Experimental determination of the lateral dose response functions of detectors to be applied in the measurement of narrow photon-beam dose profiles

被引:21
作者
Poppinga, D. [1 ]
Meyners, J. [1 ]
Delfs, B. [1 ]
Muru, A. [1 ]
Harder, D. [2 ]
Poppe, B. [1 ]
Looe, H. K. [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Med Campus Pius Hosp, Univ Clin Med Radiat Phys, Oldenburg, Germany
[2] Univ Gottingen, Med Phys & Biophys, D-37073 Gottingen, Germany
关键词
small fields; response function; volume effect; convolution kernel; dosimetry; OUTPUT FACTOR MEASUREMENT; MONTE-CARLO; ENERGY-DEPENDENCE; REFERENCE DOSIMETRY; SPATIAL RESPONSE; FILM DOSIMETRY; SIZE; DECONVOLUTION; CONVOLUTION; MV;
D O I
10.1088/0031-9155/60/24/9421
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aims at the experimental determination of the detector-specific 1D lateral dose response function K(x) and of its associated rotational symmetric counterpart K(r) for a set of high-resolution detectors presently used in narrow-beam photon dosimetry. A combination of slit-beam, radiochromic film, and deconvolution techniques served to accomplish this task for four detectors with diameters of their sensitive volumes ranging from 1 to 2.2 mm. The particular aim of the experiment was to examine the existence of significant negative portions of some of these response functions predicted by a recent Monte-Carlo-simulation (Looe et al 2015 Phys. Med. Biol. 60 6585-607). In a 6 MV photon slit beam formed by the Siemens Artiste collimation system and a 0.5 mm wide slit between 10 cm thick lead blocks serving as the tertiary collimator, the true cross-beam dose profile D(x) at 3 cm depth in a large water phantom was measured with radiochromic film EBT3, and the detector-affected cross-beam signal profiles M(x) were recorded with a silicon diode, a synthetic diamond detector, a miniaturized scintillation detector, and a small ionization chamber. For each detector, the deconvolution of the convolution integral M(x) = K(x) * D(x) served to obtain its specific 1D lateral dose response function K(x), and K(r) was calculated from it. Fourier transformations and back transformations were performed using function approximations by weighted sums of Gaussian functions and their analytical transformation. The 1D lateral dose response functions K(x) of the four types of detectors and their associated rotational symmetric counterparts K(r) were obtained. Significant negative curve portions of K(x) and K(r) were observed in the case of the silicon diode and the diamond detector, confirming the Monte-Carlobased prediction (Looe et al 2015 Phys. Med. Biol. 60 6585-607). They are typical for the perturbation of the secondary electron field by a detector with enhanced electron density compared with the surrounding water. In the cases of the scintillation detector and the small ionization chamber, the negative curve portions of K(x) practically vanish. It is planned to use the measured functions K(x) and K(r) to deconvolve clinical narrow-beam signal profiles and to correct the output factor values obtained with various high-resolution detectors.
引用
收藏
页码:9421 / 9436
页数:16
相关论文
共 53 条
[11]   The effect of detector size to the broadening of the penumbra - A computer simulated study [J].
Chang, KS ;
Yin, FF ;
Nie, KW .
MEDICAL PHYSICS, 1996, 23 (08) :1407-1411
[12]   The use of deconvolution and total least squares in recovering a radiation detector line spread function [J].
Charland, P ;
El-Khatib, E ;
Wolters, J .
MEDICAL PHYSICS, 1998, 25 (02) :152-160
[13]   The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambers [J].
Crop, F. ;
Reynaert, N. ;
Pittomvils, G. ;
Paelinck, L. ;
De Wagter, C. ;
Vakaet, L. ;
Thierens, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :2951-2969
[14]   Dosimetric characteristics of an unshielded p-type Si diode: linearity, photon energy dependence and spatial resolution [J].
Djouguela, Armand ;
Griessbach, Irmgard ;
Harder, Dietrich ;
Kollhoff, Ralf ;
Chofor, Ndimofor ;
Ruehmann, Antje ;
Willborn, Kay ;
Poppe, Bjoern .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2008, 18 (04) :301-306
[15]   Using cavity theory to describe the dependence on detector density of dosimeter response in non-equilibrium small fields [J].
Fenwick, John D. ;
Kumar, Sudhir ;
Scott, Alison J. D. ;
Nahum, Alan E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (09) :2901-2923
[16]   Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system [J].
Francescon, P. ;
Kilby, W. ;
Satariano, N. ;
Cora, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12) :3741-3758
[17]   Calculation of kQclin,Qmsrfclin,fmsr for several small detectors and for two linear accelerators using Monte Carlo simulations [J].
Francescon, P. ;
Cora, S. ;
Satariano, N. .
MEDICAL PHYSICS, 2011, 38 (12) :6513-6527
[18]   Experimental determination of the convolution kernel for the study of the spatial response of a detector [J].
Garcia-Vicente, F ;
Delgado, JM ;
Peraza, C .
MEDICAL PHYSICS, 1998, 25 (02) :202-207
[19]   Exact analytical solution of the convolution integral equation for a general profile fitting function and Gaussian detector kernel [J].
García-Vicente, F ;
Delgado, JM ;
Rodríguez, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03) :645-650
[20]   A convolution model for obtaining the response of an ionization chamber in static non standard fields [J].
Gonzalez-Castano, D. M. ;
Brualla Gonzalez, L. ;
Gago-Arias, M. A. ;
Pardo-Montero, J. ;
Gomez, F. ;
Luna-Vega, V. ;
Sanchez, M. ;
Lobato, R. .
MEDICAL PHYSICS, 2012, 39 (01) :482-491