Taser-induced rapid ventricular myocardial capture demonstrated by pacemaker intracardiac electrograms

被引:46
作者
Cao, Michael
Shinbane, Jerold S.
Gillberg, Jeffrey M.
Saxon, Leslie A.
机构
[1] Univ So Calif, Keck Sch Med, Div Cardiovasc Med, Los Angeles, CA 90033 USA
[2] Medtronic Inc, Minneapolis, MN USA
关键词
conducted energy devices; intracardiac electrograms; myocardial capture; pacemaker; Taser;
D O I
10.1111/j.1540-8167.2007.00881.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: A Taser weapon is designed to incapacitate violent individuals by causing temporary neuromuscular paralysis due to current application. We report the first case of a Taser application in a person with a dual-chamber pacemaker demonstrating evidence of Taser-induced myocardial capture. Methods and Results: Device interrogation was performed in a 53-year-old man with a dual-chamber pacemaker who had received a Taser shot consisting of two barbs delivered simultaneously. Assessment of pacemaker function after Taser application demonstrated normal sensing, pacing thresholds, and lead impedances. Stored event data revealed two high ventricular rate episodes corresponding to the exact time of the Taser application. Conclusions: This report describes the first human case of ventricular myocardial capture at a rapid rate resulting from a Taser application. This raises the issue as to whether conducted energy devices can cause primary myocardial capture or capture only in association with cardiac devices providing a preferential pathway of conduction to the myocardium.
引用
收藏
页码:876 / 879
页数:4
相关论文
共 14 条
[1]  
ALLEN TB, 1992, J FORENSIC SCI, V37, P956
[2]  
ANGLEN R, 2006, 167 CASES DEATH FOLL
[3]  
[Anonymous], PACING CLIN ELECTROP
[4]   Effect of a Taser shot to the chest of a patient with an implantable defibrillator [J].
Haegeli, LM ;
Sterns, LD ;
Adam, DC ;
Leather, RA .
HEART RHYTHM, 2006, 3 (03) :339-341
[5]   Cardiovascular and Physiologic effects of conducted electrical weapon discharge in resting adults [J].
Ho, Jeffrey D. ;
Miner, James R. ;
Lakireddy, Dhanunjaya R. ;
Bultman, Laura L. ;
Heegaard, William G. .
ACADEMIC EMERGENCY MEDICINE, 2006, 13 (06) :589-595
[6]   Ventricular fibrillation after stun-gun discharge. [J].
Kim, PJ ;
Franklin, WH .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 353 (09) :958-959
[7]  
KORNBLUM RN, 1991, J FORENSIC SCI, V36, P434
[8]   EfFects of cocaine intoxication on the threshold for stun gun induction of ventricular fibrillation [J].
Lakkireddy, Dhanunjaya ;
Wallick, Donald ;
Ryschon, Kay ;
Chung, Mina K. ;
Butany, JagdIsh ;
Martin, David ;
Saliba, Walid ;
Kowalewski, William ;
Natale, Andrea ;
Tchou, Patrick J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (04) :805-811
[9]  
LEVINE SD, 2005, ACAD EMERG MED S1, V12, P71
[10]   Stun guns: A new source of electromagnetic interference for implanted cardiac devices [J].
Marine, JE .
HEART RHYTHM, 2006, 3 (03) :342-344