Total body irradiation-an attachment free sweeping beam technique

被引:12
作者
Haertl, Petra M. [1 ]
Treutwein, Marius [1 ]
Hautmann, Matthias G. [1 ]
Maerz, Manuel [1 ]
Pohl, Fabian [1 ]
Koelbl, Oliver [1 ]
Dobler, Barbara [1 ]
机构
[1] Univ Regensburg, Med Ctr, Dept Radiotherapy, D-93053 Regensburg, Germany
关键词
Total body irradiation; Sweeping beam technique; Lung shielding; Dosimetry; 3D treatment planning; STEM-CELL TRANSPLANTATION; INTERSTITIAL PNEUMONITIS; DOSE-RATE; VERIFICATION; TOXICITY; THERAPY; TBI; ARC;
D O I
10.1186/s13014-016-0658-y
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: A sweeping beam technique for total body irradiation in standard treatment rooms and for standard linear accelerators (linacs) is introduced, which does not require any accessory attached to the linac. Lung shielding is facilitated to reduce the risk of pulmonary toxicity. Additionally, the applicability of a commercial radiotherapy planning system (RTPS) is examined. Material and Methods: The patient is positioned on a low couch on the floor, the longitudinal axis of the body in the rotational plane of the linac. Eight arc fields and five additional fixed beams are applied to the patient in supine and prone position respectively. The dose distributions were measured in a solid water phantom and in an Alderson phantom. Diode detectors were calibrated for in-vivo dosimetry. The RTPS Oncentra was employed for calculations of the dose distribution. Results: For the cranial 120 cm the longitudinal dose profile in a slab phantom measured with ionization chamber varies between 94 and 107 % of the prescription dose. These values were confirmed by film measurements and RTPS calculations. The transmittance of the lung shields has been determined as a function of the thickness of the absorber material. Measurements in an Alderson phantom and in-vivo dosimetry of the first patients match the calculated dose. Discussion and conclusion: A treatment technique with clinically good dose distributions has been introduced, which can be applied with each standard linac and in standard treatment rooms. Dose calculations were performed with a commercial RTPS and should enable individual dose optimization.
引用
收藏
页数:8
相关论文
共 36 条
[1]  
Abraham D, 2000, Med Dosim, V25, P219, DOI 10.1016/S0958-3947(00)00049-2
[2]  
Achterberg N, 1997, 28 WISS TAG DTSCH GE, V28, P83
[3]   Total body irradiation before an allogeneic stem cell transplantation: is there a magic dose? [J].
Adkins, Douglas R. ;
DiPersio, John F. .
CURRENT OPINION IN HEMATOLOGY, 2008, 15 (06) :555-560
[4]  
APPELBAUM FR, 1993, SEMIN ONCOL, V20, P3
[5]  
Belkacemi Y, 1998, STRAHLENTHER ONKOL, V174, P92, DOI 10.1007/BF03038482
[6]  
Bölling T, 2011, STRAHLENTHER ONKOL, V187, P311, DOI 10.1007/s00066-011-2190-1
[7]  
Born C, 2001, STRAHLENTHER ONKOL S, V177, P114
[8]  
Cegla R., 2007, THESIS FACHHOCHSCHUL
[9]   Impact of drug therapy, radiation dose, and dose rate on renal toxicity following bone marrow transplantation [J].
Cheng, Jonathan C. ;
Schultheiss, Timothy E. ;
Wong, Jeffrey Y. C. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (05) :1436-1443
[10]   Total body irradiation with an arc and a gravity-oriented compensator [J].
Chui, CS ;
Fontenla, DP ;
Mullokandov, E ;
Kapulsky, A ;
Lo, YC ;
Lo, CJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (05) :1191-1195