Acetylcholinesterase biosensor based on prussian blue-modified electrode for detecting organophosphorous pesticides

被引:173
|
作者
Sun, Xia [1 ]
Wang, Xiangyou [1 ]
机构
[1] Shandong Univ Technol, Sch Agr & Food Engn, Zibo 255049, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
Biosensor; Prussian blue; Acetylcholinesterase; Pesticide residue; Modified electrode; NANOPARTICLES; CARBAMATE;
D O I
10.1016/j.bios.2010.04.028
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this study, a novel acetylcholinesterase (AChE) biosensor was developed based on dual-layer membranes (chitosan membrane and prussian blue membrane) modifying glassy carbon electrode (GCE). A chitosan membrane was used for immobilizing AChE through glutaraldehyde cross-linking attachment to recognize pesticides selectively. A Prussian blue (PB) membrane was electrodeposited on the surface of GCE to enhance electron transfer. Before the detection, the chitosan enzyme membrane was quickly fixed on the surface of PB/GCE with O-ring to prepare an amperometric AChE-PB/GCE sensor for organophosphorus (OP) pesticides. The electrochemical behaviour of AChE-PB/GCE was studied, and the results showed that the chitosan membrane as carrier can absorb a large amount of enzyme, and PB has a significant synergistic effect towards enzymatic catalysis. As a result of these two important enhancement factors, the proposed biosensor exhibited extreme sensitivity to OP pesticides compared to the other kinds of AChE biosensor. The influences of phosphate buffer pH, substrate concentration, incubation time of pesticide on the response of the fabricated biosensor were investigated. Under optimum conditions, the inhibition rates of these pesticides were proportional to their concentrations in the range of 0.01-10 mu g l(-1), 0.05-10 mu g l(-1), 0.03-5 mu g l(-1), and 0.05-10 mu g l(-1), respectively. The detection limits were found to be 2.5 ng l(-1) for dichlorvos, 15 ng l(-1) for omethoate. 5 ng l(-1) for trichlorfon and 10 ng l(-1) for phoxim. Moreover, the biosensor exhibited good reproducibility and stability, and it was suitable for trace detection of OP pesticide residue. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:2611 / 2614
页数:4
相关论文
共 50 条
  • [1] Detection of carbamic and organophosphorous pesticides in water samples using a cholinesterase biosensor based on Prussian Blue-modified screen-printed electrode
    Arduini, Fabiana
    Ricci, Francesco
    Tuta, Catalin S.
    Moscone, Danila
    Amine, Aziz
    Palleschi, Giuseppe
    ANALYTICA CHIMICA ACTA, 2006, 580 (02) : 155 - 162
  • [2] Multiwall Carbon Nanotube-Based Acetylcholinesterase Biosensor for Detecting Organophosphorous Pesticides
    Sun, Xia
    Wang, Xiangyou
    Zhao, Wenping
    SENSOR LETTERS, 2009, 7 (06) : 1065 - 1071
  • [3] Multiwall Carbon Nanotube-Based Acetylcholinesterase Biosensor for Detecting Organophosphorous Pesticides
    Sun, Xia
    Wang, Xiangyou
    Zhao, Wenping
    SENSOR LETTERS, 2010, 8 (02) : 247 - 252
  • [4] A novel biosensor based on acetylecholinesterase/prussian blue-chitosan modified electrode for detection of carbaryl pesticides
    Song, Yonghai
    Zhang, Mei
    Wang, Li
    Wan, Lingli
    Xiao, Xianping
    Ye, Shuhong
    Wang, Jingran
    ELECTROCHIMICA ACTA, 2011, 56 (21) : 7267 - 7271
  • [5] Nonenzymatic hydrogen peroxide sensor based on a Prussian Blue-modified carbon ionic liquid electrode
    Li, Yonghong
    Liu, Xiaoying
    Zeng, Xiandong
    Liu, Yong
    Liu, Xinsheng
    Wei, Wanzhi
    Luo, Shenglian
    MICROCHIMICA ACTA, 2009, 165 (3-4) : 393 - 398
  • [6] Amperometric biosensor based on Prussian Blue-modified screen-printed electrode for lipase activity and triacylglycerol determination
    Rejeb, Ines Ben
    Arduini, Fabiana
    Amine, Aziz
    Gargouri, Mohamed
    Palleschi, Giuseppe
    ANALYTICA CHIMICA ACTA, 2007, 594 (01) : 1 - 8
  • [7] Behavior of PQQ glucose dehydrogenase on Prussian blue-modified carbon electrode
    Laurinavicius, Valdas
    Kurtinaitiene, Bogumila
    Stankeviciute, Rasa
    ELECTROANALYSIS, 2008, 20 (13) : 1391 - 1395
  • [8] Amperometric detection of hydrogen peroxide at a Prussian Blue-modified FTO electrode
    Tseng, KS
    Chen, LC
    Ho, KC
    SENSORS AND ACTUATORS B-CHEMICAL, 2005, 108 (1-2) : 738 - 745
  • [9] Prussian blue-modified nanoporous gold film electrode for amperometric determination of hydrogen peroxide
    Ghaderi, Seyran
    Mehrgardi, Masoud Ayatollahi
    BIOELECTROCHEMISTRY, 2014, 98 : 64 - 69
  • [10] Amperometric thiol sensor based on Prussian blue-modified glassy carbon electrode
    Kafi, A. K. M.
    Yin, Fan
    Shin, Hoon-Kyu
    Kwon, Young-Soo
    CURRENT APPLIED PHYSICS, 2007, 7 (05) : 496 - 499