Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

被引:39
|
作者
Michalak, Gregory [1 ]
Grimes, Joshua [1 ]
Fletcher, Joel [1 ]
Halaweish, Ahmed [2 ]
Yu, Lifeng [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA 19355 USA
关键词
dual-energy CT; monoenergetic; stability; EXPERIMENTAL-VERIFICATION; DIAGNOSTIC PERFORMANCE; MULTIDETECTOR CT; NOISE-REDUCTION; QUALITY; STONES; SINGLE; SUBTRACTION; SELECTION; ACCURACY;
D O I
10.1118/1.4939128
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside +/- 10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors' report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:513 / 517
页数:5
相关论文
共 50 条
  • [41] Comparing fully automated AI body composition biomarkers at differing virtual monoenergetic levels using dual-energy CT
    Toia, Giuseppe V.
    Garret, John W.
    Rose, Sean D.
    Szczykutowicz, Timothy P.
    Pickhardt, Perry J.
    ABDOMINAL RADIOLOGY, 2024,
  • [42] Virtual Monochromatic Imaging in Dual-source and Dual-energy CT for Visualization of Acute Ischemic Stroke
    Hara, Hidetake
    Muraishi, Hiroshi
    Matsuzawa, Hiroki
    Inoue, Toshiyuki
    Nakajima, Yasuo
    Satoh, Hitoshi
    Abe, Shinji
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 103 - 107
  • [43] The optimal energy level of virtual monochromatic imaging in dual-energy CT arthrography of the wrist
    Chung, Bo Mi
    Lee, Guen Young
    Kim, Sujin
    BRITISH JOURNAL OF RADIOLOGY, 2024, 98 (1166): : 305 - 311
  • [44] Dual-Energy CT in Head and Neck Imaging
    Roele E.D.
    Timmer V.C.M.L.
    Vaassen L.A.A.
    van Kroonenburgh A.M.J.L.
    Postma A.A.
    Current Radiology Reports, 5 (5)
  • [45] Clinical utility of virtual noncalcium dual-energy CT in imaging of the pelvis and hip
    Baffour, Francis I.
    Glazebrook, Katrina N.
    Morris, Jonathan M.
    Michalak, Gregory J.
    Fletcher, Joel G.
    Leng, Shuai
    McCollough, Cynthia H.
    SKELETAL RADIOLOGY, 2019, 48 (12) : 1833 - 1842
  • [46] Clinical photon-counting CT increases CT number precision and reduces patient size dependence compared to single- and dual-energy CT
    Flores, Jessica D.
    Poludniowski, Gavin
    Szum, Adrian
    Walther, Georg
    Lundberg, Johan
    Nowik, Patrik
    Granberg, Tobias
    BRITISH JOURNAL OF RADIOLOGY, 2025,
  • [47] Dual-Energy CT: Applications in Abdominal Imaging
    Bauer R.W.
    Fischer S.
    Current Radiology Reports, 3 (4)
  • [48] Numerical Simulation for Dual-energy CT Imaging
    Wang, Lixin
    Sun, Fengrong
    Babyn, Paul
    Zhong, Hai
    Zhou, Zhenzhen
    Huang, Hao
    Qin, Jiaojiao
    PROCEEDINGS OF THE 2015 2ND INTERNATIONAL CONFERENCE ON ELECTRICAL, COMPUTER ENGINEERING AND ELECTRONICS (ICECEE 2015), 2015, 24 : 316 - 320
  • [49] Clinical utility of virtual noncalcium dual-energy CT in imaging of the pelvis and hip
    Francis I. Baffour
    Katrina N. Glazebrook
    Jonathan M. Morris
    Gregory J. Michalak
    Joel G. Fletcher
    Shuai Leng
    Cynthia H. McCollough
    Skeletal Radiology, 2019, 48 : 1833 - 1842
  • [50] Dual-Energy CT of Pediatric Abdominal Oncology Imaging: Private Tour of New Applications of CT Technology
    Kamps, Shawn E.
    Otjen, Jeffrey P.
    Stanescu, A. Luana
    Mileto, Achille
    Lee, Edward Y.
    Phillips, Grace S.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (05) : 967 - 975