Alumina-on-alumina total hip arthroplasty has been used for 32 years in Europe. The theoretical advantages of this combination are represented by its remarkable sliding characteristics, its very low wear debris generation, and its improved fracture toughness. These advantages are achieved if the material is processed properly with high density, high purity and small grains. We summarize the results obtained with the alumina-on-alumina combination concerning in vitro and in vivo wear behavior with special emphasis on wear debris characterization and quantification and histologic tissue examinations. Alumina-on-alumina seems to be one of the best choices in young and active patients provided that sound socket fixation is maintained in the long term.
引用
收藏
页码:62 / 71
页数:10
相关论文
共 80 条
[31]   Phase transformation of a zirconia ceramic head after total hip arthroplasty [J].
Haraguchi, K ;
Sugano, N ;
Nishii, T ;
Miki, H ;
Oka, K ;
Yoshikawa, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (07) :996-1000
[32]   TISSUE REACTION TO CERAMIC IMPLANT MATERIAL [J].
HARMS, J ;
MAUSLE, E .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1979, 13 (01) :67-87
[33]   PROSTHETIC COMPONENT FAILURES IN HIP-ARTHROPLASTY SURGERY [J].
HECK, DA ;
PARTRIDGE, CM ;
REUBEN, JD ;
LANZER, WL ;
LEWIS, CG ;
KEATING, EM .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :575-580
[34]   5 YEARS EXPERIENCE WITH CERAMIC-METAL-COMPOSITE HIP ENDOPROSTHESES .2. MECHANICAL EVALUATIONS AND IMPROVEMENTS [J].
HEIMKE, G ;
GRISS, P .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1981, 98 (03) :165-171
[35]   Catastrophic failure of modular zirconia-ceramic femoral head components after total hip arthroplasty [J].
Hummer, CD ;
Rothman, RH ;
Hozack, WJ .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :848-850
[36]  
HUO MH, 1996, CLIN ORTHOP RELAT R, V332, P143
[37]   A comparison of polyethylene wear in hips with cobalt-chrome or zirconia heads - A prospective. Randomised study [J].
Kim, YH ;
Kim, JS ;
Cho, SH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (05) :742-750
[38]   LOW WEAR RATE OF UHMWPE AGAINST ZIRCONIA CERAMIC (Y-PSZ) IN COMPARISON TO ALUMINA CERAMIC AND SUS 316L ALLOY [J].
KUMAR, P ;
OKA, M ;
IKEUCHI, K ;
SHIMIZU, K ;
YAMAMURO, T ;
OKUMURA, H ;
KOTOURA, Y .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :813-828
[39]  
Le Mouel S, 1998, REV CHIR ORTHOP, V84, P338
[40]  
Lerouge S, 1996, J BIOMED MATER RES, V32, P627, DOI 10.1002/(SICI)1097-4636(199612)32:4<627::AID-JBM16>3.0.CO