Experimental determination of the lateral dose response functions of detectors to be applied in the measurement of narrow photon-beam dose profiles

被引:21
作者
Poppinga, D. [1 ]
Meyners, J. [1 ]
Delfs, B. [1 ]
Muru, A. [1 ]
Harder, D. [2 ]
Poppe, B. [1 ]
Looe, H. K. [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Med Campus Pius Hosp, Univ Clin Med Radiat Phys, Oldenburg, Germany
[2] Univ Gottingen, Med Phys & Biophys, D-37073 Gottingen, Germany
关键词
small fields; response function; volume effect; convolution kernel; dosimetry; OUTPUT FACTOR MEASUREMENT; MONTE-CARLO; ENERGY-DEPENDENCE; REFERENCE DOSIMETRY; SPATIAL RESPONSE; FILM DOSIMETRY; SIZE; DECONVOLUTION; CONVOLUTION; MV;
D O I
10.1088/0031-9155/60/24/9421
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aims at the experimental determination of the detector-specific 1D lateral dose response function K(x) and of its associated rotational symmetric counterpart K(r) for a set of high-resolution detectors presently used in narrow-beam photon dosimetry. A combination of slit-beam, radiochromic film, and deconvolution techniques served to accomplish this task for four detectors with diameters of their sensitive volumes ranging from 1 to 2.2 mm. The particular aim of the experiment was to examine the existence of significant negative portions of some of these response functions predicted by a recent Monte-Carlo-simulation (Looe et al 2015 Phys. Med. Biol. 60 6585-607). In a 6 MV photon slit beam formed by the Siemens Artiste collimation system and a 0.5 mm wide slit between 10 cm thick lead blocks serving as the tertiary collimator, the true cross-beam dose profile D(x) at 3 cm depth in a large water phantom was measured with radiochromic film EBT3, and the detector-affected cross-beam signal profiles M(x) were recorded with a silicon diode, a synthetic diamond detector, a miniaturized scintillation detector, and a small ionization chamber. For each detector, the deconvolution of the convolution integral M(x) = K(x) * D(x) served to obtain its specific 1D lateral dose response function K(x), and K(r) was calculated from it. Fourier transformations and back transformations were performed using function approximations by weighted sums of Gaussian functions and their analytical transformation. The 1D lateral dose response functions K(x) of the four types of detectors and their associated rotational symmetric counterparts K(r) were obtained. Significant negative curve portions of K(x) and K(r) were observed in the case of the silicon diode and the diamond detector, confirming the Monte-Carlobased prediction (Looe et al 2015 Phys. Med. Biol. 60 6585-607). They are typical for the perturbation of the secondary electron field by a detector with enhanced electron density compared with the surrounding water. In the cases of the scintillation detector and the small ionization chamber, the negative curve portions of K(x) practically vanish. It is planned to use the measured functions K(x) and K(r) to deconvolve clinical narrow-beam signal profiles and to correct the output factor values obtained with various high-resolution detectors.
引用
收藏
页码:9421 / 9436
页数:16
相关论文
共 53 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]  
[Anonymous], THESIS
[3]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[4]   Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams [J].
Azangwe, Godfrey ;
Grochowska, Paulina ;
Georg, Dietmar ;
Izewska, Joanna ;
Hopfgartner, Johannes ;
Lechner, Wolfgang ;
Andersen, Claus E. ;
Beierholm, Anders R. ;
Helt-Hansen, Jakob ;
Mizuno, Hideyuki ;
Fukumura, Akifumi ;
Yajima, Kaori ;
Gouldstone, Clare ;
Sharpe, Peter ;
Meghzifene, Ahmed ;
Palmans, Hugo .
MEDICAL PHYSICS, 2014, 41 (07)
[5]  
Bassinet C, 2013, MED PHYS, V40
[6]   Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams [J].
Bednarz, G ;
Huq, MS ;
Rosenow, UF .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20) :3643-3649
[7]   Dosimetric characterization of the Exradin W1 plastic scintillator detector through comparison with an in-house developed scintillator system [J].
Beierholm, A. R. ;
Behrens, C. F. ;
Andersen, C. E. .
RADIATION MEASUREMENTS, 2014, 69 :50-56
[8]   Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo study [J].
Benmakhlouf, Hamza ;
Sempau, Josep ;
Andreo, Pedro .
MEDICAL PHYSICS, 2014, 41 (04)
[9]  
BRAHME A, 1981, STRAHLENTHER ONKOL, V157, P258
[10]   An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamlets [J].
Capote, R ;
Sánchez-Doblaclo, F ;
Leal, A ;
Lagares, JI ;
Arráns, R ;
Hartmann, GH .
MEDICAL PHYSICS, 2004, 31 (09) :2416-2422