摘要
Background: Residual paravalvular moderate or severe aortic regurgitation (AR) has been an independent risk factor for mortality after transcatheter aortic valve implantation (TAVI). The design of the third generation Edwards Sapien 3 (ES3; Edwards Lifesciences, Irvine, CA, USA) valve was optimized with an outer skirt to address the issue of paravalvular AR. Methods and results: We compared 100 consecutive patients treated with the ES3 for severe aortic stenosis with 100 patients treated with the Medtronic CoreValve (CV; Medtronic, Minneapolis, MN, USA) (Clinical Trial Registration: NCT02162069). We evaluated post-procedural AR, rate of permanent pacemaker implantation, device success, and 30-day clinical outcome according to the criteria of the Second Valve Academic Research Consortium (VARC-2). Frequency of post-procedural moderate or severe AR was significantly lower with ES3 compared to CV (0% vs. 20%, p < 0.01), none or trace AR significantly higher with ES3 (69% vs. 38%, p < 0.01) as well as device success (97% vs. 73%, p < 0.01). There was a significantly lower need for permanent pacemaker implantation with ES3 compared with CV (14% vs. 31%, p < 0.01). Cardiovascular mortality at 30 days was significantly lower with ES3 (0% vs. 6%, p = 0.01), and the combined endpoint "early safety" was met significantly less with ES3 (10% vs. 21% with CV, p = 0.03). Conclusions: Transfemoral TAVI with the ES3 compared with the CV was associated with a significantly lower rate of moderate or severe AR, significantly lower need for pacemaker implantation, and a significantly higher rate of device success according to VARC-2. (C) 2016 Japanese College of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:877 / 882
页数:6
相关论文
共 32 条
[1]   1-Year Outcomes After Transcatheter Aortic Valve Replacement With Balloon-Expandable Versus Self-Expandable Valves Results From the CHOICE Randomized Clinical Trial [J].
Abdel-Wahab, Mohamed ;
Neumann, Franz-Josef ;
Mehilli, Julinda ;
Frerker, Christian ;
Richardt, Doreen ;
Landt, Martin ;
Jose, John ;
Toelg, Ralph ;
Kuck, Karl-Heinz ;
Massberg, Steffen ;
Robinson, Derek R. ;
El-Mawardy, Mohamed ;
Richardt, Gert .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 66 (07) :791-800
[2]   Comparison of Balloon-Expandable vs Self-expandable Valves in Patients Undergoing Transcatheter Aortic Valve Replacement The CHOICE Randomized Clinical Trial [J].
Abdel-Wahab, Mohamed ;
Mehilli, Julinda ;
Frerker, Christian ;
Neumann, Franz-Josef ;
Kurz, Thomas ;
Toelg, Ralph ;
Zachow, Dirk ;
Guerra, Elena ;
Massberg, Steffen ;
Schaefer, Ulrich ;
El-Mawardy, Mohamed ;
Richardt, Gert .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2014, 311 (15) :1503-1514
[3]   Transcatheter Aortic-Valve Replacement with a Self-Expanding Prosthesis [J].
Adams, David H. ;
Popma, Jeffrey J. ;
Reardon, Michael J. ;
Yakubov, Steven J. ;
Coselli, Joseph S. ;
Deeb, G. Michael ;
Gleason, Thomas G. ;
Buchbinder, Maurice ;
Hermiller, James, Jr. ;
Kleiman, Neal S. ;
Chetcuti, Stan ;
Heiser, John ;
Merhi, William ;
Zorn, George ;
Tadros, Peter ;
Robinson, Newell ;
Petrossian, George ;
Hughes, G. Chad ;
Harrison, J. Kevin ;
Conte, John ;
Maini, Brijeshwar ;
Mumtaz, Mubashir ;
Chenoweth, Sharla ;
Oh, Jae K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (19) :1790-1798
[4]   Comparison of Hemodynamic Performance of the Balloon-Expandable SAPIEN 3 Versus SAPIEN XT Transcatheter Valve [J].
Amat-Santos, Ignacio J. ;
Dahou, Abdellaziz ;
Webb, John ;
Dvir, Danny ;
Dumesnil, Jean G. ;
Allende, Ricardo ;
Ribeiro, Henrique B. ;
Urena, Marina ;
Paradis, Jean-Michel ;
DeLarochelliere, Robert ;
Dumont, Eric ;
Bergeron, Sebastien ;
Thompson, Christopher R. ;
Pasian, Sergio ;
Bilodeau, Sylvie ;
Leipsic, Jonathon ;
Larose, Eric ;
Pibarot, Philippe ;
Rodes-Cabau, Josep .
