Structural conformation and leaching from in vitro aged and retrieved Invisalign appliances

被引:97
作者
Schuster, S
Eliades, G
Zinelis, S
Eliades, T
Bradley, TG
机构
[1] Marquette Univ, Sch Dent, Dept Orthodont, Div Dev Sci, Milwaukee, WI 53233 USA
[2] Univ Athens, Sch Dent, Dept Biomat, Sect Basic Sci & Oral Biol, GR-10679 Athens, Greece
[3] Univ Athens, Sch Dent, Biomat Lab, Sect Basic Sci & Oral Biol, GR-10679 Athens, Greece
[4] Aristotle Univ Thessaloniki, Dept Orthodont, GR-54006 Thessaloniki, Greece
关键词
D O I
10.1016/j.ajodo.2004.04.021
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The objectives of this study were to investigate the structure of Invisalign appliances (Align Technology, Santa Clara, Calif) after intraoral exposure, and to qualitatively and quantitatively characterize the substances leached from the aligners after accelerated in vitro aging. Samples of Invisalign appliances were randomly selected from 10 patients before intraoral placement and after retrieval, and the prepared specimens were subjected to (1) bright-field optical reflection microscopy to study the surface morphology; (2) Fourier transform infrared microspectroscopy to characterize the in vivo changes in molecular composition induced on appliance surfaces, (3) scanning electron microscopy and energy dispersive X-ray microanalysis to identify the elemental composition of integuments formed on the surface, and (4) Vickers hardness (HV200) testing. Another set of reference and retrieved appliances was subjected to artificial aging for 2 weeks, and the extracts were subjected to gas chromatography-mass spectroscopy. The retrieved appliances demonstrated substantial morphological variation relative to the as-received specimens involving abrasion at the cusp tips, adsorption of integuments, and localized calcification of the precipitated biofilm at stagnation sites. Buccal segments of retrieved appliances showed an increase in hardness, which might be attributed to mastication-induced cold work; however, the clinical implication of this effect on mechanotherapy is unknown. In vitro aged and retrieved appliances were found to leach no traceable amount of substances in an ethanol aging solution.
引用
收藏
页码:725 / 728
页数:4
相关论文
共 17 条
  • [1] *AL TECHN INC, 2002, INV REF GUID, P8
  • [2] *AL TECHN INC, 2003, MAT SAFET DAT SHEET
  • [3] CHIA L, 1988, BASIC TECHNIQUES EXP, P134
  • [4] Eliades G., 2001, ORTHODONTIC MAT SCI, P50
  • [5] Eliades T, 2003, DENTAL MATERIALS IN VIVO: AGING AND RELATED PHENOMENA, P155
  • [6] Structural conformation of in vitro and in vivo aged orthodontic elastomeric modules
    Eliades, T
    Eliades, G
    Watts, DC
    [J]. EUROPEAN JOURNAL OF ORTHODONTICS, 1999, 21 (06) : 649 - 658
  • [7] GLANTZ PO, 1980, DENTAL PLAQUE SURFAC, P49
  • [8] OBSERVATIONS ON THE ELASTIC BEHAVIOR OF A SYNTHETIC ORTHODONTIC ELASTOMER
    HUGET, EF
    PATRICK, KS
    NUNEZ, LJ
    [J]. JOURNAL OF DENTAL RESEARCH, 1990, 69 (02) : 496 - 501
  • [9] KASEMO B, 1991, BONE-BIOMATERIAL INTERFACE, P19
  • [10] SPECTROSCOPIC APPROACHES TO THE INVESTIGATION OF INTERACTIONS BETWEEN ARTIFICIAL SURFACES AND PROTEINS
    LEININGER, RI
    HUTSON, TB
    JAKOBSEN, RJ
    [J]. ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 516 : 173 - 183