Thermal damage of the specimen during breast biopsy with the use of the Breast Lesion Excision System: does it affect diagnosis?

被引:8
作者
Al-Harethee, Wasim Ali [1 ]
Kalles, Vasileios [1 ]
Papapanagiotou, Ioannis [1 ]
Matiatou, Maria [1 ]
Georgiou, Georgia [1 ]
Nonni, Afrodite [2 ]
Koulocheri, Dimitra [3 ]
Liakou, Paraskevi [1 ]
Theodoropoulos, George [1 ]
Zografos, George C. [1 ]
机构
[1] Univ Athens, Hippocratio Gen Hosp, Breast Unit, Propaedeut Surg Clin 1, Athens 17121, Greece
[2] Univ Athens, Sch Med, Dept Pathol, Athens 17121, Greece
[3] Univ Athens, Hippocratio Gen Hosp, Dept Radiol, Athens 17121, Greece
关键词
Breast Lesion Excision System; Non-palpable mammographic lesions; Radiofrequency; Percutaneous breast biopsy; Thermal damage; UNDERESTIMATION;
D O I
10.1007/s12282-013-0458-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Stereotactic vacuum assisted breast biopsy (VABB) procedures remain of key importance in the diagnostic evaluation of suspicious non-palpable mammographic lesions. The Breast Lesion Excision System(A (R)) (BLES) is an image-guided percutaneous biopsy method that utilizes radiofrequency (RF) in order to retrieve an intact-tissue biopsy specimen. The purpose of this study is to determine the effects of RF application on the tissue specimen and the possible interference with the histopathologic results. In this study we included 265 patients with suspicious non-palpable mammographic lesions who underwent 273 stereotactic VABB procedures using the BLES. All biopsies were performed by the same surgical-radiology team under local anesthesia. The specimen thermal damage was classified according to the pathology report in 3 categories: Severe (extensive thermal damage or inability to diagnose), medium (ability to diagnose but either circumferential damage > 1.5 mm or diffuse areas of thermal damage) and mild (circumferential thermal damage < 1.5 mm but > 0.5 mm). Radiofrequency-associated thermal damage of the specimen was observed in 14 cases (5.13 %), and was classified as severe in 5, medium in 5, and mild in 4 specimens. Within the group of RF damaged specimens, we found a significant (p < 0.05) positive correlation between fat cell content and classification of thermal damage. Although thermal damage is of concern during BLES breast biopsy, the incidence is low, and the outcome of the histopathologic assessment is not affected even in severely damaged specimens. Increased thermal damage seems to correlate with higher fat cell content of the specimen.
引用
收藏
页码:84 / 89
页数:6
相关论文
共 15 条
[1]   Breast Lesion Excision System - Intact®(BLES): A Novel Stereotactic Method of Biopsy Utilizing Radiofrequency for Suspicious Non-Palpable Mammographic Lesions [J].
Al-Harethee, W. A. ;
Papanagiotou, I. ;
Grigoropoulos, P. ;
Filippakis, G. M. ;
Zografos, G. .
CANCER RESEARCH, 2010, 70
[2]   Complications of percutaneous stereotactic vacuum assisted breast biopsy system utilizing radio frequency [J].
Al-Harethee, Wasim ;
Theodoropoulos, George ;
Filippakis, Georgios Michael ;
Papapanagiotou, Ioannis ;
Matiatou, Maria ;
Georgiou, Georgia ;
Kalles, Vasileios ;
Koulocheri, Dimitra ;
Nonni, Afroditi ;
Konstadoulakis, Manousos M. ;
Manouras, Andreas ;
Zografos, George C. .
EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) :623-626
[3]   The breast lesion excision system (BLES): a novel technique in the diagnostic and therapeutic management of small indeterminate breast lesions? [J].
Allen, Steven D. ;
Nerurkar, Ashish ;
Della Rovere, Guidabaldo U. Querci .
EUROPEAN RADIOLOGY, 2011, 21 (05) :919-924
[4]   The Fat Content of Small Primary Breast Cancer Interferes with Radiofrequency-Induced Thermal Ablation [J].
Athanassiou, E. ;
Sioutopoulou, D. ;
Vamvakopoulos, N. ;
Karasavvidou, F. ;
Tzovaras, G. ;
Tziastoudi, E. ;
Sakellariou, M. ;
Hatzitheofilou, C. .
EUROPEAN SURGICAL RESEARCH, 2009, 42 (01) :54-58
[5]   Benign breast diseases: Classification, diagnosis, and management [J].
Guray, Merih ;
Sahin, Aysegul A. .
ONCOLOGIST, 2006, 11 (05) :435-449
[6]   Comparison of the diagnostic accuracy of a vacuum-assisted percutaneous intact specimen sampling device to a vacuum-assisted core needle sampling device for breast biopsy: Initial experience [J].
Killebrew, Larry K. ;
Oneson, Ruth H. .
BREAST JOURNAL, 2006, 12 (04) :302-308
[7]   Cost-effectiveness of stereotactic 11-gauge directional vacuum-assisted breast biopsy [J].
Liberman, L ;
Sama, MP .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (01) :53-58
[8]   Percutaneous imaging-guided core breast biopsy: State of the art at the millennium [J].
Liberman, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (05) :1191-1199
[9]   Efficacy and cost-effectiveness of stereotactic vacuum-assisted core biopsy of nonpalpable breast lesions: analysis of 602 biopsies performed over 5 years [J].
Luparia, A. ;
Durando, M. ;
Campanino, P. ;
Regini, E. ;
Lucarelli, D. ;
Talenti, A. ;
Mattone, G. ;
Mariscotti, G. ;
Sapino, A. ;
Gandini, G. .
RADIOLOGIA MEDICA, 2011, 116 (03) :477-488
[10]   Image-guided breast biopsy: state-of-the-art [J].
O'Flynn, E. A. M. ;
Wilson, A. R. M. ;
Michell, M. J. .
CLINICAL RADIOLOGY, 2010, 65 (04) :259-270