Use of a saline chaser in abdominal computed tomography: a systematic review

被引:6
作者
Takao, Hidemasa [1 ]
Nojo, Takeshi [2 ]
Ohtomo, Kuni [1 ]
机构
[1] Univ Tokyo, Dept Radiol, Grad Sch Med, Bunkyo Ku, Tokyo 1138655, Japan
[2] Kyoto Univ, Dept Epidemiol & Healthcare Res, Grad Sch Med & Publ Hlth, Sakyo Ku, Kyoto, Japan
关键词
Computed tomography; Contrast media; Saline chaser; Saline flush; Saline pusher; CONTRAST-ENHANCED CT; POWER INJECTOR; DIFFERENT VOLUMES; BOLUS TRACKING; HELICAL CT; SPIRAL CT; ROW CT; FLUSH; ANGIOGRAPHY; REDUCTION;
D O I
10.1016/j.clinimag.2008.10.032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To perform a systematic review of the literature regarding the use of a saline chaser in abdominal computed tomography (CT). Materials and methods: The MEDLINE database was searched from 1966 to March 2007. Studies were included if they compared the magnitude of contrast enhancement with and without a saline chaser in CT and reported CT contrast enhancement values of the liver, portal vein, or abdominal aorta. For randomized controlled trials (RCTs) comparing the same dose of contrast material with and without a saline chaser, mean differences in the magnitude of contrast enhancement were pooled by using a fixed-effects model. Results: Seventeen studies met the inclusion criteria. Three RCTs using clinical images were included in the meta-analysis. The mean differences in contrast enhancement of the liver, portal vein, and aorta were -0.4 (95% CI, -2.4 to 1.7), 7.4 (95% CI, 1.3 to 14), and 11 (95% CI, -5.3 to 27), respectively. Three RCTs using a time-density analysis were included in the meta-analysis. The mean differences in peak contrast enhancement of the liver, portal vein, and aorta were 8.2 (95% CI, 4.2 to 12), 28 ( 95% CI, 16 to 40), and 15 (95% CI, 0.3 to 30), respectively. Conclusion: In clinical images, a saline chaser did not improve contrast enhancement of the liver. In time-density analysis, however, a saline chaser improved peak contrast enhancement of the liver. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 30 条
[1]  
[Anonymous], 2001, SYSTEMATIC REV HLTH, DOI DOI 10.1002/9780470693926
[2]   Non-invasive 16-row multislice CT coronary angiography: usefulness of saline chaser [J].
Cademartiri, F ;
Mollet, N ;
van der Lugt, A ;
Nieman, K ;
Pattynama, PMT ;
de Feyter, PJ ;
Krestin, GP .
EUROPEAN RADIOLOGY, 2004, 14 (02) :178-183
[3]  
Cademartiri Filippo, 2004, Radiol Med, V107, P497
[4]   Sixteen-detector row CT angiography of carotid arteries:: Comparison of different volumes of contrast material with and without a bolus chaser [J].
de Monyé, C ;
Cademartiri, F ;
de Weert, TT ;
Siepman, DAM ;
Dippel, DWJ ;
van Der Lugt, A .
RADIOLOGY, 2005, 237 (02) :555-562
[5]   Using a saline chaser to decrease contrast media in abdominal CT [J].
Dorio, PJ ;
Lee, FT ;
Henseler, KP ;
Pilot, M ;
Pozniak, MA ;
Winter, TC ;
Shock, SA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (04) :929-934
[6]  
Glasziou P., 2001, SYSTEMATIC REV HLTH
[7]   Reduction of contrast material dose and artifacts by a saline flush using a double power injector in helical CT of the thorax [J].
Haage, P ;
Schmitz-Rode, T ;
Hübner, D ;
Piroth, W ;
Günther, RW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (04) :1049-1053
[8]   Multidetector CT angiography using a dual-head power injector and bolus tracking for the diagnosis of aortic aneurysms at reduced contrast material dosage [J].
Hirai, Nobuhiko ;
Imakita, Satoshi ;
Tanaka, Ryoichi ;
Higashi, Masahiro ;
Nishino, Toshihiro ;
Naito, Hiroaki ;
Ito, Katsuhide .
ACADEMIC RADIOLOGY, 2006, 13 (06) :694-700
[9]   Thoracic spiral CT: Delivery of contrast material pushed with injectable saline solution in a power injector [J].
Hopper, KD ;
Mosher, TJ ;
Kasales, CJ ;
TenHave, TR ;
Tully, DA ;
Weaver, JS .
RADIOLOGY, 1997, 205 (01) :269-271
[10]   Contrast-enhanced CT with saline flush technique using two automated injectors: How much contrast medium does it save? [J].
Irie, T ;
Kajitani, M ;
Yamaguchi, M ;
Itai, Y .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2002, 26 (02) :287-291