Raman chemical imaging for ingredient-specific particle size characterization of aqueous suspension nasal spray formulations: A progress report

被引:61
作者
Doub, William H. [1 ]
Adams, Wallace P.
Spencer, John A.
Buhse, Lucinda F.
Nelson, Matthew P.
Treado, Patrick J.
机构
[1] CDER, OPS, Div Pharmaceut Anal Food & Drug Adm, St Louis, MO USA
[2] US FDA, CDER, OPS, Off Gener Drugs, Rockville, MD 20857 USA
[3] ChemImage Corp, Pittsburgh, PA 15208 USA
[4] US FDA, Div Pharmaceut Anal, St Louis, MO 63101 USA
关键词
drug delivery; nasal; image analysis; spectroscopy; Raman;
D O I
10.1007/s11095-006-9211-2
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Purpose. This study was conducted to evaluate the feasibility of using Raman chemical imaging (i.e., Raman imaging microspectroscopy) to establish chemical identity, particle size and particle size distribution (PSD) for a representative corticosteroid in aqueous nasal spray suspension formulations. Materials and Methods. The Raman imaging PSD protocol was validated using polystyrene (PS) microsphere size standards (NIST-traceable). A Raman spectral library was developed for the active and inactive compounds in the formulation. Four nasal sprays formulated with beclomethasone dipropionate (BDP) ranging in size from 1.4 to 8.3 mu m were imaged by both Raman and brightfield techniques. The Raman images were then processed to calculate the PSD for each formulation. Results. Within each region examined, active pharmaceutical ingredient (API) particles are unambiguously identified and the total number of those particles, particle size and PSD of API free of excipients and PSD of API particles adhered to other excipients are reported. Conclusions. Good statistical agreement is obtained between the reported and measured sizes of the PS microspheres. BDP particles were clearly distinguishable from those of excipients. Raman chemical imaging (RCI) is able to differentiate between and identify the chemical makeup of multiple components in complex BDP sample and placebo mixtures. The Raman chemical imaging method (coupled Raman and optical imaging) shows promise as a method for characterizing particle size and shape of corticosteroid in aqueous nasal spray suspension formulations. However, rigorous validation of RCI for PSD analysis is incomplete and requires additional research effort. Some specific areas of concern are discussed.
引用
收藏
页码:934 / 945
页数:12
相关论文
共 19 条
  • [1] A RAMAN MICROSCOPY STUDY OF STRESS TRANSFER IN HIGH-PERFORMANCE EPOXY COMPOSITES REINFORCED WITH POLYETHYLENE FIBERS
    BOOGH, LCN
    MEIER, RJ
    KAUSCH, HH
    KIP, BJ
    [J]. JOURNAL OF POLYMER SCIENCE PART B-POLYMER PHYSICS, 1992, 30 (04) : 325 - 333
  • [2] LINE-SCANNED MICRO RAMAN-SPECTROSCOPY USING A COOLED CCD IMAGING DETECTOR
    BOWDEN, M
    GARDINER, DJ
    RICE, G
    GERRARD, DL
    [J]. JOURNAL OF RAMAN SPECTROSCOPY, 1990, 21 (01) : 37 - 41
  • [3] CHALMERS JM, 2002, HDB VIBRATINAL SPECT
  • [4] DANIEL WW, 1990, APPL NONPARAMETRIC S, P330
  • [5] Hyperspectral Raman line imaging of an aluminosilicate glass
    Jestel, NL
    Shaver, JM
    Morris, MD
    [J]. APPLIED SPECTROSCOPY, 1998, 52 (01) : 64 - 69
  • [6] KELLY RN, FALSE ASSUMPTIONS LA
  • [7] Levin M, 2005, AM LAB, V37, P14
  • [8] LEVIN M, 2005, LAB EQUIB, V42, P12
  • [9] Lim J. S., 1990, Two-dimensional signal and image processing
  • [10] Chemical imaging of thermoplastic olefin (TPO) surface architecture
    Morris, HR
    Turner, JF
    Munro, B
    Ryntz, RA
    Treado, PJ
    [J]. LANGMUIR, 1999, 15 (08) : 2961 - 2972