A new wavelength selection criterion for two-color pyrometer interfered with participating media

被引:11
作者
Huang, Yunwei [1 ]
Long, Mujun [1 ]
Chen, Dengfu [1 ]
Duan, Huamei [1 ]
Tan, Kai [1 ]
Gui, Lintao [1 ]
Liu, Tao [1 ]
机构
[1] Chongqing Univ, State Key Lab Coal Mine Disaster Dynam & Control, Coll Mat Sci & Engn, Chongqing 400044, Peoples R China
关键词
Two-color pyrometer; Wavelength selection criterion; Participating media; Radiation transfer; TEMPERATURE-MEASUREMENT; RADIATIVE-TRANSFER; BAND MODEL; PARAMETERS; H2O; CO2;
D O I
10.1016/j.infrared.2018.07.022
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Two-color pyrometry is well established and widely applied to surface temperature measurements. However, its accuracy degrades significantly when the transmission path between the object surface and the pyrometer is full of high concentration participating media, which attenuates the surface radiation signal to the pyrometer. To improve measurement accuracy, a theoretical optimization wavelength selection criterion for two-color pyrometry working with the interference of a participating medium is established on the basis of the Plank's law and Wien's approximation. The closer to zero the criterion value is, the smaller the measuring error is. Therefore, the optimum choice of operating wavelengths combination for two-color pyrometer is obtained through inter-comparison of the criterion values. Evaluation of the practical applicability of the new criterion is performed by employing an analytical radiation thermometry model considering radiative attenuation and augmentation in participating media. Theoretical estimations confirm that the new criterion has good applicability and scientific quality considering two typical media ordinarily encountered in engineering applications: water vapor and water mist. This selection criterion not only provides solid reference information for the proper selection of wavelengths combination, but also helps one to better understand the fundamentals of measurement errors for two-color pyrometer interfered with participating media. The analyses in the paper provide the necessary theoretical supports for the design and application of a two-color pyrometer.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 31 条
  • [1] Multi-spectral pyrometry-a review
    Araujo, Antonio
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (08)
  • [2] Monte Carlo simulations of ambient temperature uncertainty determined by dual-band pyrometry
    Araujo, Antonio
    Martins, Nelson
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2015, 26 (08)
  • [3] Bohren C F, 1998, OPT LASER TECHNOL, V31, P328
  • [4] BLACKBODY RADIATION FUNCTIONS
    CHANG, SL
    RHEE, KT
    [J]. INTERNATIONAL COMMUNICATIONS IN HEAT AND MASS TRANSFER, 1984, 11 (05) : 451 - 455
  • [5] MULTI-WAVELENGTH PYROMETRY
    COATES, PB
    [J]. METROLOGIA, 1981, 17 (03) : 103 - 109
  • [6] Dewitt D.P.N., 1988, Theory and Practice of Radiation Thermometry
  • [7] Eiveland W. A., 1991, J THERMOPHYS HEAT TR, V1, P9
  • [8] The analysis of optimization criteria for multi-band pyrometry
    Fu, TR
    Cheng, XF
    Fan, XL
    Ding, JL
    [J]. METROLOGIA, 2004, 41 (04) : 305 - 313
  • [9] Huang Y., 2016, EPD C 2016, V2016, P143
  • [10] Kim T.K., 1991, HUM RELAT, V56, P587