Accuracy and precision of a new portable ultrasound scanner, the Biocon-700, in residual urine volume measurement

被引:20
作者
Cho, Moon Kyoung [1 ]
Noh, Eun Ji [1 ]
Kim, Chul Hong [1 ]
机构
[1] Chonnam Natl Univ, Dept Obstet & Gynecol, Sch Med, 8 Hakdong, Gwangju, South Korea
关键词
Portable ultrasound scanner; Residual urine volume; Biocon-700; BLADDER; CATHETERIZATION;
D O I
10.1007/s00192-016-3224-0
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
The aim of this study was to examine the accuracy of a new portable abdominal ultrasound (US) machine (Biocon-700, Mcube Technology, Seoul, Korea) for measuring postvoid residual (PVR) urine. One-hundred and twenty-five patients with voiding problems underwent PVR measurements, which were compared with the postvoid bladder volume measurements obtained by catheterization within 5 min in the same patients. Accuracy was assessed by comparing the two methods, and correlation coefficients were obtained. The Pearson's correlation coefficient between PVR volume as measured by US and by catheterization was 0.872 (R (2) = 0.76). The mean difference was 23.59 +/- 37.32 ml (95% confidence interval, 17.5-30.65 ml), and the absolute scan error tended to increase with a greater amount of catheterized residual urine volume (R (2) = 0.49). Accuracy of the Biocon-700 scanner is as good as catheter assessments of residual volume and is therefore a sufficient alternative to catheterization for determining residual urine volume.
引用
收藏
页码:1057 / 1061
页数:5
相关论文
共 17 条
[1]  
Abdel-Fattah M, 2005, J Obstet Gynaecol, V25, P186
[2]   The accuracy of portable abdominal ultrasound equipment in measuring postvoid residual volume [J].
Alnaif B. ;
Drutz H.P. .
International Urogynecology Journal, 1999, 10 (4) :215-218
[3]   Measurement of bladder volume following cesarean section using Bladderscan [J].
Barrington, JW ;
Edwards, G ;
Ashcroft, M ;
Adekanmi, O .
INTERNATIONAL UROGYNECOLOGY JOURNAL, 2001, 12 (06) :373-374
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Chan H, 1993, J Neurosci Nurs, V25, P309
[6]   Accuracy and precision of a new portable ultrasound scanner, the BME-150A, in residual urine volume measurement: a comparison with the BladderScan BVI 3000 [J].
Choe, Jin Ho ;
Lee, Ji Yeon ;
Lee, Kyu-Sung .
INTERNATIONAL UROGYNECOLOGY JOURNAL, 2007, 18 (06) :641-644
[7]   THE ACCURACY OF PORTABLE ULTRASOUND SCANNING IN THE MEASUREMENT OF RESIDUAL URINE VOLUME [J].
COOMBES, GM ;
MILLARD, RJ .
JOURNAL OF UROLOGY, 1994, 152 (06) :2083-2085
[8]  
Cooperberg M R, 2000, Urology, V55, P590, DOI 10.1016/S0090-4295(99)00565-8
[9]  
Fuse H., 1996, International Urology and Nephrology, V28, P633, DOI 10.1007/BF02552157
[10]  
Ghadeer AS, 2009, J OBSTET GYNAECOL CA, V31, P526