Design of Noninvasive Hemodynamic Monitoring Equipment Using Impedance Cardiography

被引:0
|
作者
Vu Duy Hai [1 ]
Lai Huu Phuong Trung [1 ]
Pham Manh Hung [1 ]
Dao Viet Hung [1 ]
Pham Phuc Ngoc [1 ]
Phan Dang Hung [1 ]
Chu Quang Dan [1 ]
Tran Quoc Vi [2 ]
机构
[1] Hanoi Univ Sci & Technol, Biomed Elect Ctr, Hanoi, Vietnam
[2] Vinh Phuc Gen Hosp, Vinh Phuc, Vietnam
来源
7TH INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM (BME7): TRANSLATIONAL HEALTH SCIENCE AND TECHNOLOGY FOR DEVELOPING COUNTRIES | 2020年 / 69卷
关键词
Noninvasive hemodynamic monitor; Impedance cardiography; Stroke volume; Cardiac output; Thoracic fluid content; Left ventricular ejection time;
D O I
10.1007/978-981-13-5859-3_1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hemodynamic parameters play a very important role in supporting physicians to diagnose and treat cardiovascular diseases. This paper presents the results of the study on designing of noninvasive hemodynamic monitoring equipment using impedance cardiography (ICG). This is a measurement method that shows many advantages and is interested in the present. The equipment has ability of measurement and monitoring hemodynamic parameters include: heart rate HR; thoracic fluid content TFC; left ventricular ejection time LVET; stroke volume SV; cardiac output CO and ICG waveform. Results of measurement between the research equipment BK-NICO and the standard equipment of NICCOMO for 104 patients show that the hemodynamic parameters measured by the two equipments are quite small difference. Specifically, the average deviation of the parameters obtained, HR is 0.16 bpm, TFC is 0.73 1/k Omega, LVET is 5.95 ms, SV is 0.17 ml, CO is 0.01 lpm. With the results of the research, the device is fully capable of deploying the application into practice in the hospitals to replace the current invasive measurement method.
引用
收藏
页码:3 / 9
页数:7
相关论文
共 50 条
  • [1] Hemodynamic monitoring during hemodialysis using noninvasive impedance cardiography.
    Ahmed, Z
    Pareek, R
    Elivera, H
    Hoang, T
    Lee, J
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2003, 14 : 872A - 872A
  • [2] Bench to bedside: Electrophysiologic and clinical principles of noninvasive hemodynamic monitoring using impedance cardiography
    Summers, RL
    Shoemaker, WC
    Peacock, WF
    Ander, DS
    Coleman, TG
    ACADEMIC EMERGENCY MEDICINE, 2003, 10 (06) : 669 - 680
  • [3] Assessment of hemodynamic changes during hemodialysis using noninvasive impedance cardiography.
    Ahmed, Z
    Pareek, R
    Verma, A
    Ranganna, K
    Hoang, T
    Toh, I
    Matsumura, D
    Hasni, K
    Tachev, S
    Kancha, K
    Lee, J
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2004, 43 (04) : A15 - A15
  • [4] Noninvasive monitoring of transient cardiac changes with impedance cardiography
    Zhang, HJ
    Matonick, JP
    Li, JKJ
    PROCEEDINGS OF THE IEEE 28TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2002, : 71 - 72
  • [5] Noninvasive Monitoring of Transient Cardiac Changes with Impedance Cardiography
    Zhang, Hongjun
    Li, John K-J.
    CARDIOVASCULAR ENGINEERING, 2008, 8 (04) : 225 - 231
  • [6] The utility of impedance cardiography in hemodynamic monitoring of patients with sepsis
    Butz, James
    Shan, Yizhi
    Samayoa, Andres
    Kirton, Orlando C.
    Thai Vu
    TRAUMA SURGERY & ACUTE CARE OPEN, 2019, 4 (01)
  • [7] UTILITY OF IMPEDANCE CARDIOGRAPHY ON HEMODYNAMIC MONITORING IN SEPTIC PATIENTS
    Shan, Yizhi
    Samayoa, Andres
    Shadis, Ryan
    Yusckak, James
    Kirton, Orlando
    Vu, Thai
    CRITICAL CARE MEDICINE, 2016, 44 (12)
  • [8] Utilization of impedance cardiography in noninvasive assessment of hemodynamic status in cirrhotic patients
    Suleymanlar, Inci
    Aktas, Sevil
    Basarici, Ibrahim
    TURKISH JOURNAL OF GASTROENTEROLOGY, 2010, 21 (03): : 280 - 286
  • [9] Noninvasive cardiac monitoring in pregnancy: impedance cardiography versus echocardiography
    Burlingame, J.
    Ohana, P.
    Aaronoff, M.
    Seto, T.
    JOURNAL OF PERINATOLOGY, 2013, 33 (09) : 675 - 680
  • [10] Noninvasive cardiac monitoring in pregnancy: impedance cardiography versus echocardiography
    J Burlingame
    P Ohana
    M Aaronoff
    T Seto
    Journal of Perinatology, 2013, 33 : 675 - 680