Comparison of tunable bandpass reaction cell inductively coupled plasma mass spectrometry with conventional inductively coupled plasma mass spectrometry for the determination of heavy metals in whole blood and urine

被引:42
作者
Nixon, DE
Neubauer, KR
Eckdahl, SJ
Butz, JA
Burritt, MF [1 ]
机构
[1] Mayo Clin, Div Clin Biochem & Immunol, Met Lab, Rochester, MN 55905 USA
[2] PerkinElmer Life & Analyt Sci, Shelton, CT 06484 USA
关键词
arsenic; lead; cadmium; mercury; thallium; urine; whole blood; inductively coupled plasma mass spectrometry; dynamic reaction cell inductively coupled plasma mass spectrometry; heavy metal screen;
D O I
10.1016/j.sab.2004.05.013
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A Dynamic Reaction Cell(TM) inductively coupled plasma mass spectrometer (DRC-ICP-MS) was evaluated for the determination of arsenic, lead, cadmium, mercury, and thallium in urine and whole blood. Reaction cell conditions were evaluated for suppression of ArCl+ and CaCl+ polyatomic interferences. The reaction gas was 5% hydrogen in argon. Lead, cadmium, mercury, and thallium were determined with the reaction cell vented. Mixture of 2.5% t-butanol, 0.5% HCl, and 2 mg Au/l plus Ga, Rh, and Bi internal standards was used to dilute whole blood and urine. Calibration was achieved using aqueous acidic standards spiked into urine matrix. Urine and whole blood addition calibration curves were nearly identical for all five elements. DRC-ICP-MS detection limits were equivalent or better than conventional ICP-MS. Within run coefficients of variation (CV's) were nearly the same for DRC-ICP-MS and conventional ICP-MS for National Institute of Standards and Technology (NIST) SRM 2670 and BioRad Lyphochek Urine Metals Control. DRC-ICP-MS within run CV's for As, Pb, Cd, and Hg were 1.9%, 4%, 1.7%, and 1.7%, respectively, for NIST 2670 and 2.9%, 1.8%, 3.4%, 1.7%, and 1.0% for BioRad urine. BioRad Lyphochek Whole Blood control concentrations and CV's were: 78 mug/l (3.8%), 284 mug/l (0.52%), and 544 mug/l (0.9%). With the exception of mercury day-to-day CV's for certified whole blood and urine controls were less than 4% on both the DRC-ICP-MS and conventional ICP-MS. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1377 / 1387
页数:11
相关论文
共 18 条
[1]  
ABOUSHAKRA F, 1997, PLASMA SOURCE MASS S, P120
[2]   MEASUREMENT OF VANADIUM, NICKEL, AND ARSENIC IN SEAWATER AND URINE REFERENCE MATERIALS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH CRYOGENIC DESOLVATION [J].
ALVES, LC ;
ALLEN, LA ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1993, 65 (18) :2468-2471
[3]   Determination of the total arsenic concentration in human urine by inductively coupled plasma mass spectrometry: a comparison of the accuracy of three analytical methods [J].
Amarasiriwardena, CJ ;
Lupoli, N ;
Potula, V ;
Korrick, S ;
Hu, H .
ANALYST, 1998, 123 (03) :441-445
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Denoyer ER, 1999, SPECTROSCOPY, V14, P43
[6]   PRESERVATION OF DILUTE MERCURY SOLUTIONS [J].
FELDMAN, C .
ANALYTICAL CHEMISTRY, 1974, 46 (01) :99-102
[7]  
KERSHISNIK MM, 1992, CLIN CHEM, V38, P2197
[8]   MERCURY LOSS FROM WATER DURING STORAGE - MECHANISMS AND PREVENTION [J].
LO, JM ;
WAI, CM .
ANALYTICAL CHEMISTRY, 1975, 47 (11) :1869-1870
[9]  
Neubauer K, 1999, ATOM SPECTROSC, V20, P64
[10]   Evaluation of a tunable bandpass reaction cell for an inductively coupled plasma mass spectrometer for the determination of chromium and vanadium in serum and urine [J].
Nixon, DE ;
Neubauer, KR ;
Eckdahl, SJ ;
Butz, JA ;
Burritt, MF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (05) :951-966