Minimum lyophilized plasma requirement for ISI calibration

被引:30
作者
Poller, L
Barrowcliffe, TW
Van den Besselaar, AMHP
Jespersen, J
Tripodi, A
Houghton, D
机构
[1] Univ Manchester, Dept Pathol Sci, ECAA, Cent Facil, Manchester M13 9PT, Lancs, England
[2] Natl Inst Biol Stand & Controls, Div Haematol, London, England
[3] Univ Leiden Hosp, Haemostasis & Thrombosis Res Ctr, NL-2300 RC Leiden, Netherlands
[4] Cent Sygehuset, Dept Clin Biochem, Esbjerg, Denmark
[5] Univ Milan, A Bianchi Bonomi Hemophilia & Thrombosis Ctr, I-20122 Milan, Italy
[6] IRCCS Maggiore Hosp, I-20122 Milan, Italy
关键词
D O I
10.1093/ajcp/109.2.196
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The minimum requirement of lyophilized plasma samples for a reliable International Sensitivity Index (ISI) was assessed by calibrations based on reducing numbers from a maximum of 60 depleted and 20 plasma samples from patients receiving coumarin using a manual technique and low ISI thromboplastin. The probability of achieving an international normalized ratio (INR) result within a clinically important range of 20% deviation has been assessed at European Concerted Action on Anticoagulation (ECAA) national laboratories in calibrations of ECAA reference thromboplastin against the certified prothrombin time (PT) with a reference thromboplastin. With conventional orthogonal regression using a certified PT and linear regression using certified INR, deviations and the coefficient of variation of the calibration slope increased with reduced numbers. The INR deviations became marked when the number of abnormal plasma samples was reduced to fewer than 20. Calibrations of 3 abnormal plasma samples and 1 lyophilized normal plasma sample gave a high incidence of deviations greater than 10% with an INR of 3.0. The study demonstrates that with both methods of analysis, an optimum minimum number of lyophilized plasma samples is needed for a reliable local ISI.
引用
收藏
页码:196 / 204
页数:9
相关论文
共 14 条
  • [1] [Anonymous], 1984, THROMBOPLASTIN CALIB
  • [2] ASSESSMENT OF VALUE OF CALIBRATED LYOPHILIZED PLASMAS TO DETERMINE INTERNATIONAL SENSITIVITY INDEX FOR COAGULOMETERS
    CLARKE, K
    TABERNER, DA
    THOMSON, JM
    MORRIS, JA
    POLLER, L
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1992, 45 (01) : 58 - 60
  • [3] DRAPER NR, 1966, APPLIED REGRESSION A, P7
  • [4] HOUBOUYAN LL, 1993, THROMB HAEMOSTASIS, V69, P663
  • [5] HUBBARD AR, IN PRESS INR CALIBRA
  • [6] A COMPARISON OF LYOPHILIZED ARTIFICIALLY DEPLETED PLASMAS AND LYOPHILIZED PLASMAS FROM PATIENTS RECEIVING WARFARIN IN CORRECTING FOR COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS
    POLLER, L
    TRIPLETT, DA
    HIRSH, J
    CARROLL, J
    CLARKE, K
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (03) : 366 - 371
  • [7] A simplified statistical method for local INR using linear regression
    Poller, L
    Barrowcliffe, TW
    vandenBesselaar, AMHP
    Jespersen, J
    Tripodi, A
    Houghton, D
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 1997, 98 (03) : 640 - 647
  • [8] Poller L, 1996, THROMB HAEMOSTASIS, V76, P977
  • [9] THE VALUE OF PLASMA CALIBRANTS IN CORRECTING COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS - AN INTERNATIONAL MULTICENTER STUDY
    POLLER, L
    TRIPLETT, DA
    HIRSH, J
    CARROLL, J
    CLARKE, K
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (03) : 358 - 365
  • [10] POLLER L, IN PRESS EVALUATION