Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system

被引:32
作者
Aldelaijan, Saad [1 ,2 ,3 ,4 ]
Devic, Slobodan [1 ,2 ,5 ]
机构
[1] Jewish Gen Hosp, Radiat Oncol Dept, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Med Phys Unit, Montreal, PQ H4A 3J1, Canada
[3] Montreal Neurol Inst, Biol & Biomed Engn Dept, Montreal, PQ H3A 2B4, Canada
[4] King Faisal Specialist Hosp & Res Ctr, Biomed Phys Dept, Riyadh, Saudi Arabia
[5] McGill Univ, Jewish Gen Hosp, Segal Canc Ctr, Montreal, PQ, Canada
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2018年 / 49卷
基金
加拿大自然科学与工程研究理事会;
关键词
Radiochromic film dosimetry system; Response function; PROTOCOL;
D O I
10.1016/j.ejmp.2018.05.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Different dose response functions of EBT3 model GafChromic (TM) film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation. Methods: Pieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis. Results: Response functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy. Conclusion: Although reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy.
引用
收藏
页码:112 / 118
页数:7
相关论文
共 28 条
[1]   Use of a control film piece in radiochromic film dosimetry [J].
Aldelaijan, Saad ;
Alzorkany, Faisal ;
Moftah, Belal ;
Buzurovic, Ivan ;
Seuntjens, Jan ;
Tomic, Nada ;
Devic, Slobodan .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (01) :202-207
[2]   Radiochromic film dosimetry of HDR 192Ir source radiation fields [J].
Aldelaijan, Saad ;
Mohammed, Huriyyah ;
Tomic, Nada ;
Liang, Li-Heng ;
DeBlois, Francois ;
Sarfehnia, Arman ;
Abdel-Rahman, Wamied ;
Seuntjens, Jan ;
Devic, Slobodan .
MEDICAL PHYSICS, 2011, 38 (11) :6074-6083
[3]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[4]   The use of a reflective scanner to study radiochromic film response [J].
Alva, H ;
Mercado-Uribe, H ;
Rodríguez-Villafuerte, M ;
Brandan, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :2925-2933
[5]   Improving the energy response of external beam therapy (EBT) GafChromic™ dosimetry films at low energies (≤100 keV) [J].
Bekerat, H. ;
Devic, S. ;
DeBlois, F. ;
Singh, K. ;
Sarfehnia, A. ;
Seuntjens, J. ;
Shih, Shelley ;
Yu, Xiang ;
Lewis, D. .
MEDICAL PHYSICS, 2014, 41 (02)
[6]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[7]   Absorption spectra variations of EBT radiochromic film from radiation exposure [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) :N135-N140
[8]   Post-irradiation colouration of Gafchromic EBT radiochromic film [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) :N281-N285
[9]   Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies [J].
Chiu-Tsao, ST ;
Ho, Y ;
Shankar, R ;
Wang, L ;
Harrison, LB .
MEDICAL PHYSICS, 2005, 32 (11) :3350-3354
[10]   Calibrating page sized Gafchromic EBT3 films [J].
Crijns, W. ;
Maes, F. ;
van der Heide, U. A. ;
Van den Heuvel, F. .
MEDICAL PHYSICS, 2013, 40 (01)