Carbon fiber nanoelectrodes applied to microchip electrophoresis amperometric detection of neurotransmitter dopamine in rat pheochromocytorna (PC12) cells

被引:28
作者
Cheng, Han [1 ]
Huang, Wei-Hua [1 ]
Chen, Rong-Sheng [1 ]
Wang, Zong-Li [1 ]
Cheng, Jie-Ke [1 ]
机构
[1] Wuhan Univ, Dept Chem, Wuhan 430072, Peoples R China
关键词
carbon fiber nanoelectrodes; electrochemical detection; microfluidic chip electrophoresis; PC12; cells;
D O I
10.1002/elps.200600603
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Microelectrodes have been adopted in electrochemical detection for CE or microchip CE in recent years. In this paper, the use of nanoelectrodes (with tip diameter of 100-300 nm) as the electrochemical detector in microchip CE is firstly reported. The experimental results indicated that both the sensitivity and resolution of microchip CE with the carbon fiber nanoelectrode (CFNE) amperometric detection have been improved markedly comparing with the traditional microelectrodes. The detection limit of dopamine (S/N = 3) is 5.9 X 10(-8) M, which is one or two orders of magnitude lower than that reported so far, and the resolution of dopamine (DA) and isoprenaline (IP) has also improved from 0.6 (using 7 mu m carbon fiber microelectrodes, CFME) to 1.0. We assembled a novel and easily operated microchip CE system with end-column amperometric detection, which allows the convenient and fast replacement of the passivated electrodes. Under the optimized condition, the RSDs of peak height and migration time are 1.47 and 0.31%, respectively (n = 40), indicating that the system displays excellent reproducibility. The nanoelectrode-based microchip CE system has been successfully applied to the determination of DA in cultured rat pheochromocytorna (PC12) cells, and the average content of DA in an individual PC12 cell is 0.54 +/- 0.07 fmol, which is in good agreement with that reported in the literature.
引用
收藏
页码:1579 / 1586
页数:8
相关论文
共 55 条
[1]   Microfluidic chip for low-flow push-pull perfusion sampling in vivo with on-line analysis of amino acids [J].
Cellar, NA ;
Burns, ST ;
Meiners, JC ;
Chen, H ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 2005, 77 (21) :7067-7073
[2]  
CHEN JK, 2005, ANAL SINGLE CELLS, P110
[3]  
DOUGLASS SD, 1995, ANAL CHEM, V67, P58
[4]   Separation and detection of nitrophenols at capillary electrophoresis microchips with amperometric detection [J].
Fischer, J ;
Barek, J ;
Wang, J .
ELECTROANALYSIS, 2006, 18 (02) :195-199
[5]   Integration of single cell injection, cell lysis, separation and detection of intracellular constituents on a microfluidic chip [J].
Gao, J ;
Yin, XF ;
Fang, ZL .
LAB ON A CHIP, 2004, 4 (01) :47-52
[6]  
Gawron AJ, 2001, ELECTROPHORESIS, V22, P242, DOI 10.1002/1522-2683(200101)22:2<242::AID-ELPS242>3.0.CO
[7]  
2-W
[8]   POSTCOLUMN DERIVATIZATION IN NARROW-BORE CAPILLARIES FOR THE ANALYSIS OF AMINO-ACIDS AND PROTEINS BY CAPILLARY ELECTROPHORESIS WITH FLUORESCENCE DETECTION [J].
GILMAN, SD ;
PIETRON, JJ ;
EWING, AG .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1994, 6 (04) :373-384
[9]   MICROMACHINING A MINIATURIZED CAPILLARY ELECTROPHORESIS-BASED CHEMICAL-ANALYSIS SYSTEM ON A CHIP [J].
HARRISON, DJ ;
FLURI, K ;
SEILER, K ;
FAN, ZH ;
EFFENHAUSER, CS ;
MANZ, A .
SCIENCE, 1993, 261 (5123) :895-897
[10]   A method for the fabrication of low-noise carbon fiber nanoelectrodes [J].
Huang, WH ;
Pang, DW ;
Tong, H ;
Wang, ZL ;
Cheng, JK .
ANALYTICAL CHEMISTRY, 2001, 73 (05) :1048-1052