Evaluation of the image quality and dose reduction in digital radiography with an advanced spatial noise reduction algorithm in pediatric patients

被引:5
作者
Oh, Saelin [1 ,2 ,3 ]
Kim, Ji Hye [1 ,2 ]
Yoo, So-Young [1 ,2 ]
Jeon, Tae Yeon [1 ,2 ]
Kim, Yu Jin [1 ,2 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Dept Radiol, Samsung Med Ctr, 81 Irwon Ro, Seoul 06351, South Korea
[2] Sungkyunkwan Univ, Sch Med, Ctr Imaging Sci, Samsung Med Ctr, 81 Irwon Ro, Seoul 06351, South Korea
[3] Korea Univ, Coll Med, Dept Radiol, Anam Hosp, Seoul 02841, South Korea
关键词
Digital radiography; Algorithm; Pediatric patients; Dose reduction; Phantoms; FLAT-PANEL-DETECTOR; CHEST RADIOGRAPHY; OPTIMIZATION;
D O I
10.1007/s00330-021-07942-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To evaluate whether the advanced spatial noise reduction (ASNR) algorithm installed in a digital radiography system generates acceptable images at a lower dose than a conventional denoising algorithm in pediatric patients. Methods Nine sets of 30 images of pediatric patients, classified under three protocols and three age groups, were retrospectively selected. Different levels of low-dose image sets of these 270 images were generated by a noise simulation tool after validation testing using phantoms. Each image set was obtained with both the ASNR and conventional algorithm, and grouped randomly and blinded. Three experienced pediatric radiologists were asked to pick the "image with optimum dose" among images of different dose levels with an ALARA (as low as reasonably achievable) perspective. Dose reduction rates for each protocol and age group were calculated, and entrance skin exposure (ESE) was calculated using the values of kVp and mAs, assuming a standard body depth for each age group. Results With the ASNR algorithm, estimated dose reduction rates were highest for abdominal radiographs (45.0%, 27.3%, and 24.3% in infants, children, and adolescents, respectively, p < 0.001). The mean dose reductions for all age groups in the abdominal, chest, and skull radiographs were 32.8%, 12.9%, and 23.2%, respectively (p < 0.001). Average of the calculated ESE was lower with the ASNR algorithm than with the conventional algorithm group (p < 0.001). Conclusions The ASNR algorithm facilitated optimization of image quality with a higher reduction in radiation dose than the conventional algorithm, making it more acceptable for use in pediatric patients.
引用
收藏
页码:8937 / 8946
页数:10
相关论文
共 21 条
  • [1] [Anonymous], 2014, GEN CLIN PHARM CONS
  • [2] Method of simulating dose reduction for digital radiographic systems
    Båth, M
    Håkansson, M
    Tingberg, A
    Månsson, LG
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 253 - 259
  • [3] European Commission, 2000, European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Paediatrics
  • [4] EuroSafe Imaging, 2014, EUR DIGN REF LEV PAE
  • [5] Food and Drug Administration, 2017, PED INF XRAY IM DEV
  • [6] Investigation of optimum energies for chest imaging using film-screen and computed radiography
    Honey, ID
    MacKenzie, A
    Evans, DS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (929) : 422 - 427
  • [7] National Survey of Radiation Doses of Pediatric Chest Radiography in Korea: Analysis of the Factors Affecting Radiation Doses
    Kim, Bo Hyun
    Do, Kyung-Hyun
    Goo, Hyun Woo
    Yang, Dong Hyun
    Oh, Sang Young
    Kim, Hyeog Ju
    Lee, Kwang Yong
    Lee, Jung Eun
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (05) : 610 - 617
  • [8] A robust and fast non-local means algorithm for image denoising
    Liu, Yan-Li
    Wang, Jin
    Chen, Xi
    Guo, Yan-Wen
    Peng, Qun-Sheng
    [J]. JOURNAL OF COMPUTER SCIENCE AND TECHNOLOGY, 2008, 23 (02) : 270 - 279
  • [9] Paediatric x-ray radiation dose reduction and image quality analysis
    Martin, L.
    Ruddlesden, R.
    Makepeace, C.
    Robinson, L.
    Mistry, T.
    Starritt, H.
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2013, 33 (03) : 621 - 633
  • [10] Matrecano M, 2012, VISAPP C PAP 2012