Preconcentration of Pb with Aminosilanized Fe3O4 Nanopowders in Environmental Water Followed by Electrothermal Atomic Absorption Spectrometric Determination

被引:3
作者
Kusutaki, Tomoharu [1 ]
Furukawa, Mai [1 ]
Tateishi, Ikki [2 ]
Katsumata, Hideyuki [1 ]
Kaneco, Satoshi [1 ]
机构
[1] Mie Univ, Grad Sch Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
[2] Mie Univ, Global Environm Ctr Educ & Res, Tsu, Mie 5148507, Japan
关键词
aminosilanized Fe3O4; ETAAS; Pb; preconcentration; environmental sample;
D O I
10.3390/chemengineering3030074
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A new preconcentration method to determine lead in environmental waters using the aminosilanized magnetite Fe3O4 powder sorbent has been developed. The preconcentration method was combined with electrothermal atomization atomic absorption spectrometry (ETAAS) and a graphite atomizer. Trace amount of sorbent (3 mg) could be applied into the preconcentration of Pb. According the preconcentration, the detection limits were 14 and 19 pg.mL(-1) with bare and aminosilanized Fe3O4, respectively. The effect of interferent elements such as Al, Ca, Co, Fe, K, Mg, Na, Ni, and Zn (1000 ng.mL(-1) versus Pb 1 ng.mL(-1)) on the preconcentration of Pb using bare magnetite was evaluated, and some elements (Al, Ni, and Zn) were found to interfere with the Pb preconcentration. The aminosilanized Fe3O4 sorbent was found to be effective in eliminating the severe interferences. The enrichment factors were 34 for the preconcentration with aminosilanized Fe3O4. The recovery of spiked Pb in the case of the sorbent with aminosilanized Fe3O4 was in the range of 75 to 110%. From the analytical data, the preconcentration technique was found to be useful for the determination of trace lead in environmental waters.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 29 条
  • [1] Electrothermal atomic absorption spectrometric determination of lead in calcium drug samples by direct atomization technique
    Ahsan, S
    Kaneco, S
    Ohta, K
    Mizuno, T
    Suzuki, T
    Miyada, M
    Taniguchi, Y
    [J]. ANALYTICA CHIMICA ACTA, 1998, 362 (2-3) : 279 - 284
  • [2] Solid phase extraction of heavy metal ions from agricultural samples with the aid of a novel functionalized magnetic metal-organic framework
    Babazadeh, Mirzaagha
    Hosseinzadeh-Khanmiri, Rahim
    Abolhasani, Jafar
    Ghorbani-Kalhor, Ebrahim
    Hassanpour, Akbar
    [J]. RSC ADVANCES, 2015, 5 (26) : 19884 - 19892
  • [3] Bahadir Z, 2014, J IND ENG CHEM, V20, P1030
  • [4] Hollow-fiber liquid-liquid-solid micro-extraction of lead in soft drinks and determination by graphite furnace atomic absorption spectrometry
    Carletto, Jeferson S.
    Carasek, Eduardo
    Welz, Bernhard
    [J]. TALANTA, 2011, 84 (03) : 989 - 994
  • [5] Extraction of trace amounts of mercury with sodium dodecyle sulphate-coated magnetite nanoparticles and its determination by flow injection inductively coupled plasma-optical emission spectrometry
    Faraji, Mohammad
    Yamini, Yadollah
    Rezaee, Mohammad
    [J]. TALANTA, 2010, 81 (03) : 831 - 836
  • [6] Simultaneous and Direct Determination of As, Bi, Pb, Sb, and Se and Co, Cr, Cu, Fe, and Mn in Milk by Electrothermal Atomic Absorption Spectrometry
    Freschi, Gian Paulo G.
    Fortunato, Felipe M.
    Freschi, Carolina D.
    Gomes Neto, Jose Anchieta
    [J]. FOOD ANALYTICAL METHODS, 2012, 5 (04) : 861 - 866
  • [7] A metal-organic framework nanocomposite made from functionalized magnetite nanoparticles and HKUST-1 (MOF-199) for preconcentration of Cd(II), Pb(II), and Ni(II)
    Ghorbani-Kalhor, Ebrahim
    [J]. MICROCHIMICA ACTA, 2016, 183 (09) : 2639 - 2647
  • [8] Silica-coated magnetic nanoparticles modified with γ-mercaptopropyltrimethoxysilane for fast and selective solid phase extraction of trace amounts of Cd, CuHg, and Pb in environmental and biological samples prior to their determination by inductively coupled plasma mass spectrometry
    Huang, Chaozhang
    Hu, Bin
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2008, 63 (03) : 437 - 444
  • [9] Dispersive liquid-liquid microextraction combined with flow injection inductively coupled plasma mass spectrometry for simultaneous determination of cadmium, lead and bismuth in water samples
    Jia, Xiaoyu
    Han, Yi
    Liu, Xinli
    Duan, Taicheng
    Chen, Hangting
    [J]. MICROCHIMICA ACTA, 2010, 171 (1-2) : 49 - 56
  • [10] Facile synthesis of smartaminosilane modified-SnO2/porous silica nanocomposite for high efficiency removal of lead ions and bacterial inactivation
    Kheshtzar, Iman
    Ghorbani, Mohsen
    Gatabi, Maliheh Pashai
    Lashkenari, Mohammad Soleimani
    [J]. JOURNAL OF HAZARDOUS MATERIALS, 2018, 359 : 19 - 30