Fatigue testing of a proximal femoral hip component

被引:3
|
作者
Macdonald, W
Carlsson, LV
Gathercole, N
Jacobsson, CM
机构
[1] Univ Bath, Dept Engn & Appl Sci, Mat Res Ctr, SMS Consultants, Bath BA2 7AY, Avon, England
[2] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Orthopaed, Gothenburg, Sweden
[3] Univ Gothenburg, Inst Surg Sci, Dept Biomat Res, Gothenburg, Sweden
关键词
hip replacement; fatigue testing; femoral component; GFRE bushings; GOT implant;
D O I
10.1243/09544110360579358
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Gothenburg Osseointegrated Titanium (GOT) implant is a novel total hip replacement including a metaphyseal loading proximal femoral component fixed in the retained femoral neck. Endurance testing was performed under conditions analogous to ISO 7206-4: 1989. The cement-free implant is not fixed distally within the intramedullary canal, so distal embedding (as specified in the standard) would have been unrealistic. Instead glass-fibre-reinforced epoxy (GFRE) bushings were used to model reduced bone support mid-length at the medial cortex and distally at the lateral cortex. Such support simulated proximal bone loss, realistically reproducing the effect of osteolysis or fixation failure. Under such conditions the component survived unbroken for 10 million cycles at 3.0 kN peak load.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [21] Cement mantle fatigue failure in total hip replacement: Experimental and computational testing
    Jeffers, Jonathan R. T.
    Browne, Martin
    Lennon, Alexander B.
    Prendergast, Patrick J.
    Taylor, Mark
    JOURNAL OF BIOMECHANICS, 2007, 40 (07) : 1525 - 1533
  • [22] Comparison of proximal femoral nail antirotation internal fixation and artificial hip replacement for elderly patients with intertrochanteric fractures
    Jiang, Wenhui
    Ye, Keyong
    Lv, Guomin
    Cheng, Zhengyue
    Lin, Lixiang
    AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2024, 16 (08): : 4154 - 4162
  • [23] Recurrent total hip arthroplasty dislocation in combination with massive osteolysis: an outline of management using a proximal femoral replacement
    Kundra, R. K.
    Bhattacharya, S.
    Lawrence, T.
    HIP INTERNATIONAL, 2007, 17 (04) : 237 - 240
  • [24] Premature failure of a femoral hip replacement
    Oravcova, Monika
    Palcek, Peter
    Chalupova, Maria
    MATERIALS TODAY-PROCEEDINGS, 2016, 3 (04) : 1114 - 1117
  • [25] Influence of femoral anteversion on proximal femoral loading: measurement and simulation in four patients
    Heller, MO
    Bergmann, G
    Deuretzbacher, G
    Claes, L
    Haas, NP
    Duda, GN
    CLINICAL BIOMECHANICS, 2001, 16 (08) : 644 - 649
  • [26] Accuracy of femoral templating in reproducing anatomical femoral offset in total hip replacement
    Davies, H.
    Foote, J.
    Spencer, R. F.
    HIP INTERNATIONAL, 2007, 17 (03) : 155 - 159
  • [27] Osteosynthesis and hip replacement in proximal femur fractures
    Zarattini, Guido
    AGING CLINICAL AND EXPERIMENTAL RESEARCH, 2013, 25 : S55 - S58
  • [28] Osteosynthesis and hip replacement in proximal femur fractures
    Guido Zarattini
    Aging Clinical and Experimental Research, 2013, 25 : 55 - 58
  • [29] Precoated femoral component in primary hybrid total hip arthroplasty - Results at a mean 10-year follow-up
    Lachiewicz, PF
    Messick, P
    JOURNAL OF ARTHROPLASTY, 2003, 18 (01): : 1 - 5
  • [30] Total hip arthroplasty using an uncemented femoral component with taper design: outcome at 10-year follow-up
    M. Schramm
    F. Keck
    D. Hohmann
    R. P. Pitto
    Archives of Orthopaedic and Trauma Surgery, 2000, 120 : 407 - 412