MR imaging of the liver: Comparison between single-shot echo-planar and half-Fourier rapid acquisition with relaxation enhancement sequences

被引:18
作者
Yamashita, Y [1 ]
Tang, Y [1 ]
Namimoto, T [1 ]
Mitsuzaki, K [1 ]
Takahashi, M [1 ]
机构
[1] Kumamoto Univ, Sch Med, Dept Radiol, Kumamoto 860, Japan
关键词
liver neoplasms; MR; magnetic resonance (MR); echo planar; pulse sequences; rapid imaging;
D O I
10.1148/radiology.207.2.9577477
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate single-shot T2-weighted magnetic resonance sequences and their role in clinical practice in patients with hepatic lesions. MATERIALS AND METHODS: Prospective comparison of echo-planar and half-Fourier rapid acquisition with relaxation enhancement (RARE) imaging was performed in 80 patients with focal hepatic lesions. Spin-echo (SE) single-shot echo-planar (echo times, 47 and 80 msec) and half-Fourier RARE (echo time, 59 msec) images were compared with turbo SE (repetition time msec/echo time msec = 3,200-7,600/90) images. Quantitative, qualitative, and receiver operating characteristic (ROC) analyses were performed. RESULTS: For liver, signal-to-noise ratios on half-Fourier RARE images were significantly higher than those on echo-planar images (P < .01). For cystic lesions, contrast on half-Fourier RARE and echo-planar images was slightly higher tan that on turbo SE images. For solid lesions, contrast on echo-planar images was better than that on half-Fourier RARE or turbo SE images. Artifacts including ghosting, bowel motion, susceptibility difference, and chemical shift were negligible on half-Fourier RARE images in all patients, whereas susceptibility difference and chemical shift of various degrees were seen on all echo-planar images. On the basis of ROC analyses, tumor detection rates were significantly higher with half-Fourier RARE and turbo SE images than with echo-planar images (P < .01). CONCLUSION: Echo-planar images provide sufficient contrast to allow detection of both solid and cystic masses, but severe artifacts preclude routine use. Half-Fourier RARE images are free from artifacts (chemical shift and susceptibility) and diagnostic performance with them is similar to that with turbo SE images.
引用
收藏
页码:331 / 337
页数:7
相关论文
共 19 条
  • [1] ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    FELMLEE, JP
    GRIMM, RC
    [J]. RADIOLOGY, 1993, 189 (01) : 259 - 264
  • [2] CHIEN D, 1992, AM J NEURORADIOL, V13, P1097
  • [3] ECHO-PLANAR MR-IMAGING
    EDELMAN, RR
    WIELOPOLSKI, P
    SCHMITT, F
    [J]. RADIOLOGY, 1994, 192 (03) : 600 - 612
  • [4] Liver masses: Replacement of conventional T2-weighted spin-echo MR imaging with breath-hold MR imaging
    Gaa, J
    Hatabu, H
    Jenkins, RL
    Finn, JP
    Edelman, RR
    [J]. RADIOLOGY, 1996, 200 (02) : 459 - 464
  • [5] VALUE OF T1-RELAXATION AND T2-RELAXATION TIMES FROM ECHOPLANAR MR IMAGING IN THE CHARACTERIZATION OF FOCAL HEPATIC-LESIONS
    GOLDBERG, MA
    HAHN, PF
    SAINI, S
    COHEN, MS
    REIMER, P
    BRADY, TJ
    MUELLER, PR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (05) : 1011 - 1017
  • [6] RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR
    HENNIG, J
    NAUERTH, A
    FRIEDBURG, H
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) : 823 - 833
  • [7] ECHO-PLANAR PERFUSION-SENSITIVE MR-IMAGING OF ACUTE CEREBRAL-ISCHEMIA
    KUCHARCZYK, J
    VEXLER, ZS
    ROBERTS, TP
    ASGARI, HS
    MINTOROVITCH, J
    DERUGIN, N
    WATSON, AD
    MOSELEY, ME
    [J]. RADIOLOGY, 1993, 188 (03) : 711 - 717
  • [8] MR cholangiopancreatography using HASTE (half-Fourier acquisition single-shot turbo spin-echo) sequences
    Miyazaki, T
    Yamashita, Y
    Tsuchigame, T
    Yamamoto, H
    Urata, J
    Takahashi, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (06) : 1297 - 1303
  • [9] ABDOMINAL MR-IMAGING - EVALUATION OF A FAST SPIN-ECHO SEQUENCE
    OUTWATER, EK
    MITCHELL, DG
    VINITSKI, S
    [J]. RADIOLOGY, 1994, 190 (02) : 425 - 429
  • [10] COMPARISON OF BREATH-HOLD FAST SPIN-ECHO AND CONVENTIONAL SPIN-ECHO PULSE SEQUENCES FOR T2-WEIGHTED MR-IMAGING OF LIVER-LESIONS
    RYDBERG, JN
    LOMAS, DJ
    COAKLEY, KJ
    HOUGH, DM
    EHMAN, RL
    RIEDERER, SJ
    [J]. RADIOLOGY, 1995, 194 (02) : 431 - 437