摘要
Al3+ selective sensor has been fabricated from poly(vinyl chloride) (PVC) matrix membranes containing neutral carrier morin as ionophore. Best performance was exhibited by the membrane having composition as morin:PVC: sodium tetraphenyl borate: tri-n-butylphosphate in the ratio 5:150:5:150 (w/w, mg). This membrane worked well over a wide activity range of 5.0 x 10(-7) to 1.0 x 10(-1) M of Al3+, with a Nernstian slope of 19.7 +/- 10.1 mV/decade of Al3+ activity and a limit of detection 3.2 x 10(-7) M. The response time of the sensor is similar to 5 s and membrane could be used over a period of 2 months with good reproducibility. The proposed sensor works well over a pH range (3.5-5.0) and demonstrates good discriminating power over a number of mono-, di- and trivalent cations. The sensor can also be used in partially non-aqueous media having up to 20% (v/v) methanol, ethanol or acetone content with no significant change in the value of slope or working activity range. The sensor has also been used in the potentiometric titration of Al3+, with EDTA and for its determination in zinc plating mud and red mud. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1469 / 1473
页数:5
相关论文
共 22 条
[1]   Aluminium(III)-selective electrode based on a newly synthesized tetradentate Schiff base [J].
Abbaspour, A ;
Esmaeilbeig, AR ;
Jarrahpour, AA ;
Khajeh, B ;
Kia, R .
TALANTA, 2002, 58 (02) :397-403
[2]   Reverse flow-injection manifold for spectrofluorimetric determination of aluminum in drinking water [J].
Albendín, G ;
Mánuel-Vez, MP ;
Moreno, C ;
García-Vargas, M .
TALANTA, 2003, 60 (2-3) :425-431
[3]   Quantification of trace elements by sector field inductively coupled plasma mass spectrometry in urine, serum, blood and cerebrospinal fluid of patients with Parkinson's disease [J].
Bocca, B ;
Alimonti, A ;
Petrucci, F ;
Violante, N ;
Sancesario, G ;
Forte, G ;
Senofonte, O .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (04) :559-566
[4]  
CHRISTIAN GD, 2003, ANAL CHEM, P212
[5]   PVC MATRIX MEMBRANE ION-SELECTIVE ELECTRODES - CONSTRUCTION AND LABORATORY EXPERIMENTS [J].
CRAGGS, A ;
MOODY, GJ ;
THOMAS, JDR .
JOURNAL OF CHEMICAL EDUCATION, 1974, 51 (08) :541-544
[6]   DETERMINATION OF SELECTIVITY COEFFICIENTS OF ION-SELECTIVE ELECTRODES BY A MATCHED-POTENTIAL METHOD [J].
GADZEKPO, VPY ;
CHRISTIAN, GD .
ANALYTICA CHIMICA ACTA, 1984, 164 (OCT) :279-282
[7]   THE CLINICAL BIOCHEMISTRY OF ALUMINUM [J].
KING, SW ;
SAVORY, J ;
WILLS, MR .
CRC CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 1981, 14 (01) :1-20
[8]   Direct determination of trace aluminum with quercetin by reversed-phase high performance liquid chromatography [J].
Lian, HZ ;
Kang, WF ;
Bi, SP ;
Arkin, Y ;
Shao, DL ;
Li, D ;
Chen, YJ ;
Dai, LM ;
Gan, N ;
Tian, LC .
TALANTA, 2004, 62 (01) :43-50
[9]  
Lo Balbo A, 1998, ANAL LETT, V31, P2717
[10]  
Mousavi MF, 2001, ELECTROANAL, V13, P1125, DOI 10.1002/1521-4109(200109)13:13<1125::AID-ELAN1125>3.0.CO