Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

被引:39
|
作者
Michalak, Gregory [1 ]
Grimes, Joshua [1 ]
Fletcher, Joel [1 ]
Halaweish, Ahmed [2 ]
Yu, Lifeng [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA 19355 USA
关键词
dual-energy CT; monoenergetic; stability; EXPERIMENTAL-VERIFICATION; DIAGNOSTIC PERFORMANCE; MULTIDETECTOR CT; NOISE-REDUCTION; QUALITY; STONES; SINGLE; SUBTRACTION; SELECTION; ACCURACY;
D O I
10.1118/1.4939128
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside +/- 10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors' report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:513 / 517
页数:5
相关论文
共 50 条
  • [31] Dual-Energy Computed Tomography Virtual Monoenergetic Imaging of Lung Cancer: Assessment of Optimal Energy Levels
    Kaup, Moritz
    Scholtz, Jan-Erik
    Engler, Alexander
    Albrecht, Moritz H.
    Bauer, Ralf W.
    Kerl, J. Matthias
    Beeres, Martin
    Lehnert, Thomas
    Vogl, Thomas J.
    Wichmann, Julian L.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2016, 40 (01) : 80 - 85
  • [32] Noise-optimised virtual monoenergetic imaging of dual-energy CT: effect on metal artefact reduction in patients with lumbar internal fixation
    Zeng, Yanwei
    Deng, Kai
    Yang, Haitao
    Tan, Yi
    Liu, Jun
    Geng, Daoying
    Zhang, Jun
    EUROPEAN SPINE JOURNAL, 2019, 28 (08) : 1783 - 1792
  • [33] Dual-energy CT and coronary imaging
    Kay, Fernando Uliana
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2020, 10 (04) : 1090 - 1107
  • [34] Imaging the Spine with Dual-Energy CT
    Murray N.
    Le M.
    Ebrahimzadeh O.
    Alharthy A.
    Mohammed M.F.
    Ouellette H.A.
    Khosa F.
    Current Radiology Reports, 5 (9)
  • [35] Virtual Monoenergetic Images of Dual-Energy CT-Impact on Repeatability, Reproducibility, and Classification in Radiomics
    Euler, Andre
    Laqua, Fabian Christopher
    Cester, Davide
    Lohaus, Niklas
    Sartoretti, Thomas
    dos Santos, Daniel Pinto
    Alkadhi, Hatem
    Baessler, Bettina
    CANCERS, 2021, 13 (18)
  • [36] Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Yoshida, Ryoichi
    Katayama, Takuto
    Arimoto, Makoto
    Kataoka, Jun
    Nagata, Hiroji
    Kobayashi, Satoshi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, : 143 - 153
  • [37] Endoleaks after endovascular aortic aneurysm repair: Improved detection with noise-optimized virtual monoenergetic dual-energy CT
    Martin, Simon S.
    Wichmann, Julian L.
    Weyer, Hendrik
    Scholtz, Jan-Erik
    Leithner, Doris
    Spandorfer, Adam
    Bodelle, Boris
    Jacobi, Volkmar
    Vogl, Thomas J.
    Albrecht, Moritz H.
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 94 : 125 - 132
  • [38] Dual-Energy CT: New Horizon in Medical Imaging
    Goo, Hyun Woo
    Goo, Jin Mo
    KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (04) : 555 - 569
  • [39] Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging
    Martin Beeres
    Jesko Trommer
    Claudia Frellesen
    Nour-Eldin A. Nour-Eldin
    Jan E. Scholtz
    Eva Herrmann
    Thomas J. Vogl
    Julian L. Wichmann
    The International Journal of Cardiovascular Imaging, 2016, 32 : 137 - 144
  • [40] Dual-Energy CT for Pediatric Thoracic Imaging: A Review
    Rapp, Jordan B.
    Biko, David M.
    Siegel, Marilyn J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 221 (04) : 526 - 538