AMERICAN JOURNAL OF CARDIOLOGY, 2014, 114 (07) :1075-1082
[5]   Incidence, Predictors, and Outcomes of Aortic Regurgitation After Transcatheter Aortic Valve Replacement Meta-Analysis and Systematic Review of Literature [J].
Athappan, Ganesh ;
Patvardhan, Eshan ;
Tuzcu, E. Murat ;
Svensson, Lars Georg ;
Lemos, Pedro A. ;
Fraccaro, Chiara ;
Tarantini, Giuseppe ;
Sinning, Jan-Malte ;
Nickenig, Georg ;
Capodanno, Davide ;
Tamburino, Corrado ;
Latib, Azeem ;
Colombo, Antonio ;
Kapadia, Samir R. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (15) :1585-1595
[6]   2-Year Follow-Up of Patients Undergoing Transcatheter Aortic Valve Implantation Using a Self-Expanding Valve Prosthesis [J].
Buellesfeld, Lutz ;
Gerckens, Ulrich ;
Schuler, Gerhard ;
Bonan, Raoul ;
Kovac, Jan ;
Serruys, Patrick W. ;
Labinaz, Marino ;
den Heijer, Peter ;
Mullen, Michael ;
Tymchak, Wayne ;
Windecker, Stephan ;
Mueller, Ralf ;
Grube, Eberhard .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (16) :1650-1657
[7]   Long-Term Outcomes for Patients With Severe Symptomatic Aortic Stenosis Treated With Transcatheter Aortic Valve Implantation [J].
Codner, Pablo ;
Orvin, Katia ;
Assali, Abid ;
Sharony, Ram ;
Vaknin-Assa, Hanna ;
Shapira, Yaron ;
Schwartzenberg, Shmuel ;
Bental, Tamir ;
Sagie, Alexander ;
Kornowski, Ran .
AMERICAN JOURNAL OF CARDIOLOGY, 2015, 116 (09) :1391-1398
[8]   Chronic pacing and adverse outcomes after transcatheter aortic valve implantation [J].
Dizon, Jose' M. ;
Nazif, Tamim M. ;
Hess, Paul L. ;
Biviano, Angelo ;
Garan, Hasan ;
Douglas, Pamela S. ;
Kapadia, Samir ;
Babaliaros, Vasilis ;
Herrmann, Howard C. ;
Szeto, Wilson Y. ;
Jilaihawi, Hasan ;
Fearon, William F. ;
Tuzcu, E. Murat ;
Pichard, Augusto D. ;
Makkar, Raj ;
Williams, Mathew ;
Hahn, Rebecca T. ;
Xu, Ke ;
Smith, Craig R. ;
Leon, Martin B. ;
Kodali, Susheel K. .
HEART, 2015, 101 (20) :1665-1671
[9]   Complications and Mortality in Patients Undergoing Transcatheter Aortic Valve Replacement With Edwards SAPIEN & SAPIEN XT Valves: A Meta-Analysis of World-Wide Studies and Registries Comparing the Transapical and Transfemoral Accesses [J].
Ghatak, Abhijit ;
Bavishi, Chirag ;
Cardoso, Rhanderson N. ;
Macon, Conrad ;
Singh, Vikas ;
Badheka, Apurva O. ;
Padala, Santosh ;
Cohen, Mauricio G. ;
Mitrani, Raul ;
O'neill, William ;
De Marchena, Eduardo .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2015, 28 (03) :266-278
[10]   Registry of Transcatheter Aortic-Valve Implantation in High-Risk Patients [J].
Gilard, Martine ;
Eltchaninoff, Helene ;
Iung, Bernard ;
Donzeau-Gouge, Patrick ;
Chevreul, Karine ;
Fajadet, Jean ;
Leprince, Pascal ;
Leguerrier, Alain ;
Lievre, Michel ;
Prat, Alain ;
Teiger, Emmanuel ;
Lefevre, Thierry ;
Himbert, Dominique ;
Tchetche, Didier ;
Carrie, Didier ;
Albat, Bernard ;
Cribier, Alain ;
Rioufol, Gilles ;
Sudre, Arnaud ;
Blanchard, Didier ;
Collet, Frederic ;
Dos Santos, Pierre ;
Meneveau, Nicolas ;
Tirouvanziam, Ashok ;
Caussin, Christophe ;
Guyon, Philippe ;
Boschat, Jacques ;
Le Breton, Herve ;
Collart, Frederic ;
Houel, Remi ;
Delpine, Stephane ;
Souteyrand, Geraud ;
Favereau, Xavier ;
Ohlmann, Patrick ;
Doisy, Vincent ;
Grollier, Gilles ;
Gommeaux, Antoine ;
Claudel, Jean-Philippe ;
Bourlon, Francois ;
Bertrand, Bernard ;
Van Belle, Eric ;
Laskar, Marc .
NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (18) :1705-1715