The safe limits of mechanical factors in the apnea testing for the diagnosis of brain death

被引:6
作者
Olguner, Cimen [1 ]
Koca, Ugur [1 ]
Akan, Mert [1 ]
Karci, Ayse [1 ]
Elar, Zahide [1 ]
机构
[1] Dokuz Eylul Univ, Sch Med, Dept Anesthesiol & Reanimat, TR-35340 Izmir, Turkey
关键词
brain death; apnea testing; barotrauma; organ donation; airway simulator;
D O I
10.1620/tjem.211.115
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Apneic oxygenation is an apnea testing method in the diagnosis of brain death. In this method, oxygen (O-2) is delivered into the trachea via an O-2 catheter (O2C). However, barotrauma may develop during O-2 insufflation into the trachea. Oxygen catheter diameters, O-2 catheter tip position in the trachea, and O-2 flow rate have been proposed as causes of barotrauma. This study was designed to highlight the airway pressure changes during apneic oxygenation in a model consisting of an anesthesia bag, which was connected to a pressure transducer and to an endotracheal tube (ETT). The pressure of the system was monitored while delivering O-2 continuously to the system through O2C of different diameters, which were placed in the ETT. Tested variables were ETT/O2C ratio, O2C tip position in ETT (proximal 1/3 of the ETT, mid point of the ETT, and distal 1/3 of the ETT) and O-2 flow rate (6, 8, and 10 L min(-1)). The increase in the airway pressure significantly correlated with O2C tip position in ETT (p = 0.017). ETT/O2C ratio smaller than 1.75 caused significantly high airway pressures (p < 0.05). The pressure was significantly higher at the flow rate of 10 L min(-1) O-2 compared with the flow rate of 6 L min(-1) O-2 (p < 0.01). Thus, ETT/O2C ratio, O2C tip position in ETT and O-2 flow rate are the important factors that determine the airway pressure in the trachea during O-2 insufflation. In conclusion, overlooked mechanical factors dangerously increase airway pressure during apnea testing.
引用
收藏
页码:115 / 120
页数:6
相关论文
共 13 条
  • [1] [Anonymous], 1995, Neurology, V45, P1012
  • [2] Tension pneumothorax during apnea testing for the determination of brain death
    Bar-Joseph, G
    Bar-Lavie, Y
    Zonis, Z
    [J]. ANESTHESIOLOGY, 1998, 89 (05) : 1250 - 1251
  • [3] Effect of catheter design on tracheal pressures during tracheal gas insufflation
    Dyer, IR
    Esmail, M
    Findlay, G
    Mecklenburgh, JS
    Dingley, J
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2003, 20 (09) : 740 - 744
  • [4] Complications during apnea testing in the determination of brain death: Predisposing factors
    Goudreau, JL
    Wijdicks, EFM
    Emery, SF
    [J]. NEUROLOGY, 2000, 55 (07) : 1045 - 1048
  • [5] Apnea testing for the diagnosis of brain death
    Lang, CJG
    Heckmann, JG
    [J]. ACTA NEUROLOGICA SCANDINAVICA, 2005, 112 (06): : 358 - 369
  • [6] Efficacy of a T-piece system and a continuous positive airway pressure system for apnea testing in the diagnosis of brain death
    Levesque, Simon
    Lessard, Martin R.
    Nicole, Pierre C.
    Langevin, Stephan
    LeBlanc, Francois
    Lauzier, Francois
    Brochu, Jacques G.
    [J]. CRITICAL CARE MEDICINE, 2006, 34 (08) : 2213 - 2216
  • [7] Potential pitfalls in apnea testing
    Rudolf, J
    Haupt, WF
    Neveling, M
    Grond, M
    [J]. ACTA NEUROCHIRURGICA, 1998, 140 (07) : 659 - 663
  • [8] Pneumothorax and pneumoperitoneum during the apnea test - How safe is this procedure?
    Saposnik, G
    Rizzo, G
    Deluca, JL
    [J]. ARQUIVOS DE NEURO-PSIQUIATRIA, 2000, 58 (3B) : 905 - 908
  • [9] Saposnik G, 2004, NEUROL INDIA, V52, P342
  • [10] Sharples A, 1997, ANAESTHESIA, V52, P503