Design of Noninvasive Hemodynamic Monitoring Equipment Using Impedance Cardiography

被引:0
作者
Vu Duy Hai [1 ]
Lai Huu Phuong Trung [1 ]
Pham Manh Hung [1 ]
Dao Viet Hung [1 ]
Pham Phuc Ngoc [1 ]
Phan Dang Hung [1 ]
Chu Quang Dan [1 ]
Tran Quoc Vi [2 ]
机构
[1] Hanoi Univ Sci & Technol, Biomed Elect Ctr, Hanoi, Vietnam
[2] Vinh Phuc Gen Hosp, Vinh Phuc, Vietnam
来源
7TH INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM (BME7): TRANSLATIONAL HEALTH SCIENCE AND TECHNOLOGY FOR DEVELOPING COUNTRIES | 2020年 / 69卷
关键词
Noninvasive hemodynamic monitor; Impedance cardiography; Stroke volume; Cardiac output; Thoracic fluid content; Left ventricular ejection time;
D O I
10.1007/978-981-13-5859-3_1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hemodynamic parameters play a very important role in supporting physicians to diagnose and treat cardiovascular diseases. This paper presents the results of the study on designing of noninvasive hemodynamic monitoring equipment using impedance cardiography (ICG). This is a measurement method that shows many advantages and is interested in the present. The equipment has ability of measurement and monitoring hemodynamic parameters include: heart rate HR; thoracic fluid content TFC; left ventricular ejection time LVET; stroke volume SV; cardiac output CO and ICG waveform. Results of measurement between the research equipment BK-NICO and the standard equipment of NICCOMO for 104 patients show that the hemodynamic parameters measured by the two equipments are quite small difference. Specifically, the average deviation of the parameters obtained, HR is 0.16 bpm, TFC is 0.73 1/k Omega, LVET is 5.95 ms, SV is 0.17 ml, CO is 0.01 lpm. With the results of the research, the device is fully capable of deploying the application into practice in the hospitals to replace the current invasive measurement method.
引用
收藏
页码:3 / 9
页数:7
相关论文
共 50 条
  • [21] Design of Measurement Device for Impedance Cardiography
    Snajdarova, Michaela
    Borik, Stefan
    Cap, Ivo
    PROCEEDINGS OF 2017 18TH INTERNATIONAL CONFERENCE ON COMPUTATIONAL PROBLEMS OF ELECTRICAL ENGINEERING (CPEE), 2017,
  • [22] Non-invasive cuffless blood pressure and heart rate monitoring using impedance cardiography
    Ghosh, Sudipta
    Chattopadhyay, Bhabani Prasad
    Roy, Ram Mohan
    Mukherjee, Jayanta
    Mahadevappa, Manjunatha
    INTELLIGENT MEDICINE, 2022, 2 (04): : 199 - 208
  • [23] Impedance cardiography to assess hemodynamic status: a comparison with transpulmonary thermodilution
    A Donati
    R Nardella
    V Gabbanelli
    C Valentini
    F Gabriele
    P Pelaia
    Critical Care, 11 (Suppl 2):
  • [24] Hemodynamic Patterns Identified by Impedance Cardiography Predict Mortality in the General Population: The PREVENCION Study
    Medina-Lezama, Josefina
    Narvaez-Guerra, Offdan
    Herrera-Enriquez, Karela
    Morey-Vargas, Oscar L.
    Bolanos-Salazar, Juan Francisco
    Abugattas, Juan P.
    Zea-Diaz, Humberto
    Chirinos-Revilla, Jose L.
    Fernandez-Sivincha, Jenny G.
    Delgado-Lazo, Victor
    Chirinos, Diana A.
    Townsend, Raymond R.
    Chirinos, Julio A.
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (18):
  • [25] Utility of Impedance Cardiography for the Detection of Hemodynamic Changes in Stable Patients With Sickle Cell Disease
    Das, Bibhuti B.
    Raj, Ashok
    Recto, Michael
    Kong, Maiying
    Bertolone, Salvatore
    JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2012, 34 (05) : 336 - 339
  • [26] Noninvasive continuous hemodynamic monitoring
    Jasper Truijen
    Johannes J. van Lieshout
    Wilbert A. Wesselink
    Berend E. Westerhof
    Journal of Clinical Monitoring and Computing, 2012, 26 : 267 - 278
  • [27] Electrical impedance cardiography using artificial neural networks
    Mulavara, AP
    Timmons, WD
    Nair, MS
    Gupta, V
    Kumar, AAR
    Taylor, BC
    ANNALS OF BIOMEDICAL ENGINEERING, 1998, 26 (04) : 577 - 583
  • [28] Motion Artifacts Reduction for Noninvasive Hemodynamic Monitoring of Conscious Patients Using Electrical Impedance Tomography: A Preliminary Study
    Dang, Thi Hang
    Jang, Geuk Young
    Lee, Kyounghun
    Oh, Tong In
    SENSORS, 2023, 23 (11)
  • [29] Continuous measurement of hemodynamic alterations during pharmacologic cardiovascular stress using automated impedance cardiography
    Scherhag, AW
    Pfleger, S
    deMey, C
    Schreckenberger, AB
    Staedt, U
    Heene, DL
    JOURNAL OF CLINICAL PHARMACOLOGY, 1997, 37 (01) : S21 - S28
  • [30] Impedance cardiography as tool for continuous hemodynamic monitoring during cesarean section: randomized, prospective double blind study
    D'Ambrosio, Alessandro
    Cotoia, Antonella
    Beck, Renata
    Salatto, Potito
    Zibar, Lada
    Cinnella, Gilda
    BMC ANESTHESIOLOGY, 2018, 18