Measurements of restricted diffusion using an oscillating gradient spin-echo sequence

被引:97
作者
Schachter, M [1 ]
Does, MD
Anderson, AW
Gore, JC
机构
[1] Yale Univ, Dept Appl Phys, New Haven, CT 06520 USA
[2] Yale Univ, Dept Diagnost Radiol, New Haven, CT 06520 USA
关键词
NMR; restricted diffusion; oscillating gradient spin echo; porous media;
D O I
10.1006/jmre.2000.2203
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An oscillating gradient spin-echo (OGSE) pulse sequence was used to measure the apparent diffusion coefficient (D-app) of water in the short diffusion time regime in the presence of restrictions. The diffusion coefficients of water in a simple water sample and a water and oil mixture were measured to be the same for different periods of the gradient oscillation, as expected when there are no restriction effects. The D-app of water in the spaces between closely packed beads was also measured as a function of the gradient oscillation periods in the range 11 to 80 ms. The D-app of water in restricted systems varies with the period of the gradient oscillation and the dispersion depends on the scale of the restriction. For a sample of packed beads of diameter 9.1 +/- 0.7 mum, the pore surface-to-volume ratio was estimated experimentally by this method to be 1.3 +/- 0.1 mum(-1), corresponding to a mean pore diameter of 6.4 +/- 0.7 mum. A Monte Carlo computer simulation of the NMR OGSE signal from the spins diffusing in a system of compartments was also implemented and the D-app demonstrated similar behavior with gradient oscillation periods. (C) 2000 Academic Press.
引用
收藏
页码:232 / 237
页数:6
相关论文
共 50 条
  • [1] Systematic changes to the apparent diffusion tensor of in vivo rat brain measured with an oscillating-gradient spin-echo sequence
    Kershaw, Jeff
    Leuze, Christoph
    Aoki, Ichio
    Obata, Takayuki
    Kanno, Iwao
    Ito, Hiroshi
    Yamaguchi, Yuki
    Handa, Hiroshi
    NEUROIMAGE, 2013, 70 : 10 - 20
  • [2] Restricted diffusion characteristics in oscillating gradient spin echo with mesoscopic phantom
    Oshiro, Hinako
    Hata, Junichi
    Nakashima, Daisuke
    Oshiro, Rintaro
    Hayashi, Naoya
    Haga, Yawara
    Hagiya, Kei
    Yoshimaru, Daisuke
    Okano, Hideyuki
    HELIYON, 2024, 10 (04)
  • [3] Pulsed-gradient spin-echo monitoring of restricted diffusion in multilayered structures
    Grebenkov, Denis S.
    JOURNAL OF MAGNETIC RESONANCE, 2010, 205 (02) : 181 - 195
  • [4] Improved gradient waveforms for oscillating gradient spin-echo (OGSE) diffusion tensor imaging
    Hennel, Franciszek
    Michael, Eric Seth
    Pruessmann, Klaas P.
    NMR IN BIOMEDICINE, 2021, 34 (02)
  • [5] Oscillating Gradient Spin-Echo (OGSE) Diffusion Tensor Imaging of the Human Brain
    Baron, Corey A.
    Beaulieu, Christian
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (03) : 726 - 736
  • [6] Pulsed-Gradient Spin-Echo Monitoring of Restricted Diffusion in Multilayered Structures: Challenges and Solutions
    Grebenkov, Denis S.
    MAGNETIC RESONANCE IN POROUS MEDIA, 2011, 1330 : 65 - 68
  • [7] Time-dependent Diffusion in Brain Abscesses Investigated with Oscillating-gradient Spin-echo
    Maekawa, Tomoko
    Hori, Masaaki
    Murata, Katsutoshi
    Feiweier, Thorsten
    Kamiya, Kouhei
    Andica, Christina
    Hagiwara, Akifumi
    Fujita, Shohei
    Kamagata, Koji
    Wada, Akihiko
    Abe, Osamu
    Aoki, Shigeki
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2022, 21 (04) : 525 - 530
  • [8] Changes in the ADC of diffusion-weighted MRI with the oscillating gradient spin-echo (OGSE) sequence due to differences in substrate viscosities
    Tomoko Maekawa
    Masaaki Hori
    Katsutoshi Murata
    Thorsten Feiweier
    Issei Fukunaga
    Christina Andica
    Akifumi Hagiwara
    Koji Kamagata
    Saori Koshino
    Osamu Abe
    Shigeki Aoki
    Japanese Journal of Radiology, 2018, 36 : 415 - 420
  • [9] Impact of tissue properties on time-dependent alterations in apparent diffusion coefficient: a phantom study using oscillating-gradient spin-echo and pulsed-gradient spin-echo sequences
    Kazushige Ichikawa
    Toshiaki Taoka
    Masanori Ozaki
    Mayuko Sakai
    Hiroshi Yamaguchi
    Shinji Naganawa
    Japanese Journal of Radiology, 2022, 40 : 970 - 978
  • [10] Impact of tissue properties on time-dependent alterations in apparent diffusion coefficient: a phantom study using oscillating-gradient spin-echo and pulsed-gradient spin-echo sequences
    Ichikawa, Kazushige
    Taoka, Toshiaki
    Ozaki, Masanori
    Sakai, Mayuko
    Yamaguchi, Hiroshi
    Naganawa, Shinji
    JAPANESE JOURNAL OF RADIOLOGY, 2022, 40 (09) : 970 - 978