Comparison between the Transcutaneous and Total Serum Bilirubin Measurement in Malay Neonates with Neonatal Jaundice

被引:12
|
作者
Mohamed, Mazrah [1 ]
Ibrahim, Nor Rosidah [1 ,2 ]
Ramli, Noraida [1 ,2 ]
Majid, Noorizan Abdul [1 ,2 ]
Yacob, Najib Majdi [3 ]
Nasir, Ariffin [1 ,2 ]
机构
[1] Univ Sains Malaysia, Sch Med Sci, Dept Paediat, Kubang Kerian 16150, Kelantan, Malaysia
[2] Univ Sains Malaysia, Hosp Univ Sains Malaysia, Kubang Kerian, Kelantan, Malaysia
[3] Univ Sains Malaysia, Sch Med Sci, Biostat & Res Methodol Unit, Kubang Kerian, Kelantan, Malaysia
来源
MALAYSIAN JOURNAL OF MEDICAL SCIENCES | 2022年 / 29卷 / 01期
关键词
transcutaneous bilirubin; total serum bilirubin; JM-105; bilirubinometer; neonatal jaundice; TERM; HYPERBILIRUBINEMIA; IMPLEMENTATION; NEWBORNS; ACCURACY; DEVICE;
D O I
10.21315/mjms2022.29.1.5
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: This study aims to investigate the reliability of the Drager Jaundice Meter JM-105 for the screening of neonatal jaundice in Malay neonates. Methods: A cross-sectional study was conducted in a university hospital involving 130 jaundiced neonates requiring serum bilirubin determination from day 2 to day 7 of life. Results: The mean total serum bilirubin (TSB) was 232 mu mol/L, whereas the mean transcutaneous bilirubin (TcB) measured at the forehead and sternum were 222 mu mol/L and 223 mu mol/L, respectively. Further, TcB underestimates TSB with a mean difference of 10.10 mu mol/L at the forehead and 9.27 mu mol/L at the sternum. A positive linear relationship was observed between TSB with TcB forehead (r = 0.82) and TcB sternum (r = 0.80). A good discriminations ability was observed for both the TcB forehead (receiver operating characteristics [ROC] curve = 89.8%) and sternum (ROC curve = 89.7%) at a TSB level of 205 mu mol/L. The sensitivity ranges from 84.4% to 85.3%, while the specificity ranges from 77.4% to 76.4%. Conclusion: Our study demonstrates a strong linear relationship and good diagnostic accuracy of TcB values compared to TSB values. To conclude, TcB measured at the forehead or sternum is a good alternative as a non-invasive screening tool for non-severe hyperbilirubinemia in Malay neonates.
引用
收藏
页码:43 / 54
页数:12
相关论文
共 50 条
  • [31] Should transcutaneous bilirubin be measured in preterm infants receiving phototherapy? The relationship between transcutaneous and total serum bilirubin in preterm infants with and without phototherapy
    Hulzebos, Christian V.
    Vader-van Imhoff, Deirdre E.
    Bos, Arend F.
    Dijk, Peter H.
    PLOS ONE, 2019, 14 (06):
  • [32] Optimal site for applying transcutaneous bilirubinometer as an outpatient screening tool for neonatal jaundice: a comparison between the sternum and forehead
    Zhang, Emily
    Wu, Tzong-Jin
    Hudak, Mark L.
    Yan, Ke
    Teng, Ru-Jeng
    FRONTIERS IN PEDIATRICS, 2024, 12
  • [33] Implementation of a reformulated Roche® bilirubin Gen.3 reagent did not affect the relationship between BiliChek transcutaneous and Roche total serum bilirubin
    Aguirre, Jose C. Jara
    Wockenfus, Amy M.
    Krekelberg, Benjamin J.
    Fine, Kara L.
    Cook, Walter J.
    Karon, Brad S.
    CLINICAL BIOCHEMISTRY, 2019, 63 : 148 - 151
  • [34] The use of a simple and affordable skin patch for measurement of transcutaneous bilirubin levels in neonates during phototherapy
    Kallimath, Aditya
    Patnaik, Suprabha
    Suryawanshi, Pradeep
    Deshmukh, Rupeshkumar
    Malshe, Nandini
    FRONTIERS IN PEDIATRICS, 2024, 12
  • [35] Transcutaneous Bilirubin Measurement Methods in Neonates and Its Utility for Emergency Department Use
    Yamamoto, Loren G.
    Killeen, Jeffrey
    French, Gina M.
    PEDIATRIC EMERGENCY CARE, 2012, 28 (04) : 380 - 384
  • [36] Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates
    Deepak Chawla
    Suksham Jain
    Gurjit Kaur
    Vikas Sinhmar
    Vishal Guglani
    European Journal of Pediatrics, 2014, 173 : 173 - 179
  • [37] Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates
    Chawla, Deepak
    Jain, Suksham
    Kaur, Gurjit
    Sinhmar, Vikas
    Guglani, Vishal
    EUROPEAN JOURNAL OF PEDIATRICS, 2014, 173 (02) : 173 - 179
  • [38] Comparison of Transcutaneous and Total Serum Bilirubin Measurements at Five Different Sites in Newborns before and after Phototherapy
    Eken, Seda
    Ilhan, Ozkan
    Hakan, Nilay
    KLINISCHE PADIATRIE, 2024, 236 (01): : 31 - 38
  • [39] Comparison between transcutaneous bilirubinometry and total serum bilirubin measurements in preterm infants <35 weeks gestation
    Ahmed, M.
    Mostafa, S.
    Fisher, G.
    Reynolds, T. M.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2010, 47 : 72 - 77
  • [40] Discrepancies Between Transcutaneous and Serum Bilirubin Measurements
    Taylor, James A.
    Burgos, Anthony E.
    Flaherman, Valerie
    Chung, Esther K.
    Simpson, Elizabeth A.
    Goyal, Neera K.
    Von Kohorn, Isabelle
    Dhepyasuwan, Nui
    PEDIATRICS, 2015, 135 (02) : 224 - 231