Objectives: The aim of this study was to compare the diagnostic quality of a new wireless handheld unit (ADX4000; Dexcowin Co. Ltd, Korea) on conventional bitewings and its LCD screen for the detection of approximal caries in primary teeth. Methods: In total, 108 approximal surfaces of primary teeth were examined in vitro by 3 observers. Conventional films were viewed under subdued lighting conditions on a conventional view box. Digital 3.5 inch images were displayed on the built-in monitor of the ADX4000 and digital 17 inch images were viewed on a 17 inch monitor. The true caries diagnosis was based on histological assessment of the approximal surfaces after sectioning the primary teeth. Receiver operating characteristic (ROC) curve (A(z)) analysis was used to assess the diagnostic quality of imaging modalities. Results: The areas under the ROC curves ranged from 0.786 (digital 17 inch) to 0.813 (digital 3.5 inch). No statistically significant differences were found between the three modalities for detecting approximal caries. Conclusions: It was concluded that the diagnostic quality of conventional film and digital images, which were exposed and viewed by a new wireless handheld unit, was comparable. Dentomaxillofacial Radiology (2010) 39, 91-94. doi: 10.1259/dmfr/15182314
引用
收藏
页码:91 / 94
页数:4
相关论文
共 17 条
[1]   Performance of RVGui sensor and Kodak Ektaspeed Plus film for proximal caries detection [J].
Abreu, M ;
Mol, A ;
Ludlow, JB .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2001, 91 (03) :381-385
[2]   In vitro comparison of conventional film and direct digital imaging in the detection of approximal caries [J].
Castro, V. M. ;
Katz, J. O. ;
Hardman, P. K. ;
Glaros, A. G. ;
Spencer, P. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (03) :138-142
[3]   Factors affecting the diagnostic quality of bitewing radiographs: a review [J].
Chadwick, BL ;
Dummer, PHM .
BRITISH DENTAL JOURNAL, 1998, 184 (02) :80-84
[4]   RECEIVER OPERATING CHARACTERISTICS OF RADIO VISIOGRAPHY [J].
DAGENAIS, ME ;
CLARK, BG .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1995, 79 (02) :238-245
[5]   Conventional, digital and contrast-enhanced bitewing radiographs in the decision to restore approximal carious lesions [J].
Haak, R ;
Wicht, MJ ;
Noack, MJ .
CARIES RESEARCH, 2001, 35 (03) :193-199
[6]  
Hellen-Halme Kristina, 2007, Swed Dent J Suppl, P9
[7]   Influence of number of surfaces and observers on statistical power in a multiobserver ROC radiographic caries detection study [J].
Hintze, H ;
Frydenberg, M ;
Wenzel, A .
CARIES RESEARCH, 2003, 37 (03) :200-205
[8]   Influence of the validation method on diagnostic accuracy for caries. A comparison of six digital and two conventional radiographic systems [J].
Hintze, H ;
Wenzel, A .
DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (01) :44-49
[9]   End-user survey for digital sensor characteristics: a pilot questionnaire study [J].
Li, G. ;
van der Stelt, P. F. ;
Verheij, J. G. C. ;
Speller, R. ;
Galbiati, A. ;
Psomadellis, F. ;
Turchetta, R. ;
Theodoridis, S. ;
Hall, G. ;
Avset, B. S. ;
Triantis, F. A. ;
Longo, R. .
DENTOMAXILLOFACIAL RADIOLOGY, 2006, 35 (03) :147-151
[10]   THE EFFECT OF CATHODE-RAY TUBE DISPLAY FORMAT ON OBSERVER PERFORMANCE IN DENTAL DIGITIZED RADIOGRAPHY - COMPARISON WITH PLAIN FILMS [J].
MOYSTAD, A ;
SVANAES, DB ;
LARHEIM, TA ;
GRONDAHL, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1994, 23 (04) :206-210