Analysis of a surface imaging system using a six degree-of-freedom couch

被引:5
作者
Zhao, Xiaodong [1 ]
Covington, Elizabeth L. [1 ]
Popple, Richard A. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, 1719 6th Ave S, Birmingham, AL 35294 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2022年 / 23卷 / 08期
关键词
six degree-of-freedom couch; stereotactic radiosurgery; surface-guided radiotherapy; surface imaging; INSPIRATION BREATH-HOLD; CANCER PATIENTS; RADIATION; RADIOTHERAPY; ACCURACY; SETUP;
D O I
10.1002/acm2.13697
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To validate surface imaging (SI)-reported offsets using a six degree-of-freedom couch and an anthropomorphic phantom for commissioning and routine quality assurance of an SI system used for stereotactic radiosurgery (SRS). Methods An anthropomorphic phantom with a radiopaque ball bearing (BB) placed either anterior, midline, or posterior, was tracked with SI with a typical SRS region of interest. Couch motion in all six degrees of freedom was programmed and delivered on a linac. SI system logs were synchronized with linac trajectory logs. Ten random couch positions were selected at couch 0 degrees, 45 degrees, 90 degrees, 270 degrees, 315 degrees with megavolt (MV) images taken to account for couch walkout. The SI residual error (epsilon), the difference between SI reported offset and MV or trajectory log position, was calculated. Residual errors were measured with and without one SI pod blocked. Results The median [range] of magnitude of translational epsilon was 0.13 [0.07, 0.21], 0.16 [0.11, 0.26], 0.61 [0.50, 0.68], 0.49 [0.42, 0.55], 0.55 [0.38, 0.72] mm for couch rotations of 0 degrees, 45 degrees, 90 degrees, 270 degrees, 315 degrees, respectively, for the midline BB and no pod blocked. The range of all translational epsilon from all couch angles (with and without pod block) at different BB positions is [0.05, 0.96] mm. The absolute range of difference when changing BB position when no pod is blocked in median translational epsilon is [0.01, 0.40] mm with the maximum at BB posterior. The absolute range of difference when not changing BB positions with and without pod block in median translational epsilon is [0.01, 0.37] mm with the maximum at BB posterior and couch 315 degrees. Conclusion SI system and linac trajectory log analysis can be used to assess SI system performance with automated couch motion to validate SI accuracy.
引用
收藏
页数:7
相关论文
共 21 条
  • [1] Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation Therapy
    Alderliesten, Tanja
    Sonke, Jan-Jakob
    Betgen, Anja
    Honnef, Joeri
    van Vliet-Vroegindeweij, Corine
    Remeijer, Peter
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 536 - 542
  • [2] A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup
    Bert, C
    Metheany, KG
    Doppke, K
    Chen, GTY
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 2753 - 2762
  • [3] Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging
    Betgen, Anja
    Alderliesten, Tanja
    Sonke, Jan-Jakob
    van Vliet-Vroegindeweij, Corine
    Bartelink, Harry
    Remeijer, Peter
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 106 (02) : 225 - 230
  • [4] Stereotactic radiosurgery with MLC-defined arcs: Verification of dosimetry, spatial accuracy, and end-to-end tests
    Brezovich, Ivan A.
    Wu, Xingen
    Popple, Richard A.
    Covington, Elizabeth
    Cardan, Rex
    Shen, Sui
    Fiveash, John
    Bredel, Markus
    Guthrie, Barton
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 84 - 98
  • [5] Evaluation of radiotherapy setup accuracy for head and neck cancer using a 3-D surface imaging system
    Cho, H. -L.
    Park, E. -T.
    Kim, J. -Y.
    Kwak, K. -S.
    Kim, C. -J.
    Ahn, K. -J.
    Suh, T. -S.
    Lee, Y. -K.
    Kim, S. -W.
    Kim, J. -K.
    Lim, S.
    Choi, Y. -M.
    Park, S. -K.
    [J]. JOURNAL OF INSTRUMENTATION, 2013, 8
  • [6] A Low-Cost Method to Assess the Performance of Surface Guidance Imaging Systems at Non-Zero Couch Angles
    Covington, Elizabeth L.
    Popple, Richard A.
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (04)
  • [7] Surface guided imaging during stereotactic radiosurgery with automated delivery
    Covington, Elizabeth L.
    Stanley, Dennis N.
    Fiveash, John B.
    Thomas, Evan M.
    Marcrom, Samuel R.
    Bredel, Marcus
    Willey, Christopher D.
    Riley, Kristen O.
    Popple, Richard A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (12): : 90 - 95
  • [8] The Role of Optical Surface Imaging Systems in Radiation Therapy
    Hoisak, Jeremy D. P.
    Pawlicki, Todd
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2018, 28 (03) : 185 - 193
  • [9] Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence
    Kang, Hyejoo
    Patel, Rakesh
    Roeske, John C.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 127 - 133
  • [10] Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy
    Kugele, Malin
    Edvardsson, Anneli
    Berg, Lovisa
    Alkner, Sara
    Ljus, Carina Andersson
    Ceberg, Sofie
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 25 - 38