an isocentric bearing for Winston-Lutz test. Tilt legs and rotational stage were designed for rotational verification of optical surface mapping system and RF tracking system, respectively. The phantom was scanned using a CT scanner and a QA plan was created. This prototype phantom was evaluated against established QA techniques. ResultsThe QA result between the proposed procedure and established QA technique are 1.12 0.31 and 1.14 +/- 0.31 mm, respectively, for RF tracking system and 0.18 +/- 0.06 and 0.18 +/- 0.05 mm for Winston-Lutz test. There is no significant difference for the QA results between the established QA and proposed procedure (P > 0.05, t test). The accuracy of rotational verification for surface mapping system and RF tracking system are less than 0.5 and 1 degrees compared the predefined value. The isocenter deviation of each location system is around l mm. ConclusionWe have designed and evaluated a novel-integrated phantom for radiographic and nonradiographic localization and positioning systems for radiotherapy. With this phantom, we will reduce the variation in measurements and simplify the QA procedures.
引用
收藏
页码:2857 / 2863
页数:7
相关论文
共 42 条
  • [41] 3-D Quality Assurance in CyberKnife Radiotherapy Using a Novel N-(3-methoxypropyl) Acrylamide Polymer Gel Dosimeter and Optical CT
    Awad, Samer, I
    Moftah, Belal
    Basfer, Ahmed
    Almousa, Akram A.
    Al Kafi, M. Abdullah
    Eyadeh, Molham M.
    Rabaeh, Khalid A.
    RADIATION PHYSICS AND CHEMISTRY, 2019, 161 : 34 - 41
  • [42] Long-term experience in quality assurance of on-rail computed tomography systems for image-guided radiotherapy using in-house multifunctional phantoms
    Masahide Saito
    Naoki Sano
    Hidekazu Suzuki
    Takafumi Komiyama
    Kan Marino
    Koji Ueda
    Hikaru Nemoto
    Hiroshi Onishi
    Radiological Physics and Technology, 2023, 16 : 292 - 298