Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

被引:39
|
作者
Michalak, Gregory [1 ]
Grimes, Joshua [1 ]
Fletcher, Joel [1 ]
Halaweish, Ahmed [2 ]
Yu, Lifeng [1 ]
Leng, Shuai [1 ]
McCollough, Cynthia [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA 19355 USA
关键词
dual-energy CT; monoenergetic; stability; EXPERIMENTAL-VERIFICATION; DIAGNOSTIC PERFORMANCE; MULTIDETECTOR CT; NOISE-REDUCTION; QUALITY; STONES; SINGLE; SUBTRACTION; SELECTION; ACCURACY;
D O I
10.1118/1.4939128
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside +/- 10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors' report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:513 / 517
页数:5
相关论文
共 50 条
  • [1] Virtual Monoenergetic Imaging of Lower Extremities Using Dual-Energy CT Angiography in Patients with Diabetes Mellitus
    Bucolo, Giuseppe Mauro
    D'Angelo, Tommaso
    Yel, Ibrahim
    Koch, Vitali
    Gruenewald, Leon D.
    Othman, Ahmed E.
    Alizadeh, Leona Soraja
    Overhoff, Daniel P.
    Waldeck, Stephan
    Martin, Simon S.
    Mazziotti, Silvio
    Ascenti, Giorgio
    Blandino, Alfredo
    Vogl, Thomas J.
    Booz, Christian
    DIAGNOSTICS, 2023, 13 (10)
  • [2] Technical Note: Relation between dual-energy subtraction of CT images for electron density calibration and virtual monochromatic imaging
    Saito, Masatoshi
    MEDICAL PHYSICS, 2015, 42 (07) : 4088 - 4093
  • [3] Virtual monoenergetic dual-layer dual-energy CT images in colorectal cancer: CT diagnosis could be improved?
    Arico, Francesco Marcello
    Trimarchi, Renato
    Portaluri, Antonio
    Barilla, Claudia
    Migliaccio, Nicola
    Bucolo, Giuseppe Mauro
    Cicero, Giuseppe
    Sofia, Carmelo
    Booz, Christian
    Vogl, Thomas J. J.
    Marino, Maria Adele
    Ascenti, Velio
    D'Angelo, Tommaso
    Mazziotti, Silvio
    Ascenti, Giorgio
    RADIOLOGIA MEDICA, 2023, 128 (08): : 891 - 899
  • [4] Dual-energy CT in patients with colorectal cancer: Improved assessment of hypoattenuating liver metastases using noise-optimized virtual monoenergetic imaging
    Lenga, Lukas
    Czwikla, Rouben
    Wichmann, Julian L.
    Leithner, Doris
    Albrecht, Moritz H.
    Booz, Christian
    Arendt, Christophe T.
    Yel, Ibrahim
    D'Angelo, Tommaso
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 106 : 184 - 191
  • [5] Dual-Energy CT in Oncologic Imaging
    Foti, Giovanni
    Ascenti, Giorgio
    Agostini, Andrea
    Longo, Chiara
    Lombardo, Fabio
    Inno, Alessandro
    Modena, Alessandra
    Gori, Stefania
    TOMOGRAPHY, 2024, 10 (03) : 299 - 319
  • [6] Maximizing Iodine Contrast-to-Noise Ratios in Abdominal CT Imaging through Use of Energy Domain Noise Reduction and Virtual Monoenergetic Dual-Energy CT
    Leng, Shuai
    Yu, Lifeng
    Fletcher, Joel G.
    McCollough, Cynthia H.
    RADIOLOGY, 2015, 276 (02) : 562 - 570
  • [7] Improved Detection of Endoleaks in Virtual Monoenergetic Images in Dual-Energy CT Angiography Following EVAR
    Kazimierczak, Wojciech
    Kazimierczak, Natalia
    Lemanowicz, Adam
    Nowak, Ewa
    Migdalski, Arkadiusz
    Jawien, Arkadiusz
    Jankowski, Tomasz
    Serafin, Zbigniew
    ACADEMIC RADIOLOGY, 2023, 30 (12) : 2813 - 2824
  • [8] Virtual monoenergetic dual-energy CT for evaluation of hepatic and splenic lacerations
    Sun, Ellen X.
    Wortman, Jeremy R.
    Uyeda, Jennifer W.
    Lacson, Roger
    Sodickson, Aaron D.
    EMERGENCY RADIOLOGY, 2019, 26 (04) : 419 - 425
  • [9] Dual-Energy CT for the Detection of Portal Vein Thrombosis: Improved Diagnostic Performance Using Virtual Monoenergetic Reconstructions
    Martin, Simon S.
    Kolaneci, Jetlir
    Czwikla, Rouben
    Booz, Christian
    Gruenewald, Leon D.
    Albrecht, Moritz H.
    Thompson, Zachary M.
    Lenga, Lukas
    Yel, Ibrahim
    Vogl, Thomas J.
    Wichmann, Julian L.
    Koch, Vitali
    DIAGNOSTICS, 2022, 12 (07)
  • [10] Virtual monoenergetic dual-energy CT for evaluation of hepatic and splenic lacerations
    Ellen X. Sun
    Jeremy R. Wortman
    Jennifer W. Uyeda
    Roger Lacson
    Aaron D. Sodickson
    Emergency Radiology, 2019, 26 : 419 - 425