A compact automatic low-level liquid scintillation system for radon-in water measurement by pulse pair counting

被引:8
|
作者
Gudjonsson, GI [1 ]
Theodorsson, P [1 ]
机构
[1] Univ Iceland, Inst Sci, IS-107 Reykjavik, Iceland
关键词
D O I
10.1016/S0969-8043(00)00156-1
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A novel compact liquid scintillation counting system with automatic changing of seven samples, dedicated to the assay of low levels of Rn-222 in water, is described. Its has a single phototube, and its operating mode is Rn-222 series selective, as only the close pulse pairs given by Bi-214 and its short lived decay product Po-214 (half life 0.16 ms), are counted. Its Bi-214/Po-214 counting efficiency is 90.4% and the background, with an unshielded detector, is 3 pulse pairs per hour. 75% of the radon in 200 mi water samples is transferred to the scintillator (15 mi toluene) by circulating air, which is bubbled through the two liquids for 4 min. With a counting time of 3 h, the minimum detection level is 8 mBq l(-1). The method is also suitable for various environmental and geophysical studies of Rn-222 and Ra-226. (C) 2000 Elsevier Science Ltd. All rights reserved.