Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma

被引:11
作者
Zhang, Qiaoxuan [1 ,2 ]
Zhang, Lu [2 ]
Lin, Haibiao [1 ,2 ]
Cai, Zhiliang [2 ]
Yan, Jun [1 ]
Wang, Qiqin [3 ,4 ]
Han, Liqiao [1 ]
Wang, Jianbing [1 ]
Ke, Peifeng [1 ]
Zhuang, Junhua [1 ]
Huang, Xianzhang [1 ]
机构
[1] Guangzhou Univ Chinese Med, Dept Lab Med, Affiliated Hosp 2, Guangzhou, Guangdong, Peoples R China
[2] Guangzhou Univ Chinese Med, Clin Med Coll, Guangzhou, Guangdong, Peoples R China
[3] Jinan Univ, Dept Pharm, Guangzhou 510632, Guangdong, Peoples R China
[4] Jinan Univ, Guangdong Prov Key Lab Phannacodynam Constituents, Guangzhou 510632, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
17; alpha-Hydroxyprogesterone; Reference measurement procedure; Tandem mass spectrometry; Clinical laboratory measurements; CONGENITAL ADRENAL-HYPERPLASIA; QUALITY SPECIFICATIONS; HUMAN SERUM; 17-HYDROXYPROGESTERONE;
D O I
10.1007/s00216-019-02086-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A candidate reference measurement procedure (RMP) based on isotope dilution coupled with liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) was developed and validated for the quantification of 17 alpha-hydroxyprogesterone (17-OHP) in human plasma. 17-OHP spiked with a deuterium-labeled internal standard was extracted from plasma by liquid-liquid extraction with 1 mL n-hexane/ethyl acetate (3:2, v/v). Reversed-phase chromatography and positive electrospray ionization were used in the ID-LC-MS/MS. Gradient elution coupled with use of a C-18-packed ultrahigh-performance liquid chromatography column allowed complete baseline resolution of 17-OHP from its structural analogue desoxycorticosterone in 6 min. To determine the 17-OHP level in human plasma, a bracketing calibration method was used to give higher accuracy and precision. The limit of detection and the lower limit of the measuring interval for the candidate RMP were 2.1 pg/mL (6.4 pmol/L) and 4.6 pg/mL (13.9 pmol/L), respectively. Extraction recovery was determined to be (96.08 +/- 3.03)% (n = 3). Imprecision (intra-assay and interassay) was 4.03% or less at 0.83, 15.19, 64.22, and 313.46 ng/mL (2.51, 45.97, 194.34, and 948.56 nmol/L, respectively). Recoveries ranged from 98.05% to 102.24%. When comparing our RMP results with those obtained with an established RMP via International Federation of Clinical Chemistry and Laboratory Medicine external quality assessment scheme for reference laboratories in laboratory medicine (RELA) samples, we found that the biases ranged from -1.99% to 3.08% against the targets. No interference was observed, and the linear response ranged from 0.47 to 958.63 ng/mL (1.42 to 2900.90 nmol/L). Moreover, the candidate RMP was used to measure the concentration of 17-OHP in human plasma and was compared with an immunoassay using 40 plasma samples. The performance of the method meets the needs of an RMP (total coefficient of variation of 5% or less and bias of 3.08% or less). This method can be used for reference material value assignment of 17-OHP in human plasma matrix. It could also serve as an accurate reference baseline for routine methods to increase the accuracy and precision of certain clinical laboratory measurements.
引用
收藏
页码:7095 / 7104
页数:10
相关论文
共 31 条
[1]  
[Anonymous], 2019, 17511 ISO
[2]   Evaluation of an Isotope Dilution HPLC Tandem Mass Spectrometry Candidate Reference Measurement Procedure for Total 17-β Estradiol in Human Serum [J].
Botelho, Julianne Cook ;
Ribera, Ashley ;
Cooper, Hans C. ;
Vesper, Hubert W. .
ANALYTICAL CHEMISTRY, 2016, 88 (22) :11123-11129
[3]  
Clinical and Laboratory Standards Institute, 2014, C62A CLIN LAB STAND
[4]  
Clinical and Laboratory Standards Institute, 2003, EP06 CLIN LAB STAND
[5]  
Clinical and Laboratory Standards Institute, 2014, EP10A3 CLIN LAB STAN
[6]   Pilot proficiency testing study for second tier congenital adrenal hyperplasia newborn screening [J].
De Jesus, Victor R. ;
Simms, David A. ;
Schiffer, Jarad ;
Kennedy, Meredith ;
Mei, Joanne V. ;
Hannon, W. Harry .
CLINICA CHIMICA ACTA, 2010, 411 (21-22) :1684-1687
[7]   Clinical determination of 17-hydroxyprogesterone in serum by LC-MS/MS:: Comparison to Coat-A-Count™ RIA method [J].
Etter, Michele L. ;
Eichhorst, Jeff ;
Lehotay, Denis C. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 840 (01) :69-74
[8]   False positive rate in newborn screening for congenital adrenal hyperplasia (CAH)-ether extraction reveals two distinct reasons for elevated 17α-hydroxyprogesterone (17-OHP) values [J].
Fingerhut, Ralph .
STEROIDS, 2009, 74 (08) :662-665
[9]  
Fraser CG, 1999, SCAND J CLIN LAB INV, V59, P477
[10]   Basal levels of 17-hydroxyprogesterone can distinguish children with isolated precocious pubarche [J].
Grandone, Anna ;
Marzuillo, Pierluigi ;
Luongo, Caterina ;
Toraldo, Roberto ;
Mariani, Michela ;
del Giudice, Emanuele Miraglia ;
Perrone, Laura .
PEDIATRIC RESEARCH, 2018, 84 (04) :533-536