A Simple Low-Cost Electrocardiogram Synchronizer

被引:0
|
作者
Amoros, Susana [1 ]
Galvez-Monton, Carolina [2 ,3 ]
Rodriguez-Leor, Oriol [2 ,3 ]
Manuel O'Callaghan, Juan [1 ]
机构
[1] Univ Politecn Cataluna, Sch Telecommun Engn, Barcelona 08034, Spain
[2] Hlth Sci Res Inst Germans Trias & Pujol, ICREC Res Program, Can Ruti Campus, Badalona 08916, Spain
[3] Inst Salud Carlos III, CIBERCV, Madrid 28029, Spain
关键词
biomedical signals; ECG; bioinstrumentation; synchronization; trigger; medical sensor development; CORONARY CT ANGIOGRAPHY;
D O I
10.3390/s21175885
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electrocardiogram (ECG) synchronization is useful to avoid the effects of cardiac motion in medical measurements, and is widely used in standard medical imaging. A number of medical equipment include embedded commercial synchronizers. However, the use of independent synchronization modules is sometimes needed when several non-integrated instruments are used, or in the development of new medical instruments and procedures. We present a simple low-cost ECG synchronizer module based on an Arduino controller board that converts the ECG signal into a transistor-transistor-logic (TTL) one, allowing real-time medical measurements triggered at specific phases of the cardiac cycle. The device and conversion algorithm developed is optimized in vitro using synthetic and human ECG signals, and tested in vivo on three swine specimens. Error rates during the in vivo testing stage remain below the 2% of the cycles in all animals and critical false positives are less than 1%, which is sufficient for most applications. Possible algorithm updates are discussed if its performance needs to be improved.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] A Flexible Low-cost Time Synchronizer for GNSS/INS Integration
    Du, Gang
    Wang, Liduan
    Zhai, Chuanrun
    Zhan, Xingqun
    PROCEEDINGS OF THE 23RD INTERNATIONAL TECHNICAL MEETING OF THE SATELLITE DIVISION OF THE INSTITUTE OF NAVIGATION (ION GNSS 2010), 2010, : 2100 - 2105
  • [2] LOW-COST ON LINE STORAGE AND RETRIEVAL OF ELECTROCARDIOGRAM
    BENEDEK, M
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (05): : A77 - A77
  • [3] A SIMPLE LOW-COST FOOTSWITCH
    LIGGINS, AB
    BOWKER, P
    JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (01): : 87 - 88
  • [4] A SIMPLE, LOW-COST SCANNER
    GODWIN, J
    LASER FOCUS WITH FIBEROPTIC TECHNOLOGY, 1981, 17 (10): : 91 - 93
  • [5] Low-cost, Simple, Diagnostics
    Whitesides, George
    NEW BIOTECHNOLOGY, 2012, 29 : S1 - S1
  • [6] A low-cost simple lens
    Noriaki Horiuchi
    Nature Photonics, 2012, 6 : 418 - 418
  • [7] Simple, low-cost, subsystem fabrication
    Liu, HZ
    Huang, HK
    Wang, CC
    Chiu, RJ
    Lin, CH
    Chu, CK
    Houng, MP
    Wang, YH
    Chang, CH
    Wu, CL
    Chang, CS
    IEEE CIRCUITS & DEVICES, 2005, 21 (04): : 12 - 17
  • [8] A SIMPLE AND LOW-COST DNA AMPLIFIER
    ROLLO, F
    AMICI, A
    SALVI, R
    NUCLEIC ACIDS RESEARCH, 1988, 16 (07) : 3105 - 3106
  • [9] SIMPLE LOW-COST CIRCULAR PONDS
    HAWKE, SP
    FIELDDODGSON, MS
    PROGRESSIVE FISH-CULTURIST, 1987, 49 (01): : 75 - 76
  • [10] LOW-COST, SIMPLE GATE VALVE
    ALONSO, EV
    JIMENEZRODRIGUEZ, JJ
    MARTINEZALBERTOS, JL
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (03): : 2738 - 2739