Accuracy of cup placement in total hip arthroplasty by means of a mechanical positioning device: a comprehensive cadaveric 3d analysis of 16 specimens

被引:3
作者
Kievit, Arthur J. [1 ]
Dobbe, Johannes G. G. [2 ]
Mallee, Wouter H. [1 ]
Blankevoort, Leendert [1 ]
Streekstra, Geert J. [2 ]
Schafroth, Matthias U. [1 ]
机构
[1] Univ Amsterdam, Orthopaed Res Ctr Amsterdam, Amsterdam Univ Med Ctr, Dept Orthopaed, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Amsterdam, Dept Biomed Engn & Phys, Amsterdam Univ Med Ctr, Amsterdam, Netherlands
关键词
3D analysis; accuracy study; anterior pelvic plane; anteversion; inclination; Lewinnek's safe zone; total hip arthroplasty; ANTERIOR PELVIC PLANE; ACETABULAR COMPONENT; NAVIGATION; ORIENTATION; DISLOCATION; GUIDE; RISK;
D O I
10.1177/1120700019874822
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: We tested whether a mechanical device (such as Hipsecure) to pinpoint the anterior pelvic plane (APP) as a guide can improve acetabular cup placement. To assess accuracy we asked: (1) is the APP an effective guide to position acetabular cup placement within acceptable degrees of divergence from the optimal 40 degrees inclination and 15 degrees anteversion; (2) could a mechanical device increase the number of acetabular cup placements within Lewinnek's safe zone (i.e. inclination 30 degrees to 50 degrees; anteversion 5 degrees to 25 degrees)? Methods: 16 cadaveric specimens were used to assess the 3D surgical success of using a mechanical device APP to guide acetabular cup placement along the APP. We used the Hipsecure mechanical device to implant acetabular cups at 40 degrees inclination and 15 degrees anteversion. Subequently, all cadaveric specimens with implants were scanned with a CT and 3D models were created of the pelvis and acetabular cups to assess the outcome in terms of Lewinnek's safe zones. Results: The mean inclination of the 16 implants was 40.6 degrees (95% CI, 37.7-43.4) and the mean anteversion angle was 13.4 degrees (95% CI, 10.7-16.1). All 16 cup placements were within Lewinnek's safe zone for inclination (between 30 degrees and 50 degrees) and all but 2 were within Lewinnek's safe zone for anteversion (between 5 degrees and 25 degrees). Conclusion: In cadaveric specimens, the use of a mechanical device and the APP as a guide for acetabular cup placement resulted in good positioning with respect to both of Lewinnek's safe zones.
引用
收藏
页码:58 / 65
页数:8
相关论文
共 31 条
[1]   The reliability of the anterior pelvic plane for computer navigated acetabular component placement during total hip arthroplasty: Prospective study with the EOS imaging system [J].
Barbier, O. ;
Skalli, W. ;
Mainard, L. ;
Mainard, D. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (06) :S287-S291
[2]   Accuracy of Acetabular Component Position in Hip Arthroplasty [J].
Barrack, Robert L. ;
Krempec, Jeffrey A. ;
Clohisy, John C. ;
McDonald, Douglas J. ;
Ricci, William M. ;
Ruh, Erin L. ;
Nunley, Ryan M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (19) :1760-1768
[3]  
Biedermann R, 2005, J BONE JOINT SURG BR, V87B, P762, DOI 10.1302/0301-620X.87B6
[4]   Image-free cusp navigation inaccuracy: A two-study approach [J].
Blendea, Sorin ;
Troccaz, Jocelyne ;
Ravey, Jean-Noel ;
Merloz, Philippe .
COMPUTER AIDED SURGERY, 2007, 12 (03) :176-180
[5]   Poor accuracy of freehand cup positioning during total hip arthroplasty [J].
Bosker, B. H. ;
Verheyen, C. C. P. M. ;
Horstmann, W. G. ;
Tulp, N. J. A. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2007, 127 (05) :375-379
[6]   The John Charnley Award Risk Factors for Cup Malpositioning Quality Improvement Through a Joint Registry at a Tertiary Hospital [J].
Callanan, Mark C. ;
Jarrett, Bryan ;
Bragdon, Charles R. ;
Zurakowski, David ;
Rubash, Harry E. ;
Freiberg, Andrew A. ;
Malchau, Henrik .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) :319-329
[7]  
Dandachli W, 2006, Comput Aided Surg, V11, P322, DOI 10.1080/10929080601090706
[8]   Comparison of a mechanical acetabular alignment guide with computer placement of the socket [J].
DiGioia, AM ;
Jaramaz, B ;
Plakseychuk, AY ;
Moody, JE ;
Nikou, C ;
LaBarca, RS ;
Levison, TJ ;
Picard, F .
JOURNAL OF ARTHROPLASTY, 2002, 17 (03) :359-364
[9]   Functional pelvic orientation measured from lateral standing and sitting radiographs [J].
DiGioia, Anthony M., III ;
Hafez, Mahmoud A. ;
Jaramaz, Branislav ;
Levison, Timothy J. ;
Moody, James E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (453) :272-276
[10]   Computer-Assisted Planning and Navigation for Corrective Distal Radius Osteotomy, Based on Pre- and Intraoperative Imaging [J].
Dobbe, J. G. G. ;
Strackee, S. D. ;
Schreurs, A. W. ;
Jonges, R. ;
Carelsen, B. ;
Vroemen, J. C. ;
Grimbergen, C. A. ;
Streekstra, G. J. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (01) :182-190