Measurement of platelet aggregation in ovine blood using a new impedance aggregometer

被引:20
|
作者
Baumgarten, Andrea
Wilhelmi, Mathias [2 ]
Kalbantner, Kerstin
Ganter, Martin [3 ]
Mischke, Reinhard [1 ]
机构
[1] Tierarztliche Hsch Hannover, Klin Kleintiere, Hannover Sch Vet Med, D-30173 Hannover, Germany
[2] Hannover Med Sch, Clin Cardiac Thorac Transplant & Vasc Surg, D-3000 Hannover, Germany
[3] Hannover Sch Vet Med, Clin Small Ruminants, D-3000 Hannover, Germany
关键词
ADP; collagen; impedance aggregometry; platelet aggregation; sheep; WHOLE-BLOOD; SPECIES VARIABILITY; MODEL; SHEEP; DEVICE; SHOCK; CLOT;
D O I
10.1111/j.1939-165X.2009.00198.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Whole blood platelet aggregometry (impedance) is an important method to investigate platelet function disorders. Examination of hemostatic function in sheep is important with respect to their role as an animal model of human disease. Objective The aim of this study was to evaluate and optimize selected methodological aspects (anticoagulant, agonist concentration) of impedance aggregometry in ovine blood using the new Multiplate 5.0 analyzer. Methods Blood samples were collected in hirudin anticoagulant from 40 clinically healthy sheep. Samples from selected sheep were collected in citrate, with or without the addition of calcium chloride. The agonists adenosine diphosphate (ADP), collagen, ristocetin, arachidonic acid, and thrombin receptor-activating peptide (TRAP) were added in several concentrations to induce aggregation. Results Based on maximum aggregation values and internal precision, no significant difference was found between ADP concentrations of 3-10 mu mol/L and collagen concentrations of 3-5 mu g/mL (P >.05). The lowest interindividual variation of approximately 3-4-fold was seen with 4 and 5 mu mol/L ADP and 4 and 5 mu g/mL collagen. Ristocetin, arachidonic acid, and TRAP did not induce significant aggregation at any concentration. Aggregation results were significantly lower when measured in citrate- vs hirudin-anticoagulated blood, regardless of the presence of calcium chloride. Conclusions Our results indicate that the multiplate impedance aggregometer is suitable for the measurement of platelet aggregation in sheep using optimal agonist concentrations of 4-5 mu mol/L ADP and 4-5 mu g/mL collagen. Hirudin-anticoagulated blood is the preferred sample material.
引用
收藏
页码:149 / 156
页数:8
相关论文
共 50 条
  • [1] Measurement of platelet function in dogs using a novel impedance aggregometer
    Kalbantner, K.
    Baumgarten, A.
    Mischke, R.
    VETERINARY JOURNAL, 2010, 185 (02): : 144 - 151
  • [2] COMPARISON OF CITRATED AND HIRUDINATED BLOOD ON WHOLE BLOOD PLATELET AGGREGATION USING THE CHRONO-LOG IMPEDANCE AGGREGOMETER.
    Shipley, E. A.
    Hogan, D. F.
    Green, H. W., III
    Fiakpui, N. N.
    Magee, A. N.
    Sederquist, K. A.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2011, 25 (03) : 697 - 698
  • [3] Evaluation of platelet functions using impedance aggregometer
    Isiksacan, N.
    Koser, M.
    Cemsitoglu, F.
    Kucuksezer, U. C.
    Bakir, I.
    Deniz, G.
    IMMUNOLOGY, 2012, 137 : 525 - 525
  • [4] A study of whole blood platelet and white cell aggregation using a laser flow aggregometer
    Sun, J
    Abel, EW
    Bancroft, A
    McLaren, M
    Belch, JJF
    PLATELETS, 2003, 14 (02) : 103 - 108
  • [5] Role of nitric oxide (NO) on whole blood aggregation measured by impedance aggregometer
    Yoshimoto, H
    Suehiro, A
    Marumo, M
    Matsuda, T
    Higasa, S
    Kakishita, E
    THROMBOSIS AND HAEMOSTASIS, 1997, : P3041 - P3041
  • [7] Reference Values of Platelet Aggregation in Impedance Aggregometry with Adenosine Diphosphoric Acid on Aggregometer Multiplate
    Petrova, O. V.
    Shashin, S. A.
    Tarasov, D. G.
    SOVREMENNYE TEHNOLOGII V MEDICINE, 2016, 8 (03) : 100 - 103
  • [8] Evaluation of platelet adhesive function and aggregation using the filter aggregometer.
    Pareti, FI
    Bressi, C
    Castaman, G
    Mannucci, PM
    Ruggeri, ZM
    BLOOD, 1995, 86 (10) : 3608 - 3608
  • [9] LASER AGGREGOMETER IN STUDIES OF PLATELET-AGGREGATION
    SAMAL, AB
    CHERENKEVICH, SN
    KHMARA, NF
    LABORATORNOE DELO, 1987, (09): : 666 - 669
  • [10] DIFFICULTIES IN INTERPRETATION OF PLATELET-AGGREGATION IN TOTAL BLOOD VIA MEASUREMENT OF IMPEDANCE
    NGUYEN, P
    BEDUCHAUD, MP
    PROULT, M
    CORNILLET, P
    POTRON, G
    ANNALES DE BIOLOGIE CLINIQUE, 1988, 46 (07) : 621 - 621