Analysis of total arsenic, total selenium and total chromium in urine by inductively coupled plasma-dynamic reaction cell-mass spectrometry

被引:36
作者
McShane, William J.
Pappas, R. Steven
Paschal, Dan
机构
[1] Ctr Dis Control & Prevent, Div Sci Lab, Battelle, Atlanta, GA 30341 USA
[2] Ctr Dis Control & Prevent, Div Sci Lab, Atlanta, GA 30341 USA
关键词
D O I
10.1039/b613884e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An accurate and simple method for the determination of total arsenic, total selenium and total chromium in urine has been developed using inductively coupled argon plasma- dynamic reaction cell- mass spectrometry ( ICP- DRC- MS). Determination was by external calibration, using matrixmatched standards and high purity calibration materials from a commercial vendor. Aliquots of each urine specimen were diluted (1 + 9) with 2% (v/v) nitric acid containing rhodium and iridium as an internal standard. Argon cell gas was used to remove the isobaric interferences that normally adversely a. ect As-75, Se-78 and Cr-52 analysis by ICP- MS and require the use of a dynamic reaction cell (DRC). The principle isobars that interfere with these measurements are [(ArCl)-Ar-40-Cl-35](+), [(ArAr)-Ar-40-Ar-38](+) and [(ArC)-Ar-40-C-12](+) for(75) As, Se-78 and Cr-52, respectively. The counts at m/z 75 (arsenic); 78 (selenium); 52 (chromium), 103 (rhodium), and 193 ( iridium) were measured. The ratios of the counts at m/z 75 or 78 to those at m/z 193, and m/z 52 to those at m/z 103 were calculated. These ratios were compared with those from urine- based calibration standards to calculate the arsenic, selenium, and chromium concentrations in unknown specimens. The concentrations of arsenic, selenium, and chromium were calculated as mu g L-1 in the sample. No creatinine corrections were made, although these could easily be made with data for urine creatinine for each specimen. The proposed method uses the same diluted urine solutions prepared for conventional ICP- MS toxic metal biomonitoring and chemical terrorism screening analysis with the same internal standards; and uses argon as the DRC cell gas. The proposed method provides the basis for an accurate method for determining total arsenic, total selenium, and total chromium in unexposed subjects, and in persons considered to be exposed to those elements through chemical terrorism, environmental, nutritional or other pathways. Approximately 100 specimens, including blanks, calibration standards and quality-control materials, can be processed in an 8 h day.
引用
收藏
页码:630 / 635
页数:6
相关论文
共 21 条
[1]  
[Anonymous], 2000, MMWR Recomm Rep, V49, P1
[2]   Reaction chemistry and collisional processes in multipole devices for resolving isobaric interferences in ICP-MS [J].
Bandura, DR ;
Baranov, VI ;
Tanner, SD .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (05) :454-470
[3]  
*BOARD CHEM SCI, 2006, TERR CHEM INFR PROT, P45
[4]  
BURTIS CA, 2001, TIETZ FUNDAMENTALS C, P500
[5]   Determination of chromium in water and urine by reaction cell inductively coupled plasma mass spectrometry [J].
Chang, YL ;
Jiang, SJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (12) :1434-1438
[6]   Optimisation of ICP-dynamic reaction cell-MS as specific detector for the speciation analysis of vanadium at therapeutic levels in serum [J].
Chéry, CC ;
De Cremer, K ;
Cornelis, R ;
Vanhaecke, F ;
Moens, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (09) :1113-1118
[7]  
COTTON FA, 1999, ADV INORG CHEM, P386
[8]  
COTTON FA, 1999, ADV INORG CHEM, P756
[9]  
DU Z, 2002, TALANTA, V58, P121
[10]  
Francesconi KA, 1998, CROAT CHEM ACTA, V71, P343